期刊文献+

NiTi心血管支架的疲劳断裂性能分析 被引量:14

Fatigue and fracture behavior of Nitinol cardiovascular stents
下载PDF
导出
摘要 目的对NiTi合金心血管支架进行疲劳寿命的预测研究。方法采用疲劳断裂的有限元分析方法,建立了三维空间NiTi心血管支架结构的模型,并分析了在生理脉动循环载荷作用下的心血管支架结构。结果有限元分析发现支架结构结点附近区域的正应力大、疲劳寿命低、损伤大;同时,支架结构的疲劳寿命预测发现,用结构表面进行氮化、喷丸硬化等方法处理的支架疲劳寿命较高。相关的实验结果显示疲劳裂纹或断裂总是发生在支架结点附近区域。结论生理载荷下的模拟结果与相关的实验结果相吻合,最终为心血管支架结构的安全性指导和设计提供了理论依据。 Objective To carry out analysis on the fatigue lifetime prediction of nitinol cardiovascular stents.Meth-od By using the finite element method of fatigure fracture,the three-dimensional nitinol cardiovascular stents were established to analyze the structure of cardiovascular stents under a physiological pulsating cyclic loading.Results Great normal stress,low fatigue life and serious fatigue damage were found to be located near the stents’knots,and stents with nitrided surface or shot peened surface were likely to have a longer fatigue life.During miniature fatigue testing,early fatigue cracks or fractures were always formed at locations near stents’ knots.Conclusions The simulating results under the physiological loading are matched with related experimen-tal results,which provide a theoretical basis for the reliability and design of the cardiovascular stents.
出处 《医用生物力学》 EI CAS CSCD 2011年第1期1-6,共6页 Journal of Medical Biomechanics
基金 中国博士后科学基金(20080440194) 国家高技术研究发展计划863基金项目(2007AA04Z404) 科学技术项目(2009-K9-31) 高校科研项目(10KJD560003) 校级课题项目(XKY2010211)
关键词 NiTi心血管支架 有限元分析 疲劳断裂 正应力 数值模拟 应力分布 Nitinol cardiovascular stent Finite element analysis Fatigue fracture Normal stress Numerical simulation Stress distribution
  • 相关文献

参考文献3

二级参考文献33

  • 1何凡,李晓阳.血流与动脉壁的流固耦合研究[J].医用生物力学,2008,23(5):405-410. 被引量:18
  • 2乔爱科,刘有军,张松.S形动脉中的血流动力学研究[J].医用生物力学,2006,21(1):54-61. 被引量:11
  • 3Marks MP, Dake MD, Steinberg GK, et al. Stent placement for arterial and venous cerebrovascular disease preliminary experience [J].Neuroradiology,1994,191:441-446.
  • 4Higashida ,Randall ,SmithW,et al. Intravascular stent and endovascular coil placement for a ruptured fusiform aneurysm of the basilar artery [J]. Neurosurg ,1997,87:944- 949.
  • 5Horowitz B,Levy El, Koebbe J, et al. Transluminal stentassisted coil embolization of a vertebral confluence aneurysm technique report [J]. SurgNeurol,2001,552.91-296.
  • 6Ememann UU, Gronewoller E, Duffner FB, et al. Influence of geometric and hemodynamics parameters on aneurysm visualization during three-dimensional rotational angiography [ J ]. Am J Neuroradiol2003, 24:597-603.
  • 7Minsuok KM, Dale BT, Markus Tremmel, et al. Comparison of two stent in modifying cerebral aneurysm hemodynamics [J]. Annals of Biomedical Engineering,2008,36: 726-741.
  • 8Raghavan M, Harbaugh R. Quantified aneurysm shape and rupture risk[J]. Journal of Neurosurgery ,2005, 102: 355-362.
  • 9Alvaro V, Sergio B,Juan S, et al. Comparison of haemodynamics in cerebral aneurysms of different sizes located in the ophthalmic artery [J]. Numer Meth Fluids, 2007,53: 793-809.
  • 10Lieber BB,Livescu V, Hopkins LN,et al. Particle image velocimetry assessment of stent design influence on intra aneurysmal flow [J]. Ann Biomed Eng, 2002, 30:768 -775.

共引文献26

同被引文献182

引证文献14

二级引证文献82