期刊文献+

一种锥束CT散射场检测优化方法 被引量:8

Optimized method of scatter field detection for cone beam CT
下载PDF
导出
摘要 针对采用射束衰减网格的散射校正方法需额外增加较大扫描量的问题,提出了基于局部极值点扫描的检测优化方法,首先采集少量用于计算散射图像的辅助投影图像,然后采用三次样条插值由少量不同扫描角度的散射图像计算出完整的圆周扫描散射图像集,最后通过与原方法获取的散射图像及散射校正切片图像的对比分析,评估该方法的可行性与可靠性。实验结果表明,该方法仅需获取原方法1/20的散射图像即可完成锥束CT的散射校正,使扫描效率提高了47.6%,同时将散射校正切片误差控制在1%之内。 Aiming at reducing scanning amount in scatter field detection approach based on beam attenuation grid method,this paper proposes an optimized method.First,a small amount of assistant projections are acquired,which are used to calculate scatter images.And then,cubic spline interpolation is used to calculate complete circular scatter images from a few of scatter images in different scanning angles.Finally,the feasibility and reliability of the method is evaluated through comparing the obtained images with the scatter images acquired using beam attenuation grid method and the scatter correction slice images.Experiment results indicate that our method only needs 1/20 scatter images of original method to complete scatter correction,and increases the scanning efficiency by 47.6%;and the slice scatter correction error is 1%.
出处 《仪器仪表学报》 EI CAS CSCD 北大核心 2011年第4期781-786,共6页 Chinese Journal of Scientific Instrument
基金 国家科技支撑计划重点项目(2006BAF04B02) 西北工业大学基础研究基金(JC200934) 西北工业大学研究生创业种子基金(Z2010025)资助项目
关键词 锥束CT 散射 射束衰减网格 局部极值点 扫描角度插值 cone beam CT scattering beam attenuation grid local extreme value point scanning angle interpolation
  • 相关文献

参考文献15

  • 1张顺利,张定华,黄魁东,程云勇.锥束ART算法快速图像重建[J].仪器仪表学报,2009,30(4):887-892. 被引量:19
  • 2张顺利,张定华,李明君,吉军.基于SIMD技术的锥束ART算法快速并行图像重建[J].仪器仪表学报,2010,31(3):630-634. 被引量:6
  • 3SIEWERDSEN J H, JAFFRAY D A. Cone-beam computed tomography with a fiat panel imager: magnitude and effects of X-ray scatter[J]. Medical Physics, 2001, 28 (2) : 220-231.
  • 4ZHU L, STROBEL N, FAHRIG R. X-ray scatter correction for cone-beam CT using moving blocker array [ C ]. Medical Imaging 2005 : Physics of Medical Imaging, San Diego, CA, USA, 2005(5745):251-258.
  • 5KYRIAKOU Y, KALENDER W A. X-ray scatter data for flat-panel detector CT [ J ]. Physical Medica, 2007,23 (1) : 3-15.
  • 6武杰,陈家璧,张学龙,程静海.基于微焦点X射线源的相位衬度成像实验研究[J].电子测量与仪器学报,2010,24(7):621-625. 被引量:2
  • 7CAI W X, NING R L, CONOVER D. Scatter correction using beam stop array algorithm for cone-beam CT breast imaging [ C ]. Medical Imaging 2006: Physics of Medical Imaging, San Diego, CA, USA, 2006(6142): 61423E-1-61423E-9.
  • 8GRAHAM S A, MOSELEY D J, SIEWERDSEN J H, et al. Compensators for dose and scatter management in cone-beam computed tomography [ J ]. Med Phys, 2007, 34(7) : 2691- 2703.
  • 9SIEWERDSEN J H, MOSELEY D J, BAKHTIAR B, et al. The influence of antiscatter grids on soft- issue detectability in cone-beam computed tomography with flat-panel detectors[J]. Med Phys. 2004, 31(12) : 3506-3520.
  • 10MALUSEK A, SANDBORG M, CARLSSON G A. CT- mod-A toolkit for Monte Carlo simulation of projections including scatter in computed tomography [ J ]. Computer Methods and Programs in Biomedicine, 2008, 90 ( 2 ) : 167-178.

二级参考文献56

共引文献142

同被引文献180

引证文献8

二级引证文献68

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部