四种测氡仪器测量效果的比较研究 被引量:1

Comparison of measurement effect for four commonly used radon measuring instrument
原文传递
导出
摘要 目的对我国常用的四种现场氡测量仪器RAD7、RTM2100、RTM1688和PQ2000的性能和测量结果进行比较。方法选择一个普通房间和高氡地下室进行了氡气含量的实地测量和比较。结果 PQ2000稳定性和可靠性最好,两个浓度的测量值与氡室参考值的R均为2%;RTM2100实测值波动较大,低水平测量时应注意延长测量时间;RAD7型连续测氡仪需要进行干燥处理,否则结果会偏低。结论四种连续测氡仪在不同环境的测量结果的可靠性均在有关标准的允许范围之内。 Objective To compare the performance and measurement results of four types of radon measuring instruments RAD7,RTM2100,1688 and PQ2000,which are commonly used for field radon measurement.Methods A common room and a basement with high radon were selected for field measurements of the radon gas and comparison.Results The stability and reliability of the results for PQ2000 was the best,and the relative bias between measured values and reference value in the radon chamber were 2% for both two different radon concentrations.The fluctuation of actual measurements of RTM2100 was much larger,and the measurement time should be extended under low level measurement.RAD7 continuous radon monitor should be dried during measurement,otherwise the measurement results would be lower.Conclusion The reliability of measurement results for 4 continuous radon monitors were all within acceptable range of related standards in different environments.
作者 张继鏸 尚兵
出处 《中国预防医学杂志》 CAS 2011年第5期405-407,共3页 Chinese Preventive Medicine
基金 中国疾病预防控制中心青年科研基金课题(2009A201)
关键词 测氡仪器 性能 测量效果 RAD7 RTM2100 RTM1688 PQ2000 Radon monitor Performance Measurement results RAD7 RTM2100 RTM 1688 PQ2000
  • 相关文献

参考文献5

  • 1Tokonami S, Sanada T, Yang M. Contribution from thoron on the response of passive radon detectors [J]. Health Phys, 2001, 80 (6): 612 - 615.
  • 2WiegandJ, Feige S, Quingling X, et al. Radon and thoron in cave dwellings (Yan'an, China) [J]. Health Phys, 2000, 78 (4) :438- 444.
  • 3李利红,葛良全,程锋,田丽霞,敬伯科.空气中氡浓度时间变化规律探讨[J].辐射防护,2007,27(1):13-18. 被引量:8
  • 4崔宏星,曹吉生,尚兵.RTM—2100型Rn/Th连续测量仪测量特征[J].中国辐射卫生,2007,16(2):217-218. 被引量:5
  • 5Butterweck G, Schuler Ch, Paul A, et al. Intercomparison exercise of the PTb, BfS, MPA and PSI calibration facilities for radon [J].Radiat Prot Dosimetry , 2002, 98 (2): 219-222.

二级参考文献1

  • 1美国国家环保局.National Radon Proficiency Program:Guidance on Quality Assurance[R] EPA 402-R-95-012,1997.

共引文献11

同被引文献11

  • 1王南萍,肖磊,刘少敏,黄英,裴少英,高宝龙,程业勋.广东珠海地表介质中^(238)U、^(232)Th和^(40)K的活度水平[J].同位素,2005,18(1):73-78. 被引量:4
  • 2关康年,梁荣光,温小庭,黄忠.中山市居民住宅室内γ辐射水平与氡浓度调查[J].中国辐射卫生,2006,15(4):476-477. 被引量:4
  • 3Choubey V M,Ramola R C.Correlation between geology and radon levels in groundwater, soil and indoor air in Bhilangana Valley, Garhwal Himalaya, India[J].Environmental Geology,1997(4):4-32.
  • 4Vaupotic J,Andjelov M,Kobal I.Relationship between radon concentrations in indoor air and in soil gas[J].Environmental Geology,2002(42):583-587.
  • 5Andreja Popit, Janja Vaupotic. Indoor radon concentrations in relation to geology in Slovenia[J].Environmental Geology,2002( 42):330-337.
  • 6Sundal A V,Henriksen H,Soldal O.The influence of geological factors on indoor radon concentrations in Norway[J].Science of the Total Environment,2004( 328):41-53.
  • 7Wang Nanping,Xiao Lei,Li Canping,et al.Distribution and characteristics of radon gas in soil from a high-background-radiation city in China[J].Journal of Nuclear Science and Technology,2011,5(48):751-758.
  • 8Wang Nanping,Xiao Lei,Li Canping,et al.Determination of radioactivity level of 238U, 232Th and 40K in Surface Medium in Zhuhai City by in-situ Gamma-ray Spectrometry[J].Journal of Nuclear Science and Technology,2005,10(42):888-896.
  • 9涂彧,卢志娟,俞荣生,许哲.苏州市室内氡浓度水平及其影响因素研究[J].江苏预防医学,2010,21(5):8-11. 被引量:8
  • 10郭秋菊,程建平.珠海市环境空气中^(222)Rn、^(220)Rn子体水平及土壤析出率测量[J].辐射防护,2004,24(2):110-115. 被引量:10

引证文献1

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部