期刊文献+

快速血糖仪与生化分析仪检测葡萄糖结果分析

Result Analysis of Blood Glucose Detection by Fast Blood Glucose Meter and by Biochemical Analyzer
下载PDF
导出
摘要 目的探讨快速血糖仪与全自动生化分析仪检测血糖结果的差异性。方法应用快速血糖仪及全自动生化分析仪对117例患者的末梢血和静脉血分别进行血糖检测,测得的数据用SPSS10.0统计学软件进行分析。结果快速血糖仪与全自动生化分析仪测得的末梢血和静脉血血糖值分别为(8.3±4.2)mmol/L、(8.4±4.1)mmol/L和(9.0±4.1)mmol/L,快速血糖仪测得值较全自动生化分析仪测得值低,差异有统计学意义(P〈0.05)。相关分析表明,三者呈正相关(P〈0.01)。结论快速血糖仪操作简便,可用于已确诊糖尿病患者血糖水平的监测。 Objective To study the difference in the blood glucose detection results by fast glucose meter and automatic biochemical analyzer.Methods Fast blood glucose meter and automatic biochemical analyzer were applied respectively in detecting blood glucose through terminal blood and venous blood of 117 cases,the data were analyzed by the software SPSS10.0.Result The blood glucose data of terminal blood and venous blood by fast blood glucose meter were(8.3±4.2)mmol/L while those by automatic blochemical analyzer were(9.0±4.1)mmol/L;the data by fast blood glucose meter were lower than those by automatic biochemical analyzer with a statistical difference(P〈0.05),and the relative analysis showed out a positive correlation among the three(P〈0.01).Conclusions Fast blood glucose meter is easy of operation and can be used for monitoring blood glucose level of the patients with confirmed diagnosis of diabetes.
出处 《西南军医》 2011年第3期422-423,共2页 Journal of Military Surgeon in Southwest China
关键词 糖尿病 血糖 快速血糖仪 全自动生化分析仪 diabetes blood glucose fast blood glucose meter automatic biochemical analyzer
  • 相关文献

参考文献3

  • 1American DiabetesAssociation. Consensus statement of self- monitoring ofblood glucose [ J]. DiabetesCare, 1987, 10 : 95.
  • 2Clarke WL, Cox D, Gonder - Frederick LA, et aL Evaluating clinical accuracy of systems for self - monitoring of blood glucose [ J]. Dia - betes Care, 1987,10 (5) :622 - 628.
  • 3丁红香,徐晓杰,张美芬,倪莉.POCT血糖仪与生化分析仪血糖检测结果的比对试验及分析[J].中华检验医学杂志,2007,30(12):1374-1375. 被引量:49

二级参考文献7

  • 1体外诊断检验系统自测用血糖监测系统通用技术条件.国家质量监督检验检疫总局发布.
  • 2《医疗机构临床实验室管理办法》及配套文件.卫生部临床检验中心和中国医院协会临床检验管理专业委员会编写.
  • 3EP9-A. Method comparison and bias estimation using patient samples, approved guideline. CLSI, 1995.
  • 4C30-A. Ancillary (bedside) blood glucose testing in acute and chronic care facilites; approved guideline CLSI. 2002.
  • 5The Diabes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med, 1993, 329:977-986.
  • 6UK Pospective Diabetes Study (UKPDS) Group. Intensive blood- glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes( UKPDS 33). Lancet, 1998,352:837-853.
  • 7王治国.无室间质量评价计划的临床检验项目如何进行实验室间比对.第四届全国临床检验实验室管理学术会议论文汇编.青岛,2006.43-45.北京:中国医院管理协会,2006.