期刊文献+

XT1800i血液分析仪单核细胞增高的原因分析

The analysis of the reason that the XT1800i blood analyzer raising monocyte
原文传递
导出
摘要 目的:评估XT-1800i血液分析仪对单核细胞分类的可信度。方法:对XT-1800i血液分析仪提示有单核细胞增高(>10%)的标本进行手工涂片、瑞氏染色、显微镜油镜复检。结果:以手工涂片、染色显微镜油镜复检为金标准,XT-1800i血液分析仪对单核细胞分类的灵敏度为83.33%,特异性为77.65%。阳性预测值为30.97%,阴性预测值为97.48%。结论:对XT-1800i血液分析仪显示单核细胞增高的标本都应结合手工白细胞分类计数法进行复检。 Objective:Evaluating the credibility of XT-1800i blood analyzer for monocyte classification.Methods:Making smear,using Wright's staining and observing microscope's oil immersion lens as the recheck for all monocyte sample(10%)which classifying by XT-1800i blood analyzer.Results:Taking smear,Wright's staining and microscope's oil immersion lens methods as the measure,the sensitivity of XT-1800i blood analyzer for monocyte classification is 83.33% and the specificity is 77.65%.Its positive predictive value and negative predictive value is 30.97% and 97.48%.Conclusion:Using manual leukocyte classification counting method on all monocyte sample which classifying by XT-1800I blood analyzer.
出处 《中国卫生检验杂志》 CAS 2011年第3期687-687,689,共2页 Chinese Journal of Health Laboratory Technology
关键词 血液分析仪 单核细胞 瑞-吉氏染色 显微镜 Credibility Monocyte Wright's staining Microscope
  • 相关文献

参考文献2

二级参考文献2

  • 1Barnes PW, McFadden SL, Machin SJ, et al. The International Consensus Group for Hematology Review: suggested criteria for action following automated CBC and WBC differential analysis. Lab Hematol, 2005, 11:83-90.
  • 2潘柏申,王玉亭,林镇海,冯仁丰,金大鸣.Sysmex SF-3000血液细胞分析仪应用评价[J].中华医学检验杂志,1998,21(5):318-319. 被引量:47

共引文献169

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部