离子选择电极法与偶氮胂Ⅲ法测定血钙的效果比较 被引量:5

Comparison of Determination of Serum Calcium by Ion Selective Electrode Assay and Arsenazo Ⅲ Assay
原文传递
导出
摘要 目的比较离子选择电极法和偶氮胂Ⅲ法测定血钙的结果,试图找出一种简便、灵敏、快速、准确的血清钙测定方法。方法随机选择南方医院门诊和住院病人标本200份,分别用离子选择电极法(仪器:贝克曼库尔特LX20)、比色法偶氮胂Ⅲ法(仪器:RocheIPE)检测血清钙的含量。结果离子选择电极法和偶氮胂Ⅲ法的批内、批间变异系数均<5%,但两组结果比较差异有统计学意义(P<0.01),偶氮胂Ⅲ法检测值明显比前者高;准确性符合要求;线性上限均为4.00mmol/L;两种方法对血钙测定的回归方程为Y=0.9306X+0.3576(r=0.9391,P<0.01);两种方法基本上不受黄疸、溶血及脂血的干扰。结论两种方法之间有较好的相关性,适合临床批量样本测试,但两种方法测定血清钙可能存在系统误差。 Objective By comparing the ion selective electrode(ISE) assay and arsenazo Ⅲ assay to find out a more convenient,sensitive,and accurate method for serum calcium measuremnt..Method Samples from outpatients and inpatients of Nanfang hospital were determined by ISE assay(Beckmen Coulte LX20) and arsenazo Ⅲ assay(Roche IPE).Result The within-run and between run CV for ISE assay and arsenazo Ⅲ assay were all less than 5%.Values by arsenazo Ⅲ assay were higher than those by ISE assay(P0.01).Precision and linear relationship met the requirment,and the linear range of determination was 0~4.00 mmol /L of serum calcium.The linear Regression equation for two assays was Y=0.9306X+0.3576(r=0.9391,P0.01).Lipemia,icterus and hemolysis nearly had no effects on serum calcium detection in both assays..Conclusion There was a linear relationship between serum calcium levels determined by two assays,however there might be system error between the ISE assay and the arsenazo Ⅲ assay.
出处 《热带医学杂志》 CAS 2010年第12期1393-1394,1397,共3页 Journal of Tropical Medicine
基金 广东省自然科学基金(No.07300312 8451051501000580)
关键词 血钙 离子选择电极法 偶氮胂Ⅲ法 serum calcium ion selective electrode assay arsenazo Ⅲ assay
  • 相关文献

参考文献4

  • 1Wang L, Manson JE,Song Y,et al.Systematic review:Vitamin D and calcium supplementation in prevention of cardiovascular events[J].Ann Intern Med, 2010, 152(5 ) : 315-323.
  • 2Maathuis FJ.Physiological functions of mineral macronutrients [ J ].Curr Opin Plant Biol, 2009, 12 ( 3 ) : 250-258.
  • 3Gemene KL,Bakker E.Measurement of total calcium by flash chronopotentiometry at polymer membrane ion-selective electrodes [ J ] .Anal Chim Acta, 2009,648 (2) : 240-245.
  • 4Ritter D,Mayo MM.Erroneous detection of hypercalcemia in specimens stored in Greiner Bio-One Vacuette Plasma Separator Tubes and analyzed by the Arsenazo Ill methodology [J]. Arch Pathol Lab Med,2009,133(9):1363-1364.

同被引文献49

引证文献5

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部