Based on the multi-coincidence measurement, the time resolution of three liquid scintillation detectors (BC501A) were determined strictly by solving the coincidence equations, where the influence from electronics estimated by self coincidence measurement as well as the background had been considered. The result of this work agreed well with the result that was deduced from the traditional method, and it will be helpful to analyze the energy resolution of neutron time of flight spectra measured by using such detectors at CIAE (China Institute of Atomic Energy).
Based on the multi-coincidence measurement, the time resolution of three liquid scintillation detectors (BC501A) were determined strictly by solving the coincidence equations, where the influence from electronics estimated by self coincidence measurement as well as the background had been considered. The result of this work agreed well with the result that was deduced from the traditional method, and it will be helpful to analyze the energy resolution of neutron time of flight spectra measured by using such detectors at CIAE (China Institute of Atomic Energy).