摘要 通过对原子吸收光谱法测定ZL205A铝合金中铜含量的不确定度的系统分析,阐述了测量结果不确定度主要来源于测量试液中铜的浓度、试液定容体积及样品质量产生的不确定度,并对这些分量进行了量化计算,最后计算出合成标准不确定度和扩展不确定度。同时通过评估,得出影响铜含量测定不确定度的主要因素是测量试液中铜浓度引起的不确定度。 The uncertainty for the determination of Cu in ZL205A aluminium alloy by AAS is analyzed. The sources of the uncertainty result from the determination of Cu concentration in the solution. The volume of the solution and the weight of the sample are discussed and calculated. At last the combined standard uncertainty and the expanded uncertainty are reported. At the same time, through evaluation, it was indicated that the main factor affecting the uncertainty for the determination of Cu is the uncertainty of Cu concentration in the solution.
作者 任慧 冯艳秋
出处 《现代科学仪器》 2011年第2期85-87,共3页 Modern Scientific Instruments
关键词 原子吸收光谱法 ZL205A铝合金 不确定度 AAS ZL205A aluminium Copper Uncertainty
  • 相关文献

参考文献6

二级参考文献6

  • 1国家质量技术监督局计量司.测量不确定度评定与表示指南[M].北京:中国计量出版社,2000.62-124.
  • 2刘智敏.不确定度及其应用[M].北京:中国计量出版社,2000..
  • 3.ISO4693-1986.铁矿石铜含量的测定-火焰原子吸收分光光度法[S].,..
  • 4.JJF1059-1999.测量不确定度评定与表示[S].,..
  • 5中国实验室国家认可委员会.化学分析中不确定度的评估指南[M].北京:中国计量出版社,2002..
  • 6韩永志.化学测量不确定度的评估和表示[M].北京:国家标准物质研究中心,1999..

共引文献858

同被引文献16

  • 1[1]上海市计量测试技术研究院.常用测量不确定度评定方法及应用实例.北京:中国计量出版社,2000
  • 2JJF1059-1999.测量不确定度评定与表示[S].[S].,..
  • 3JJF1059-1999.测量不确定度评定与表示[S].[S].,1999..
  • 4上海市计量测试技术研究院.常用测量不确定度评定方法及应用实例[M].北京:中国计量出版社,2002.
  • 5中国实验室国家认可委员会.化学分析中不确定度的评估[M].北京:中国计量出版社,2002.
  • 6GB/T20975.11-2008,铝及铝合金化学分析方法第11部分:锌含量的测定火焰原子吸收光谱法[S].
  • 7GB/T20975.11-2008,铝及铝合金化学分析方法第11部分:铅含量的测定火焰原子吸收光谱法[S],2008.
  • 8GB/T13748.9-2005,镁及镁合金化学分析方法铁含量的测定邻二氮杂菲分光光度法[S].
  • 9中国合格评定国家认可委员会.CNAS-CL01,检测和校准实验室能力认可准则[S].
  • 10中国实验室国家认可委员会.化学分析中不确定度的评估[M]北京:中国计量出版社,2002.

引证文献6

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部