期刊文献+

Experimental and simulation research on the influence of voltage amplitude,frequency and carrier gas flow rate on the performance of FAIMS 被引量:1

Experimental and simulation research on the influence of voltage amplitude,frequency and carrier gas flow rate on the performance of FAIMS
原文传递
导出
摘要 A micro sensor chip of High-field Asymmetric Waveform Ion Mobility Spectrometry(FAIMS) was designed and fabricated by inductively coupled plasma(ICP) etching on the both sides of silicon and double silicon-glass bonding,with dimensions of 18.8 mm×12.4 mm×1.2mm.The sample ions were created at ambient pressure by VUV lamp ion source,which was equipped with a 10.6 eV photo discharge lamp(λ=116.5 nm).The 2-pentanone was adopted to illustrate the influences of high-field rectangular asymmetric waveform voltage amplitude,frequency and carrier gas flow rate on the performance of FAIMS sensor chip.The experiment results showed that with the frequency or carrier gas flow rate increasing,or voltage amplitude decreasing,the FAIMS sensitivity increases,and that the resolution decreases with the increasing of the frequency or flow rate.The FAIMS simulation results based on the SIMION software was in agreement with the experimental results.The FAIMS detection sensitivity experiment showed that the FAIMS sensor chip can detect positive and negative ions simultaneously,and has detection sensitivity as low as 0.1 ppm for acetic acid. A micro sensor chip of High-field Asymmetric Waveform Ion Mobility Spectrometry(FAIMS) was designed and fabricated by inductively coupled plasma(ICP) etching on the both sides of silicon and double silicon-glass bonding,with dimensions of 18.8 mm×12.4 mm×1.2mm.The sample ions were created at ambient pressure by VUV lamp ion source,which was equipped with a 10.6 eV photo discharge lamp(λ=116.5 nm).The 2-pentanone was adopted to illustrate the influences of high-field rectangular asymmetric waveform voltage amplitude,frequency and carrier gas flow rate on the performance of FAIMS sensor chip.The experiment results showed that with the frequency or carrier gas flow rate increasing,or voltage amplitude decreasing,the FAIMS sensitivity increases,and that the resolution decreases with the increasing of the frequency or flow rate.The FAIMS simulation results based on the SIMION software was in agreement with the experimental results.The FAIMS detection sensitivity experiment showed that the FAIMS sensor chip can detect positive and negative ions simultaneously,and has detection sensitivity as low as 0.1 ppm for acetic acid.
出处 《Science China(Technological Sciences)》 SCIE EI CAS 2011年第6期1407-1414,共8页 中国科学(技术科学英文版)
基金 supported by the National Natural Science Foundation of China (Grant Nos 60706030, 20827007) the National Basic Research Program of China ("973" Project)(Grant No 2007CB310504)
关键词 载气流量 电压幅值 模拟软件 实验 频率 性能 电感耦合等离子体 检测灵敏度 FAIMS vacuum ultraviolet lamp MEMS rectangular asymmetric waveform voltage SIMION
  • 相关文献

参考文献2

二级参考文献42

  • 1纪军,张彦,王梨平,黄启斌.新型不对称场离子迁移谱检测技术及其应用[J].现代科学仪器,2005,22(3):9-12. 被引量:12
  • 2张东风,孔德义,梅涛,陶永春.离子迁移谱仪微型化的现状与进展[J].仪器仪表学报,2006,27(2):199-204. 被引量:22
  • 3李刚,陈强,赵建龙.离子迁移谱技术及其在生命分析化学中应用[J].现代仪器,2006,12(2):31-34. 被引量:4
  • 4M. D. Krebs, A. M.Zapata, E. G. Nazarov, R. A. Miller, I. S. Costa, A. L. Sonenshein, and C. t~. Davis, IEEE Sens. J. 5, 696 (2005).
  • 5H. Schmidt, F. Tadjimukhamedov, I. V. Mohrenz, G. B. Smith, and G. A. Eiceman, Anal. Chem. 76, 5208 (2004).
  • 6G. A. Eiceman, E. G. Nazarov, B. Tadjikov, and R. A. Miller, Field Anal. Chem. Technol. 4, 297 (2000).
  • 7G. A. Eiceman, E. V. Krylov, and N. S. Krylova, Anal. Chem. 76, 4937 (2004).
  • 8R. W. Purves and R. Guevremont, Anal. Chem. 71, 2346 (1999).
  • 9R. Handy, D. A. Barnett, R. W.Purves, G. Horlick, and R. Guevremont, J. Anal. At. Spectrom. 15, 907 (2000).
  • 10E. V. Krylov, Tech. Phys. 44, 113 (1999).

共引文献10