期刊文献+

Sysmex XE-2100全自动血液分析仪计数胸腹水有核细胞的评价 被引量:7

Evaluation of automatic hematology analyzer Sysmex XE-2100 for counting nucleated cells in hydrothorax and ascite fluid
下载PDF
导出
摘要 目的观察Sysmex XE-2100全自动血液分析仪(简称XE-2100)白细胞计数通道(DIFF通道)计数胸腹水有核细胞的能力表现。方法按胸腹水中有核细胞的含量分为低浓度、高浓度2组标本,使用XE-2100(荧光化学染色光学法)检测,将所获得的DIFF通道数据与手工显微镜镜检法相比较。结果对于胸腹水总浓度、低浓度、高浓度区间DIFF通道与手工镜检法间相关系数(r)分别为0.988 2、0.928 5、0.976 8。结论 XE-2100DIFF通道与手工镜检法的相关性较好,可以适用于常规胸腹水标本的有核细胞计数。 Objective To observe the performance of counting nucleated cells in hydrothorax and ascite fluids by leucocyte counting channel(DIFF channel) of automatic hematology analyzer Sysmex XE-2100(XE-2100 analyzer).Methods The hydrothorax and ascite fluids were classified into 2 groups(low concentration group and high concentration group) according to the nucleated cell concentration.The samples were detected by XE-2100 analyzer(fluorescence dyeing optical assay),and the results from DIFF channel were compared with manual microscopy.Results The correlation coefficients(r) of total concentration,low concentration and high concentration between DIFF channel and manual microscopy in hydrothorax and ascite fluids were 0.988 2,0.928 5 and 0.976 8.Conclusions In the XE-2100 analyzer,the correlation between DIFF channel and manual microscopy is good,and it could be applied to count the nucleated cells in routine hydrothorax and ascite fluids.
出处 《检验医学》 CAS 北大核心 2011年第4期228-230,共3页 Laboratory Medicine
关键词 胸腹水 有核细胞计数 全自动血液分析仪 Hydrothorax and ascite fluid Nucleated cell counting Automatic hematology analyzer
  • 相关文献

参考文献2

  • 1National Committee for Clinical Laboratory Standards. Reference leukocyte (WBC) differential count (proportional) and evaluation of instrumental methods [S]. H20-A, NCCLS, 1992.
  • 2Guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting and cell marker applications. International Council for Standardization in Haematology: prepared by the ICSH Expert Panel on Cytometry[J]. Clin Lab Haematol, 1994,16(2):157- 174.

同被引文献42

  • 1李洪军.血液分析仪发展概况[J].邯郸医学高等专科学校学报,2004,17(6):528-528. 被引量:1
  • 2王志富,臧玉龙,付艳荣,王信强,姜波.利用三分类血细胞分析仪快速做脑脊液和胸腹水细胞计数及分类的方法[J].医学检验与临床,2006,18(5):79-80. 被引量:7
  • 3熊立凡,刘成玉.临床检验基础[M].4版.北京:人民卫生出版社,2010:250.
  • 4National Committee for Clinical Laboratory Standards.Reference leukocyte (WBC) differential count ( propor- tional) and evaluation of instrumental methods, H20-A [S]. Wayne, PA : NCCLS, 1992.
  • 5Guidelinse for the evaluation of blood cell analysers inclu- ding those used for differential leucocyte and reticulocyte counting and cell marker applications. International Coucil for Standardization in Haematology: prepared by the IC- SH Expert Panel on Cytomtry[J].Clin Lab Haematol, 1994,16(2) : 157-174.
  • 6Cruntu FA, Benea L. Spontaneous bacterial peritonitis pathogen esis,diagnosis, treatment[J]. J Gastrointestin Liver Dis, 2006,15 (1):51 56.
  • 7Rimola A, Garcla Tsao G, Navasa M, et al. Diagnosis, treatment and prophylaxis of spontaneous bacterial peritonitis: a consensus document. International Ascites Club[J]. J Hepatol, 2000,32 ( 1 ) 142 153.
  • 8Angeloni S, Nicolini G, Merli M, et al. Validation of automated blood cell counter for the determination of polymorphonuclear cellcount in the aseitic fluid oi cirrhotic patients with or without spontaneous bacterial peritonitis[J]. Am J Gastroenterol,2003,98 (8) :1844 1848.
  • 9Cereto F,Genescci J, Segura R. Validation o automated blood cell counters }'or the diagnosis o}' spontaneous bacterial peritonitis[J]. Am J Gastroenterol,2004,99(7) 1400.
  • 10周臣敏.临床胸腹水常规检查的质量控制[J].现代医药卫生,2007,23(23):3575-3576. 被引量:1