摘要 目的:建立一种氢化物原子荧光光谱法同时测定血液透析用水中的砷和锑的方法。方法:优化测定条件,采用氢化物原子荧光光谱法同时测定血液透析用水中的砷和锑。结果:砷和锑的检出限、相对标准偏差、回收率分别为:0.40μg/L和0.50μg/L;2.2%~4.2%和3.2%~5.3%;94.0%~101.2%和98.9%~103.6%。结论:本方法灵敏度高,准确、快捷,能同时测定血液透析用水中的砷和锑,省时、省力。 Objective:To establish a method for the simultaneous determination of arsenic and antimony in blood dialysis water by hydride atomic fluorescence spectrometry.Methods:The working condition of the determination were studied and optimized for the simultaneous determination of arsenic and antimony in blood dialysis water by hydride atomic fluorescence spectrometry.Results:The detection limits of this method were found to be 0.40 μg /L for arsenic and 0.50 μg /L for antimony,and the RSD were in the range of 2.2%~4.2% for arsenic and 3.2%~5.3% for antimony.The recovery tates of standard spiking for arsenic and antimony were in the range of 94.0%~101.2% and 98.9%~103.6%,respectively.Conclusion:The method is sensitive,accurate and rapid.It can be adopted for simultaneous determination of arsenic and antimony in blood dialysis water.
出处 《中国卫生检验杂志》 CAS 2011年第3期576-577,共2页 Chinese Journal of Health Laboratory Technology
关键词 氢化物原子荧光光谱法 透析用水 Hydride atomic fluorescence spectrometry Arsenic Antimony Blood dialysis water
  • 相关文献

参考文献2

二级参考文献10

  • 1刘明钟.原子荧光应用手册[M].北京:吉天仪器有限公司,2007.
  • 2NCCLS. Performance standanls for antimicrobial susceptibility testing [ S]. Twelfth Informational Supplement,2002,22( 1 ) :48 -49.
  • 3郭小伟,J Geochem Explor,1989年,33卷,237页
  • 4刘明钟,分析化学,1988年,16卷,1期,34页
  • 5张锦茂,岩矿测试,1986年,5卷,1期,27页
  • 6张锦茂,物探与化探,1984年,8卷,3期,150页
  • 7吴庭照,分析试验室,1983年,6卷,2期,99页
  • 8郭小伟,光谱学与光谱分析,1983年,3卷,2期,124页
  • 9郭小伟,分析化学,1980年,8卷,5期,466页
  • 10团体著者

共引文献58

同被引文献70

引证文献5

二级引证文献20

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部