摘要 基于SIMION 7.0离子光学仿真软件对设计的矩形离子阱(RIT)的离子捕获效率进行了仿真,研究了离子初始位置和射频电压、频率对离子捕获效率的影响。仿真结果表明,只有初始位置靠近RIT中心的离子才能被有效捕获到。针对常见毒品及爆炸物所在的质量范围,对于内建电场空间尺寸为10mm×8.4mm×44mm(x,y,z)的RIT,选择捕获射频电压为50V0-p~100V0-p、频率1MHz时,既能保证较高的离子捕获效率,同时也降低了对电路系统指标的要求。 The ion trapping efficiency of a miniature rectangular ion trap(RIT) was studied by simulation based on SIMION 7.0 ion optical simulation software.The effects of initial ion position,RF voltage and frequency on ion trapping efficiency were investigated.The simulation results showed that only ions generated near the center of RIT could be trapped successfully.For a RIT with inner field dimensions of 10mm×8.4mm×44mm(x,y,z),the trapping RF voltage of 50Vo-p^100Vo-p and RF frequency of 1MHz will ensure higher trapping efficiency for testing common drugs and explosives,at the same time the requirement on electronics can be reduced.
出处 《分析仪器》 CAS 2011年第3期44-47,共4页 Analytical Instrumentation
基金 "863"计划(2007AA04Z341) 国家自然科学基金(No.60871037)资助
  • 相关文献

参考文献3

二级参考文献10

共引文献10

同被引文献43

  • 1姜杰,费强,金伟,李明,庞晓东,李彬,范志勇,许晓宇,周建光,金钦汉.小型矩形离子阱质谱仪的研制[J].分析化学,2007,35(9):1387-1390. 被引量:13
  • 2Sparkman 0 D. Field-portable and miniature mass spectrometry[J]. Journal of the American Society for Mass Spectrometry, 2000, 11 (5) :468 -471.
  • 3West J, Becker M, Tombrink S, et al. Micro total analysis systems : Latest achievements [ J ]. Analytical Chemistry, 2008,80 ( 12 ) : 4403 -4419.
  • 4Badman E R, Graham Cooks R. Miniature mass analyzers [ J]. Journal of Mass Spectrometry ,2000,35 (6) :659 -671.
  • 5Xu W, Manicke N E, Cooks G R, et al. Miniaturization of mass spectrometry analysis systems [J]. Journal of the Association for Laboratory Automation ,2010,15 ( 6 ) :433 -439.
  • 6Syms R R A. Advances in micro-fabricated mass spectrometer- s[ J]. Analyical and Bioanalytical Chemistry, 2009, 393 ( 2 ) : 427 -429.
  • 7Stafford Jr G. Ion trap mass spectrometry:A personal perspec- tive [ J]. Journal of the American Society for Mass Spectrometry, 2002,13(6) :589 -596.
  • 8Orient O J, Chutjian A, Garkanian V. Miniature, high-resolution, quadrupole mass-spectrometer array [ J ]. Review of Scientific In- strnments, 1997,68 (3) : 1393 -1397.
  • 9Holkeboer D H, Karandy T L, Currier F C, et al. Miniature qua- drupole residual gas analyzer for process monitoring at milliTorr pressures[ J]. Journal of Vacuum Science & Technology A ,1998, 16(3) :1157-1162.
  • 10Kaiser Jr R E, Graham Cooks R, Stafford Jr G C, et al. Operation of a quadrupole ion trap mass spectrometer to achieve high mass/ charge ratios [ J ]. International Journal of Mass Spectrometry and Ion Processes, 1991,106 : 79 -115.

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部