期刊文献+

头发中内源性氢化可的松的高效液相色谱/串联质谱检测

Detection of endogenous cortisol in human hair by high performance liquid chromatography tandem mass spectrometry
下载PDF
导出
摘要 基于高效液相色谱/串联质谱(HPLC/MS/MS)建立了一种检测头发中内源性氢化可的松含量的方法,并与高效液相色谱/荧光法(HPLC/FLU)的测定结果进行比较.头发样品(50 mg)经过清洗与研磨、甲醇提取和C18固相萃取后,在电喷雾(ESI)负离子模式和多反应监测(MRM)方式下进行定性,以外标法进行定量.方法的定量限为2 pg/mg,工作曲线的线性范围为2~100 pg/mg;在空白头发添加50 pg/mg外标时,方法的回收率为(99.4±9.1)%,日内精密度为7.6%,日间精密度为8.3%.由于前处理提取方法、基质效应和定量方法不同,所以HPLC/MS/MS法的测定结果明显低于HPLC/FLU法的测定结果.尽管如此,皮尔逊相关性分析发现2种方法测定的结果具有显著相关性(r=0.749,p<0.01). A new method for detection of endogenous cortisol in human hair was developed based on high performance liquid chromatography tandem mass spectrometry(HPLC/MS/MS),and the results detected was compared with that determined by high performance liquid chromatography with fluorescence detector(HPLC/FLU).Hair sample(50 mg) was incubated in methanol after being washed and pulverized,and then the extraction was purified by C18 solid phase extraction column.Cortisol identification was performed by electro-spray ionization(ESI) in negative mode using multiple reaction monitoring(MRM) and the quantification was performed using external standard method.The limit of quantification is 2 pg/mg,and the linear range is 2 to 100 pg/mg.The recovery of cortisol spiked at 50 pg/mg is(99.4±9.1) %,intra-day precision is 7.6 %,inter-day precision is 8.3 %.Because of the differences in hair pretreatment method,cortisol extraction method,matrix effect and quantitative method,cortisol concentration detected by HPLC/MS/MS is significantly lower than that determined by HPLC/FLU.However,Pearson correlation analysis showed that there is a significant linear correlation(r=0.749,p〈0.01) between hair cortisol contents analyzed by these two methods.
出处 《东南大学学报(自然科学版)》 EI CAS CSCD 北大核心 2011年第3期579-583,共5页 Journal of Southeast University:Natural Science Edition
基金 国家自然科学基金资助项目(60771023) 教育部新世纪人才计划资助项目(NCET-08-0122) 江苏省社会科学基金重点资助项目(09IYA002) 东南大学博士生创新基金资助项目(BC0905)
关键词 高效液相色谱/串联质谱 氢化可的松 头发 高效液相色谱/荧光法 high performance liquid chromatography tandem mass spectrometry cortisol hair high performance liquid chromatography with fluorescence detector
  • 相关文献

参考文献17

  • 1谭仁祥.甾体化学[M].北京:化学工业出版社,2008:113-115.
  • 2Sauve B, Koren G, Walsh G, et al. Measurement of cortisol in human hair as a biomarker of systemic expo- sure [J]. Clin Invest Med, 2007, 30 (5): E183 - E191.
  • 3Kirschbaum C, Tietze A, Skoluda N, et al. Hair as a retrospective calendar of cortisol production-increased cortisol incorporation into hair in the third trimester of pregnancy [ J ]. Psychoneuroendocrinology, 2009, 34 (1): 32-37.
  • 4Ito N, Ito T, Kromminga A, et al. Human hair follicles display a functional equivalent of the hypothalamic-pitui- tary-adrenal (HPA) axis and synthesize cortisol [ J]. FASEB J, 2005, 19(8) : 1332 - 1334.
  • 5Irwin S, Kenny A P, Halloran O J, et al. Adaptation and validation of a radioimmunoassay kit for measuring plasma cortisol in turbot [ J]. Biochemistry and Physiol- ogy, 1999, 124(1) : 27-31.
  • 6Davenport M, Tiefenbacher S, Lutz C, et al. Analysis of endogenous cortisol concentrations in the hair of rhe- sus macaques [ J]. Gen Comp Endocrinol, 2006, 147 (3) : 255 -261.
  • 7Sarkar M, Das B C, Bora B D, et al. pplication of sen- sitive enzymeimmunoassay for determination of cortisol in blood plasma of yaks (Poephagus grunniens L. ) [ J ]. General and Comparative Endocrinology, 2007, 154(1/2/3) : 85 -90.
  • 8Magnisali P, Dracopoulou M, Mataragas M, et al. Routine method for the simultaneous quantification of 17 alpha-hydroxyprogesterone, testosterone, dehydroepi- androsterone, androstenedione, cortisol, and preg- nenolone in human serum of neonates using gas chroma- tography-mass spectrometry [J]. J Chromatogr A,2008, 1206(2) : 166 - 177.
  • 9Cirimele V, Kintz P, Dumestre V, et al. Identification of ten corticosteroids in human hair by liquid chromatog- raphy-ionspray mass spectrometry [J]. Forensic Sci Int, 2000, 107(1/2/3): 381-388.
  • 10Bevalot F, Gaillard Y, Lhermitte M A, et al. Analysis of corticosteroids in hair by liquid chromatography- electrospray ionization mass spectrometry[J]. J Chro- matogr B, 2000, 740 (2) : 227 - 236.