血清尿素参考方法的建立以及临床常用试剂盒的比对研究 被引量:5

Establishment of Reference Method for Serum Urea Detection and Comparison with Routine Urea Kits
下载PDF
导出
摘要 目的建立血清尿素测定的酶偶联平衡的参考方法(紫外分光光度法),对此方法进行性能验证,并评价4种临床常用的试剂盒是否可以接受。方法以美国CDC推荐的血清尿素酶偶联平衡候选参考方法为依据建立本实验的参考方法;通过测定标准物质(SRM909b),以及参加RELA-2009比对实验(ring-tiral),来验证参考方法的准确性;按CLSI《EP9-A2》要求,以酶偶联平衡参考方法作为比对方法,4种临床常用的检测试剂盒为待评方法,对40份血清样品的结果进行分析比较。结果用酶偶联平衡参考方法测定SRM909b水平1和2,与靶值的偏倚分别为2.19%,0.24%;参加参考实验室国际比对,血清尿素结果在可接受范围内;4种常用的试剂盒在医学决定点(Xc=9.64 mmol/L)预期偏差的95%可信区间值在实验室定义的可接受偏差(4.5%)范围内,也在生物学变异的可接受偏差(5.5%)范围内。结论本室已建立并验证了血清尿素测定的酶偶联平衡参考方法;4种常用血清尿素试剂盒的测定结果与参考方法的测定结果存在差异,但误差在临床可接受范围。 Objective To establish and validate the coupled-enzyme equilibrium reference method as a reference method for urea detection and compare it with other 4 routine urea kits. Methods The reference method was verified by testing the two levels of urea in SRM909b and participating in RELA-2009.The EP9-A2 protocol was used to compare the results in forty serum samples in 4 routine kits methods with coupled-enzyme equilibrium reference method. Results The deviation of SRM909b level 1 was 2.19% and SRM 909b level 2 were 0.24% respectively through determining by the reference method.The results for RELA-2009 were within the accepted range.Comparing the confidence interval of the predicted bias with the definition acceptable error at the medical decision point Xc(Xc=9.64 mmol/L) for the 4 routine kits,the results were all within the range of defined acceptable error(4.5%) and the range of biological variation bias(5.5%). Conclusion The coupled-enzyme equilibrium reference method for serum urea was established in our lab.The urea results measured with 4 routine kits were different from results with the reference method,but the deviation was acceptable,and it will not affect the detection results.
作者 张雅君 张捷
出处 《首都医科大学学报》 CAS 北大核心 2011年第2期276-282,共7页 Journal of Capital Medical University
基金 国家科技支撑计划资助项目(2007BA105B09)~~
关键词 尿素 参考方法 酶偶联平衡法 分光光度法 紫外线 评价研究 urea reference method coupled-enzyme equilibrium spectrophotometry ultraviolet evaluation study
  • 相关文献

参考文献6

二级参考文献21

  • 1European Union. Directive 98/79/EC of the european parliament and of the council of 27 october 1998 on in vitro diagnostic medical devices.Official Journal of the European Communities, 1998, L331 : 1-37.
  • 2International Organization for Standardization. In vitro diagnostic medical devices-measurement of quantities in samples of biological originmetrological traceability of values assigned to calibrators and control materials. ISO/DIS 17511, International Organization for Standardization,Geneva,2000.
  • 3International Organization for standardization. In vitro diagnostic medical devices-measurement of quantities in samples of biological originmetrological traceability of values for catalytic concentration of enzymes assigned to calibrators and control materials. ISO/DIS 18153,International Organization for Standardization, Geneva, 2000.
  • 4International Organization for Standardization. General requirements for the competence of testing and calibration laboratories. ISO/IEC 17025,International organization for standardization, Geneva, 1999.
  • 5International Organization for Standardization. Quality management i0n the medical laboratory. ISO/FDIS 15189, International Organization for Standardization, Geneva, 2000.
  • 6Siekmann L. Traceability of calibration for IVD industry: Process for eredentialing reference materials, reference methods and reference laboratories in europe. In: proceedings of the workshop on measurement traceability for clinical laboratory testing and in vitro diagnostic test systems. National Institute of Standards and Technology, Gaithersburg MD, 2000. 85-101.
  • 7Eckfeldt JH. History of reference systema in clinical measurements. In :proceedings of the workshop on measurement traceability for clinical laboratory testing and in vitro diagnostic test systems. National institute of standards and Technology: Gaithersburg MD, 2000. 161-172.
  • 8International Organization for Standardization. In Vitro Diagnostic Medical Devices- Measurement of Quantities in Samples of Biological Origin-Description of Reference Materials. ISO 15194, 2003.
  • 9International Organization for Standardization. Clinical Laboratory Medicine-Requirements for Reference Measurement Laboratories. 1SO 15195,2003.
  • 10International Organization for Standardization. General Requirements for the Competence of Testing and Calibration Laboratories. ISO 17025, 1999.

共引文献152

同被引文献33

引证文献5

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部