摘要 目的目前在国内使用的心脏再同步治疗(CRT)起搏器及心脏再同步治疗除颤器(CRT-D)中,仅美敦力公司的Concerto具有左心室起搏阈值管理(LVCM)功能。本文初步探索植入Concerto患者,LVCM功能的准确性及临床意义。方法5例药物治疗无效的严重心力衰竭患者植入了Concerto型CRT—D,术后打开LVCM功能,起搏器自动记录每天起搏阈值变化情况,并在每一个随访时间点收集LVCM自动测算的左心室起搏阈值并与人工测算值进行对比。结果所有患者术后在满足测试条件后自动开启LVCM功能,并在事先设置的时间点自动测试起搏阈值,根据阈值自动调整输出;术后1~2个月起搏阚值波动较大。LVCM测算的左心室起搏阈值与手动测算值完全符合率为83.3%(20/24),如果按差值0.5V计算,二者符合率为96.8%(23/24)。结论LVCM功能与手工测试具有很高的符合率;使用此功能能最大限度地保证左心室或双心室有效同步起搏,对提高CRT术后疗效具有重要临床意义。 Objective Medtronic Concerto cardiac resynchronization therapy defibrillator (CRT-D) is the only one device used in China's Mainland with left ventricular capture management (LVCM) algorithm. Our aim was to assess the clinical application and accuracy of LVCM in patients with Concerto CRT-D implantation. Methods Concerto CRT-D systems were implanted into 5 patients with drug-resistant heart failure. Daily fluctuation of LV thresholds acquired by the LVCM algorithm up to 6-month-follow-up. LV threshold measurements by LVCM were compared with manual LV threshold tests at periodic visits. Results All high threshold detections and all automatic modulation of LV pacing output were adjusted appropriately, LV threshold fluctuated markedly in first 2 months. LVCM tests were identical to manual LV threshold tests in 20 out of 24 attempts(83.3% ). The proportion of LVCM threshold tests within one programming step of the manual threshold test was 96. 8% (23/24). Conclusion This study demonstrated that LVCM algorithm is accurate, reliable, and clinically equivalent to the manual LV threshold test. LVCM could uhimately result in complete LV capture and improve clinical effects of CRT.
出处 《中华心律失常学杂志》 2011年第2期90-93,共4页 Chinese Journal of Cardiac Arrhythmias
关键词 心脏再同步治疗 左心室起搏阈值管理 起搏 阈值 Cardiac resynchronization therapy Left ventricular capture management Pacing Threshold
  • 相关文献

参考文献9

  • 1Alonso C, Leclercq C, d' Allonnes FR, et al. Six year experience of transvenous left ventricular lead implantation for permanent biven- tricular pacing in patients with advanced heart failure:technical as- pects. Heart, 2001,86:405-410.
  • 2Albertsen AE, Nielsen JC, Pedersen AK, et al. Left ventricular lead performance in cardiac resynchronization therapy :impact of lead lo- calization and complications. PACE ,2005,28:483-488.
  • 3Iuliano A, Shopova G, De Simone A, et al. Longterm performance of corona sinus leads used for cardiac resynchronization therapy. PACE, 2007,30 : S47-S49.
  • 4张澍,华伟,黄德嘉,王景峰,吴立群,杨杰孚,曹克将,黄从新,王方正,陈新,无.植入性心脏起搏器治疗——目前认识和建议(2010年修订版)[J].中华心律失常学杂志,2010,14(4):245-259. 被引量:87
  • 5Knight BP, Desai A, Coman 3,et al. Long-term retention of cardiac resynchronization therapy. J Am Coil Cardlol,2004 ,44 :72-77.
  • 6Burri H, Gerritse B, Davenport Lynn, et al. Fluctuation of left ven- tricular thresholds and required safety margin for left ventricular pacing with cardiac resynchronization therapy. Europace,2009,11 : 931-936.
  • 7Biffi M,Bertini M,Saporito D,et al. Automatic management of left ventricular stimulation. PACE ,2009,32:346-353.
  • 8Crossley GH, Mead H, Kleekner K, et al. Automated left ventricular capture management. PACE ,2007,30 : 1190-1200.
  • 9Murgatroyd FD, Helmling E, Lemke B, et al. Manual vs. automatic capture management in implantable cardiac defibrillators and cardi- ac resynchronization therapy defibrillators. Europace, 2010, 12: 811-816.

二级参考文献150

  • 1Rasmussen K.Chronic sinus node disease:natural course and indications for pacing.Eur Heart J,1981,2:455-459.
  • 2Linde-edelstam C,Nordlander R,Pehrsson SK,et al.A doubleblind study of submaximal exercise tolerance and variation in paced rate in atrial synchronous compared to activity sensor modulated ventricular pacing.PACE,1992,15:905-915.
  • 3Gammage M,Schofield S,Rankin I,et al.Benefit of single setting rate responsive ventricular pacing compared with fixed rate demand pacing in elderly patients.PACE,1991,14:174-180.
  • 4Kay R,Estioko MWI.Primary sick sinus syndrome as an indication for chronic pacemaker therapy in young adults:incidence,clinical features,and long-term evaluation.Am Heart J,1982,103:338-342.
  • 5Shaw DB,Holman RR,Gowers JI.Survival in sinoatrial disorder (sick-sinus syndrome).Br Med J,1980,280:139-141.
  • 6Rubenstein JJ,Schulman CL,Yurchak PM,et al.Clinical spectrum of the sick sinus syndrome.Circulation,1972,46:5-13.
  • 7Fisher JD.Role of electrophysiologic testing in the diagnosis and treatment of patients with known and suspected bradycardias and tachycardias.Prog Cardiovasc Dis,1981,24:25-90.
  • 8Reiffel JA,Kuehnert MJ.Electrophysiological testing of sinus node function:diagnostic and prognostic application-including updated information from sinus node electrograms.PACE,1994,17:349-365.
  • 9Friedberg CK,Donoso E,Stein WG.Nonsurgical acquired heart block.Ann N Y Acad Sci,1964,111:835-847.
  • 10Johansson BW.Complete heart block.A clinical,hemodynamic and pharmacological study in patients with and without an artificial pacemaker.Acta Med Scand Suppl,1966,451:1-127.

共引文献86

同被引文献40

  • 1郭继鸿.深入认识心脏再同步化治疗心力衰竭的机制[J].中国心脏起搏与心电生理杂志,2006,20(4):283-284. 被引量:62
  • 2郭继鸿.心脏再同步化治疗适应证的进展与循证医学[J].心电学杂志,2007,26(1):8-10. 被引量:9
  • 3Lau C,Cameron DA,Nishimura SC,et al.A cardiac evoked response algorithm providing threshold tracking:A North American multicenter study.Pacing Clin Electrophysiol,2000,23:953-959.
  • 4Sperzel J,Siemon G,Presser HJ,et al.Benefit of enhanced ventricular capture management in automated pacemakers.Europace,2004,6(suppl):51.
  • 5Sperzel J,Kennergren C,Biffi M,et al.Clinical performance of a ventricular automatic capture verification algorithm.Pacing Clin Electrophysiol,2005,28:933-937.
  • 6De Voogt WG,Vonk BF,Albers BA,et al.Understanding capture detection.Europace,2004,6:561-569.
  • 7Curtis A,Maas S,Domijan A,et al.A method for analysis of the local atrial evoked response for determination of atrial capture in permanent pacing systems.Pacing Clin Electrophysiol,1991,14:1576-1581.
  • 8Butter C,Hartung W,Kay G,et al.Clinical validation of new pacing in pacemakers.J Cardiovasc Electrophysiol,2001,10:1104-1108.
  • 9Boriani G,Biffi M,Cameron D,et al.Atrial evoked response integral for automatic capture verification in atrial pacing.Pacing Clin Electrophysiol,2003,26:248-252.
  • 10Rueter J,Heymen H,Meisel E,et al.Automatic measurement of atrial pacing thresholds by AV conduction.Pacing Clin Electrophysiol,2000,23:658.

引证文献4

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部