双源CT泌尿系成像双能量虚拟平扫的临床评估 被引量:17

Clinical Application of Dual-source Computed Tomographic Urography Using Dual-energy Virtual Non-enhanced Computed Tomography
下载PDF
导出
摘要 目的评价FLASH双源CT泌尿系成像双能量虚拟平扫临床应用的可行性。方法对240例患者进行FLASH双源CT泌尿系成像,包括常规平扫、肾实质期双能量扫描(100kVp/230mAs和Sn140kVp/178mAs)和排泄期扫描。顺序注射造影剂100ml(370mgI/ml)和100ml生理盐水,注射速度均为4.5ml/s。利用LiverVNC软件处理得到虚拟平扫CT图像,与常规平扫比较在器官CT值、噪声、信号噪声比、图像质量和放射剂量等方面的差异。结果常规平扫和虚拟平扫所测器官CT值差异均无统计学意义(P>0.05),虚拟平扫图像噪声明显低于常规平扫而信号噪声比明显高于常规平扫(P<0.05),虚拟平扫图像质量低于常规平扫但差异无统计学意义(P>0.05),肾实质期双能量扫描放射剂量高于常规平扫和排泄期扫描(P<0.05)。结论在FLASH双源CT泌尿系成像中,双能量虚拟平扫具有取代常规平扫的可能,减少1次平扫的放射剂量,具有临床应用价值。 Objective To evaluate the clinical feasibility of dual-source computed tomographic urography using dual-energy virtual non-enhanced CT.Methods Totally 240 patients received dual-source CT,which included true non-enhanced CT (TNCT),nephrographic phase scanning with dual-energy mode (100kVp/230mAs and Sn140kVp/178mAs),and excretory phase scanning.A contrast bolus injection of 100 ml (370 mgI/ml) contrast agent was applied (4.5ml/s),followed by 100ml normal saline (4.5ml/s).Virtual non-enhanced CT (VNCT) image sets were reformatted from ‘Liver VNC’ software.The mean CT number,noise,signal to noise ratio (SNR),image quality,and radiation dose were compared between TNCT and VNCT image sets.Results There was no significant difference in mean CT numbers of all organs (P0.05).However,VNCT images had significantly lower noise and higher SNR than TNCT images (both P0.05).Image quality of VNCT was lower than that of TNCT without significant difference (P0.05).Radiation dose of nephrographic phase with dual-energy mode was significantly higher than that of TNCT and excretory phase scanning (P0.05).Conclusion Dual-energy VNCT requires less radiation dose and therefore has a potential to replace TNCT in the dual-source CT urography.
出处 《中国医学科学院学报》 CAS CSCD 北大核心 2010年第6期649-654,共6页 Acta Academiae Medicinae Sinicae
关键词 X线计算机 体层摄影术 泌尿系成像 虚拟平扫 信号噪声比 放射剂量 X-ray computed tomography urography virtual non-enhanced computed tomography signal to noise ratio radiation dose
  • 相关文献

参考文献3

二级参考文献16

  • 1Behrendt FF, Schmidt B, Plumhans C, et al. hnage fusion in dual energy computed tomography :effect on contrast enhancement, signal-tonoise ratio and image quality in computed tomography angiography. Invest Radiol,2009,44 : 1.
  • 2Nakayama Y. Low tube voltage reduces contrast material and radiation doses on 16-MDCT aortography. A JR,2006,187:490.
  • 3Menzel H, Schibilla H, Teunen D, et al. European guidelines on quality criteria for computed tomography. Luxembourg: European Commission ,2000 : 16262.
  • 4Deetjen A, Mollmann S,Conradi G, et al. Use of automatic exposure control in muhislice computed tomography of the coronaries : comparison of 16-slice and 64-slice scanner data with conventional coronary angiography. Heart ,2007,93 : 1040.
  • 5Scheffel H, Stolzmann P, Frauenfelde TR, et al. Dual-energy contrastenhanced computed tomography for the detection of urinary stone disease. Invest Radio1,2007 ,42 :823.
  • 6Graser A,Johnson TR, Chandarana H, et al. Dual energy CT: preliminary observations and potential clinical applications in the abdomen. Eur Radio1,2009,19 : 13.
  • 7Petersilka M, Bruder H, Krauss B, et al. Technical principles of dual source CT. Eur J Radiol,2008 ,68 :362.
  • 8Flohr TG, McCollough CH, Bruder H, et al. First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol,2006 ,16 :256.
  • 9Johnson TR, Nikolaou K, Wintersperger BJ, et al. Dual source CT cardiac imaging : initial experience. Eur Radiol,2006,16 : 1409.
  • 10Slovis TL. The ALARA conception in pediatric CT:myth or reality? Radiology ,2002,223:5.

共引文献112

同被引文献94

  • 1Van der Molen AJ, Cowan NC, Mueller-Lisse UG, et al. CT Urography Working Group of the European Society of Urogenital Radiology (ESUR). CT urography: Definitions, indications and techniques--a guideline for clinical practice. Eur Radiol, 2008,18 (1) :4-17,.
  • 2Chlapoutakis K, Theocharopoulos N, Yarmenitis S, et al. Per- formance of computed tomographic urography in diagnosis of up- per urinary tract urothelial carcinoma, in patients presenting with hematuria: Systematic review and meta-analysis. Eur J Radiol, 2010,73(2) : 334-338.
  • 3Nolte-Ernsting C, Cowan N. Understanding multislice CT urog- raphy teehniques: Many roads Lead to Rome. Eur Radiol, 2006, 16 (12) : 2670-2686.
  • 4O'Connor OJ, Maher MM. CT urography. Am J Roentgenol, 2010,195(5) :320-324.
  • 5Maheshwari E, O'malley ME, Ghai S, et al. Split-bolus MDCT urography: Upper tract opacification and performance for upper tract tumors in patients with hematuria. AJR Am J Roentgenol, 2010,194(2) :453-458.
  • 6Kekelidze M, Dwarkasing RS, Dijkshoorn ML, et al. Kidney and urinary tract imaging: Triple-bolus multidetector CT urography as a one-stop shop--protocol design, opacificatiort, and -.mage quality analysis. Radiology, 2010,255(2):508-516.
  • 7Meindl T, Coppenrath E, Degenhart C, et al. MDCT urography: Experience with a hi-phasic excretory phase examination proto- col. Eur Radiol, 2007,17(10):2512-2518.
  • 8Chow LC, Kwan SW, Olcott EW, et al. Split-bolus MDCT urog- raphy with synchronous nephrographic and excretory phase en- hancement. AJR Am J Roentgenol, 2007,189(2):314-322.
  • 9Dillman JR, Caoili EM, Cohan RH, et al. Comparison of urinary tract distension and opacifieation using single-bolus 3-Phase vs split-bolus 2-phase multidetector row CT urography. J Comput Assist Tomogr, 2007,31 (5) : 750-757.
  • 10Portnoy O, Guranda L, Apter S, et al. Optimization of 64-MDCT urography: E{-ect of dual-phase imaging with furose- mide on collecting system opacifieation and radiation dose. AJR Am J Roentgenol, 2011,197(5) :W882-W886.

引证文献17

二级引证文献119

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部