摘要
目的探讨两种超声造影剂不同剂量和机械指数(MI)条件下的显像效果。方法分别对8只新西兰大白兔多次团注超声造影剂SonoVue(0.06ml/kg、0.12ml/kg)及Sonazoid(0.045ml/kg、0.09ml/kg),SonoVue组每个剂量水平分别设置7个MI值(O.10~0.55),Sonazoid组每个剂量水平分别设置5个MI值(O.19~0.98)。应用视频密度分析软件对显像过程做定量分析,并比较其峰值强度(PI)。结果在相同发射频率(7MHz)条件下:SonoVue组不同剂量组PI值均于MI=0.15时达最大值(P〈0.05);相同MI下,高剂量组PI值较大。Sonazoid组不同剂量组PI值均于MI=0.40时达最大值(P〈0.05);相同MI下,高剂量组PI值较大。结论SonoVue与Sonazoid显像条件不同,临床使用中需兼顾造影剂剂量、MI值及频率的影响。
Objective To compare the contrast enhancement effects of SonoVue and Sonazoid in rabbit liver at ultrasonic frequency of 7 MHz using various mechanical index(MI) and dosage. Methods Eight rabbits were bolus injected with different contrast agents:for SonoVue(0.06 ml/kg,0.12 ml/kg),MI ranging 0.10 - 0.55 ; while Sonazoid (0. 045 ml/kg, 0.09 ml/kg), MI ranging 0.19 - 0.98. Quantitative analysis of ultrasonography was done using video-intensity analysis system, and peak intensity (PI) was compared. Results With the same frequency, in SonoVue group, for both dosage groups, PI reached maximum at MI= 0.15 (P 〈0.05); with the same MI, PI of the high dosage group was higher (P〈0.05). In Sonazoid group,for both dosage groups,PI reached maximum at MI = 0.40( P 〈0.05) ;with the same MI,PI of the high dose group was higher ( P 〈0.05). Conclusions The optimistic ultrasonographic conditions provided to SonoVue and Sonazoid are different. Dosage of contrast agents, MI and frequency need to be concerned.
出处
《中华超声影像学杂志》
CSCD
北大核心
2011年第4期344-347,共4页
Chinese Journal of Ultrasonography
关键词
超声检查
微气泡
肝
Ultrasonography
Microbubbles
Liver