摘要 快速无损检测大豆种子中的脂肪含量,在大豆品质育种和质量检测上至关重要。分别用索氏提取法、核磁共振仪及近红外谷物分析仪测定了5种国产大豆种子的脂肪含量,并对3种方法的测量结果进行了分析比较。结果表明:核磁共振法测定的结果与索式提取法更接近,偏差均小于0.95个百分点,最低偏差仅为-0.02个百分点;而近红外光谱法与索式提取法的偏差较大。 The fat contents of five domestic soybean cultivars seeds were analyzed by Soxhlet method,nuclear magnetic resonance analyzer(NMR) and near-infrared(NIR) grain analyzer respectively.The results showed that results of NMR closely tallied with that of Soxhlet method,the deviation between them was less than 0.95 percentage points,and the minimum deviation was only-0.02 percentage points.However,the result of NIR showed higher deviation from that of Soxhlet method than NMR did in this study.
出处 《河南农业科学》 CSCD 北大核心 2011年第3期62-64,共3页 Journal of Henan Agricultural Sciences
基金 国家转基因重大专项(2009ZX08009-026B) 山东省自然科学杰出青年基金项目(JQ200909)
关键词 大豆 脂肪含量 索氏提取法 核磁共振法 近红外光谱法 Soybean Fat content Soxhlet method Nuclear magnetic resonance Near infrared spectroscopy