摘要
介绍上海光源BL14W1 XAFS光束线实验站的极低浓度数据采集系统。该系统采用XFlash5040硅漂移固体探测器配合气体电离室、电流放大器及数据采集卡等仪器设备,以及用Labview软件编写的控制和采集程序。对Fe浓度为5.0μg/g的FeCl_3水溶液样品和1nm厚的Ni薄膜样品进行了XAFS谱的测量。测量结果表明,该光束线实验站已成功实现极低浓度数据采集系统。
In this paper, we report development of a data acquisition system for XAFS (X-ray absorption fine structure) analysis of samples of extremely low concentrations on the BL14W1 beamline at Shanghai Synchrotron Radiation Facility (SSRF). The system includes an XFlash 5040 silicon drift detector, ion chamber, amplifier, ADC etc. Labview software was used to write the data acquisition program. Fluorescence K-edge XAFS spectra of FeCl3 aqueous solution containing 5.0 μg/g Fe and a 1 nm Ni thin film were used to demonstrate performance of the system. The test results indicate that the data acquisition system ofBL14W1 XAFS beamline is capable of measuring samples of extremely low concentrations.
出处
《核技术》
CAS
CSCD
北大核心
2011年第5期341-344,共4页
Nuclear Techniques
基金
国家自然科学基金青年基金(批准号11005148)
中国科学院院长基金特别支持项目(批准号29)资助