模拟定位CT床板对放射治疗剂量的影响 被引量:7

The study of different CT bed boards on the effects of radiation therapy
下载PDF
导出
摘要 目的研究不同定位CT床板对放射治疗剂量的影响。方法分别用碳纤床板、有机玻璃床板、普通CT凹形扫描床板、不用扫描床板对模体进行扫描,传输至计划系统,进行CT值采样,比较基于不同方法得到CT图像设计计划所得加速器输出量。结果与有机玻璃床板接触的模体部分CT值(如取样为1.25 cm及0.5 cm直径的圆相对偏差值分别为118.8%和205.6%),比不用扫描床板CT值大(P<0.01),其他扫描方法所得的CT值与不用扫描床板所得的CT值一致;在180°与有机玻璃床板接触的模体设射野时加速器输出量比不用扫描床板模体大0.9%,其他两种方法加速器输出量比不用扫描床板模体增大在0.5%内。结论射野穿过与有机玻璃床板接触的模体部位对剂量将有影响,而其他不同定位CT床板对放射治疗剂量的影响无临床意义;不同的治疗床板对剂量影响大小不一。 Objective To study the impact of different CT bed boards on radiotherapy dose distribution.Methods Phantom was scanned respectively with different bed boards which were carbon fiber bed board,plexiglass bed board,normal CT scan bed board and no bed board.Images of phantom were transferred to the planning system,and then CT values from different places of different bed boards were sampled,and the accelerator outputs obtained from different methods were compared.Results CT values of part in contact with the plexiglass bed board was 118.8% or 205.6%(sampling area was 1.25cm or 0.5cm diameter circle)higher than the part CT values with no bed.The results were statistically significant.In other locations the CT values were the same as other CT bed boards.The accelerator output with the images of plexiglass bed board was 0.9% bigger than the accelerator output with the images of no bed when gantry was 180°.Comparing with the other two methods which used carbon fiber bed board and normal CT scan bed board and the method which was with no bed board,the accelerator output error was within 0.5%.Conclusion The area in contact with the plexiglass bed board which is passed through by beam has some influence on the dose.Impact of different CT bed boards effects on radiation dose distribution will not cause clinically significant effect.Treatment of bed board has a certain impact on the dose.
出处 《临床肿瘤学杂志》 CAS 2011年第4期338-341,共4页 Chinese Clinical Oncology
关键词 模拟定位CT床板 放射肿瘤学 放射物理学 Bed boards Radiation oncology Radiophysics
  • 相关文献

参考文献4

二级参考文献13

  • 1盛延兴,隋振忠,郑芳霞,李学章,魏绪国,于书增.立体定向放疗床、头架、真空袋对吸收剂量的影响[J].现代肿瘤医学,2004,12(3):184-184. 被引量:16
  • 2马金利,蒋国梁,傅小龙,吴开良,李龙根.个体患者适形调强放疗计划模体内剂量实测验证技术的建立[J].中华放射医学与防护杂志,2007,27(1):80-84. 被引量:2
  • 3戴建荣 胡逸民.射野影像系统.肿瘤放射物理学[M].北京:原子能出版社,1999.463-466.
  • 4Intensity Modulated Radiation Therapy Collaborative Working Group. Intensity-modulated radiotherapy: current status and issues of interest. Int J Radiat Oncol Biol Phys, 2001,51(4) :880-914.
  • 5Laub WU, Wong T. The volume effect of detectors in the dosimetry of small fields used in IMRT. Med Phys, 2003, 30(3) : 341-347.
  • 6Leybovich LB, Sethi A, Dogan N. Comparison of ionization chambers of various volumes for IMRT absolute dose verification. Med Phys, 2003,30(2) : 119-123.
  • 7Klein EE,Chin LM,Rice RK,et al. The infuluence of the air cavities on interface doses for photon beams [J]. Int J Radiat Oncol Biol Phys, 1993,27(2) :419-427.
  • 8Mitine C,Radiother Oncol,1999年,52卷,19页
  • 9戴建荣,肿瘤放射物理学,1999年,463页
  • 10Bijhold J,Radiother Oncol,1992年,24卷,261页

共引文献59

同被引文献60

引证文献7

二级引证文献37

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部