同种异体骨圈椎体融合术稳定颈椎的生物力学评价 被引量:14

Biomechanical Evaluation on Cervical spine stabilizing Effects of Cervical Interbody Spinal Fusion Using Allograft Fusion Cage
原文传递
导出
摘要 目的 评价同种异体皮质骨圈(AFC)重建颈椎稳定性的即刻效果及自身强度。方法在8 具新鲜成人颈椎标本上,模拟临床术式,对C4,5 、C5,6 行椎间盘切除术,分别将髂骨、AFC 置于椎间隙,对节段进行压缩、植入物拔出测试(C4,5) 和脊柱三维运动稳定性评价(C5,6)。结果 (1)AFC组的运动范围与正常组和髂骨组相比,除后伸运动范围略有增大外,其余方向的运动范围均减小。(2) 在压缩载荷为(502 ±114)N 时,AFC组的椎骨有挤压破坏,而AFC结构正常;在300 N 的拔力作用下,骨圈与椎体间无松动。而在髂骨组中,当压力为(135 ±42)N 时,植骨块被破坏;在60 N 的拔力作用下,植骨块与椎体间产生松动移位。结论 AFC 具有足够的支撑、抗滑、维持或增加椎间隙高度的功能,符合生物力学及临床要求。 Objective\ To evaluate the immediate effects and strength of allograft fusion cage(AFC) used for reconstructing stability of cervical spine. Methods\ Discs of C 5 and C 6 were resected on 8 fresh human cervical spine specimens, and autogenous iliac bone grafts(AIBG), and AFCs were implanted into the intervertebral spaces respectively. Compression test, pull out test and segmental motion measurement were studied.Results\ Comparing with intact and AIBG groups, the range of motions of C5-6 in AFC group were decreased in all directions except for extension; with (502±114) N compressive load,the vertebrae in AFC group were broken while the AFC were intact, but the AIBG were broken at (135±42) N load; with 300 N drawing load,no loosening was found between AFCs and vertebra, but it was found between AIBG and vertebra at 60 N load. Conclusion\ AFC could provide enough support,anti slide ability and could remain or increase the height of intervertebral spaces. It completely meets clinical and biomechanical requirements.
出处 《中华骨科杂志》 CAS CSCD 北大核心 1999年第9期522-525,共4页 Chinese Journal of Orthopaedics
基金 1998 年广东省重点攻关基金
关键词 颈椎 脊柱融合术 同种移植 生物力学 Cervical vertebrae Spinal fusion Transplantation,homologous Biomechanics
  • 相关文献

参考文献3

  • 1赵定麟 陈德玉 等.颈椎前路减压术后各种内固定物的临床对比观察.第五届全国脊柱外科学术会议汇编[M].,1997.38-39.
  • 2赵定麟,天津市第五届全国脊柱外科学术会议,1997年,38页
  • 3贾连顺,现代颈椎外科学,1993年,21页

同被引文献54

引证文献14

二级引证文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部