Pencatam眼前节分析仪测量中央角膜厚度的研究

下载PDF
导出
摘要 目的:评价Pencatam眼前节分析仪测量中央角膜厚度的准确性。方法:分别使用Pencatam眼前节分析仪和A型超声测厚仪测量50例(100眼)眼科门诊患者的中央角膜厚度,验证两种方法测量结果的一致性。结果:分别采Pencatam眼前节分析仪及A型超声测厚仪测量的中央角膜厚度为543.05±30.65um和539.46±30.43um;比较差异无统计学意义(P>0.05)。结论:Pencatam眼前节分析仪测量中央角膜厚度结果准确,具有较好的应用价值。
作者 黄鹤龄
出处 《内蒙古中医药》 2011年第12期111-112,共2页 Inner Mongolia Journal of Traditional Chinese Medicine
  • 相关文献

参考文献4

  • 1Iskander NG,Anderson Penno E,Peters NT,et al.Accuracy of Orbscan pachymetry measurement and DHG ultrasound pachymetry in primary laser in situ keratomileusis and LASIK enhancement procedures[J].J Cataract Refract Surg,2001,27(5):681-685.
  • 2李坤,陆蓓,崇晓霞,陈彦玲,李晓玲.麻醉剂和泪膜对健康人角膜厚度的影响[J].眼视光学杂志,2004,6(4):244-246. 被引量:16
  • 3Ciolino J B,Khachikian S S,Belin M W.Comparison of corneal thickness measurements by ultrasound and seheimpflug photography in eyes that have undergone laser in situ keratomileusis[J].Am J Ophthalmol,2008,145 (1):75 -80.
  • 4De Sanetis U,Missolungi A,Mutani B,et al.Reprodueibility and repeatability of central corneal thickness measurements in keratoconus using the rotating seheimpflug camera and ultrasound pachymetry[J].Am J Ophthalmol,2007,144(5):712-718.

二级参考文献9

  • 1Yaylali V,Kaufman SC,Thompsom HW. Corneal thickness measurements with the Orbscan Topography System and Ultrasound Pachymetry[J]. J Cataract Refract Surg,1997,23(9):1345-1350.
  • 2李凤鸣.眼科全书(上册)[M].北京:人民卫生出版社,1997.42.
  • 3Iskander NG,Anderson Penno E,Peter NT,et al. Accuracy of Orbscan pchymetry measurements and DGH ultrasound pachymetry in primary laser in situ Keratomileusis and LASIK enhancement procedures[J]. J Cataract Surg,2001,27(5):681-685.
  • 4Nissen J,Hjortdal JO,Ehlers N,et al. A clinical comparison of optical and ultrasonic pachymetry[J]. Acta Ophthalmol Copenh,1991,69(5):659-663.
  • 5Lattimore MR Jr. Influence of extended soft contact lens wear on the comparative measurenent of central corneal thickness[J]. Acta Ophthalmol Scand,1996,74(2):239-242.
  • 6Mertz GW. Overnight swelling of the living human cornea[J]. J Am Optom Assoc,1980,51(2):211-214.
  • 7Lattimore MR Jr,Sandor Kaupp MS,Schallhorn S,et al. Orbscan pachymetry:implication of a reperted measures and diurnal variation analysis[J]. Ophthalmology,1999,106(5):977-981.
  • 8Dursun D,Monroy D,Knighton R,et al. The effects of experimental tear film removed on corneal surface regularity and barrier function[J]. Ophthalmology,2002,107(9):1754-1760.
  • 9谢立信,李赵霞,胡隆基,衣伟伟.Orbscan-Ⅱ在准分子激光原位角膜磨镶术术前检查中的应用[J].眼视光学杂志,2001,3(3):133-136. 被引量:13

共引文献15

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部