角膜生物参数对青光眼患者眼压测量影响的研究 被引量:2

Influence of corneal biomechanical properties on intraocular pressure measurements of the glaucoma patients
下载PDF
导出
摘要 目的:探讨角膜生物参数对青光眼患者眼压测量的影响。方法:对80例121眼青光眼患者进行眼反应分析仪(ocular response analyzer,ORA)与Goldmann压平眼压计(Goldmann applanation tonometer,GAT)测量,并用先进的OrbscanⅡ眼前节分析系统测量中央角膜厚度(central corneal thickness,CCT)。结果:平均矫正眼压(IOPcc)值17.41±5.62mmHg;平均GAT值15.76±6.06mmHg;IOPcc与角膜滞后性(cornealhysteresis,CH)有相关性(P=0.000;r=-0.236);IOPcc与GAT显著相关(P=0.000;r=0.857);IOPcc与CCT无相关性。结论:对已经诊断的青光眼患者,平均IOPcc值高于平均GAT值;随着CH的降低,IOPcc值有升高的趋势;且IOPcc值不受CCT值的影响。 AIM:To evaluate the influence of corneal biomechanical properties on intraocular pressure(IOP)measurements of the glaucoma patients.METHODS:The study included 121 eyes of 80 cases.All cases underwent IOP evaluation with the ocular response analyzer(ORA)and Goldmann applanation tonometry(GAT),and central corneal thickness(CCT)was measured with Orbscan Ⅱ.RESULTS:The mean IOPcc was 17.41±5.62mmHg;the mean GAT was 15.76±6.06mmHg.IOPcc was significantly associated with corneal hysteresis(CH)and GAT(P=0.000);IOPcc was not associated with CCT.CONCLUSION:For glaucoma patients,the mean IOPcc is higher than the mean GAT.The IOPcc tends to be higher,while the CH is lower.The IOPcc is not influenced by CCT.
出处 《国际眼科杂志》 CAS 2011年第4期620-621,共2页 International Eye Science
关键词 角膜生物力学参数 眼反应分析仪 角膜厚度 ORBSCAN Ⅱ眼前节分析系统 眼压 corneal biomechanics ocular responseanalyzer corneal thickness Orbscan II intraocular pressure
  • 相关文献

参考文献5

  • 1Martinez-de-la-Casa JM, Carcia-Feijoo J, Fernandez-Vidal A, et al . Ocular response analyzer versus Goldmanu applanation tonometry for intraocular pressure measurements. Invest Opbthalmol Vis Sei 2006 ; 47 (10) :4410-4414.
  • 2Doughty MJ, Jonuscbeit S. Effct of corneal thickness on intraocular pressure measurements with the pneumotonmeter, Goldmann appplanation tonometer, and Tono-pen. Graefes Arch Clin Exp Opthalmol 2007;245 (11) : 1603-1610.
  • 3Bhan A, Browing AC, Shah S, et al. Effect of corneal thickness on intraocular pressure measurements with the pneumotonometer, Goldmann applanatlon tonometer, and Tono-pen. Invest Opthalmol Vis Sci 2002 ;43 (5) :1389-1392.
  • 4Ko YC, Liu CJ, Hsu WM. Varying effects of corneal thickness on intraoeular pressure measurements with different tonometers. Eye 2005; 19 (3) :327-332.
  • 5Medeiros FA, Weinreb RN. Evaluation of the influence of corneal biomechanical properties on intraocular pressure measurements using the ocular response analyzer.J Glaucoma 2006 ; 15 ( 5 ) :364-370.

同被引文献24

  • 1Cheng AC, Fan D, Tang E, et al. Effect of corneal cmwature and corneal thickness on the assessment of intraocular pressure using noncontact tonometry in patients after myopic LASIK surgery. Cornea, 2006, 25(1): 26-28.
  • 2Chen S, Chen D, Wang J, et al. Changes in ocular response analyzer parameters after LASIK. J Refract Surg, 2010, 26 (4): 279-288.
  • 3Luce DA. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg, 2005, 31(1): 156-162.
  • 4Medeiros FA, Weinreb RN. Evaluation of the influence of corneal biomechanical properties on intraoeular pressure measurements us- ing the ocular response analyzer. J Glaucoma, 2006, 15 (5): 364-370.
  • 5Saad A, Lteif Y, Azan E, et al. Biomechanical properties of kerato- conus suspect eyes. Invest Ophthalmol Vis Sic, 2010, 51 (6): 2912-2916.
  • 6Whitacre MM, Stein RA, Hassanein K. The effect of corneal thick- ness on applanation tonometry. Am J Ophthalmol, 1993, 115(5): 592-596.
  • 7Hansen FK. A clinical study of the normal human central corneal thickness. Acta Ophthalmol (Copenh), 1971, 49(1): 82-89.
  • 8Mardelli PG, Piebenga LW, Whitacre MM, et al. The effect of ex- cimer laser photorefractive keratectomy on intraocular pressure measurements using the Goldmann applanation tonometer. Ophthal- mology, 1997, 104(6): 945-949.
  • 9Feltgen N, Leifert D, Funk J. Correlation between central corneal thickness, applanation tonometry, and direct intracameral lOP readings. Br J Ophtbalmol, 2001, 85(1): 85-87.
  • 10Liu J, Roberts CJ. Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis. J Cataract Refract Surg, 2005, 31(1): 146-155.

引证文献2

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部