正常人中央角膜厚度对不同眼压计测量值的影响 被引量:2

Effect of Central Corneal Thickness on Intraocular Pressure Measured by Three Tonometers in Normal Subjects
下载PDF
导出
摘要 目的探讨正常人中央角膜厚度(central corneal thickness,CCT)对Icare回弹式眼压计(Icare rebound to-nometer,Icare)、Goldmann压平眼压计(Goldmann Applanation Tonometer,GAT)和PASCAL动态轮廓眼压计(DynamicContour Tonometry,DCT)眼压测量值的影响。方法采用Icare、GAT和DCT同时测量99例正常人共185眼的眼压,使用NIDEK UP-1000型角膜测厚仪测定CCT。采用重复测量数据的方差分析比较Icare、GAT和DCT眼压测量值之间的差异,眼压测量值的两两比较采用多个样本均数间的多重比较,眼压测量值的两两相关关系采用Pearson线性相关分析,CCT与3种眼压计眼压测量值的相关性采用Pearson线性相关及回归分析。结果 Icare、GAT和DCT测得的眼压平均值分别为(16.89±3.57)mmHg(1mmHg=0.133kPa)、(16.28±3.33)mmHg、(15.57±2.57)mmHg,3种眼压计的测量结果有统计学差异(F=8.042,P<0.001);Icare与GAT,Icare与DCT,GAT与DCT眼压测量值均呈正相关,相关系数(r)分别为0.879(P<0.001),0.583(P<0.001),0.678(P<0.001);Icare和GAT的眼压测量值与CCT均呈正相关,相关系数分别为0.478(P<0.001),0.443(P<0.001),而DCT的眼压测量值则与CCT无显著相关关系(r=0.100,P=0.175),Icare与DCT、GAT与DCT眼压测量值的差值与CCT有显著相关性,r均为0.493(P<0.001),而Icare与GAT眼压测量值差值与CCT无显著相关性,r为0.137(P=0.063)。结论正常人DCT眼压测量值不受CCT影响,Icare、GAT眼压测量值均受CCT影响,两者受CCT影响的差别不显著。 Objective To study the effect of central corneal thickness(CCT) on the intraocular pressure(IOP) measured by Icare rebound tonometer(Icare),Goldmann Applanation Tonometer(GAT) and Dynamic Contour Tonometry(DCT).Methods IOP of 99 normal persons(185 eyes) was measured using Icare,GAT and DCT.CCT was measured using NIDEK-UP-1000.ANOVA of repeated measurement data and multiple comparison was applied to compare the IOP values measured by Icare,GAT and DCT.The pair-wise correlation was evaluated by Pearson correlation analysis.The effect of CCT on IOP measured by Icare,GAT and DCT was assessed by Pearson correlation analysis and linear regression analysis.Results The mean IOPs measured by Icare,GAT and DCT were(16.89±3.57)mmHg(1mmHg=0.133kPa),(16.28±3.33)mmHg and(15.57±2.57)mmHg respectively(significantly different,F=8.042,P0.001).They had positive pair-wise correlations,the correlation coefficient(r) between Icare and GAT was 0.879(P0.001),r between Icare and DCT was 0.583(P0.001),and r between GAT and DCT was 0.678(P0.001).IOPs measured by Icare and GAT had positive correlation with CCT(r=0.478,P0.001;r=0.443,P0.001) and IOP measured by DCT had no significant correlation with CCT(r=0.100,P=0.175).The differences of measured IOPs between Icare and DCT,GAT and DCT were significantly correlated with CCT(both r=0.493,P0.001),and the difference of IOPs between Icare and GAT had no significant correlation with CCT(r=0.137,P=0.063).Conclusion In healthy eyes,CCT doesn't affect the IOP measurement by DCT,but has similar affection the IOP measurement by Icare and GAT.
出处 《华南国防医学杂志》 CAS 2011年第3期213-216,223,共5页 Military Medical Journal of South China
关键词 眼压 中央角膜厚度 眼压计 Intraocular pressure Central corneal thickness Tonometer
  • 相关文献

参考文献14

  • 1张绍阳,周和政,江文珊,张文强,陈云辉,袁敏.中央角膜厚度校正前后三种眼压计眼压测量值的一致性分析[J].华南国防医学杂志,2010,24(2):100-103. 被引量:2
  • 2Poostchi A,Mitchell R,Nicholas S,et al.The iCare rebound tonometer:comparisons with Goldmann tonometry,and influence of central corneal thickness[J].Clin Exp Ophthalmol,2009,37(7):687-691.
  • 3Herdener S,Hafizovic D,Pache M,et al.Is the PASCAL-Tonometer suitable for measuring intraocular pressure in clinical routine? Long-and short-term reproducibility of dynamic contour tonometry[J].Eur J Ophthalmol,2008,18(1):39-43.
  • 4曾阳发,刘杏,黄晶晶,蔡小于,钟毅敏,毛真,王涛,肖辉.中央角膜厚度对Goldmann眼压计与非接触式眼压计测量值的影响[J].中国实用眼科杂志,2007,25(4):393-396. 被引量:8
  • 5徐鼎,董方田,楚洁,戴荣平,金玉梅.不同角膜厚度下动态轮廓眼压计和Goldmann压平眼压计在不同角膜厚度的测量值比较[J].眼科新进展,2007,27(5):363-365. 被引量:10
  • 6Shimmyo M,Ross AJ,Moy A,et al.Intraocular pressure,Goldmann applanation tension,corneal thickness,and corneal curvature in Caucasians,Asians,Hispanics,and African-Americans[J].Am J Ophthalmol 2003,136:603-613.
  • 7Marjanovi c' I,Konti c' D,Hentova-Sen c'ani c' P,et al.Correlation between central corneal thickness and intraocular pressure in various age groups[J].Srp Arh Celok Lek,2010,138(5-6):279-286.
  • 8Bayer A,Sahin A,Hürmeri V,et al.Intraocular pressure values obtained by ocular response analyzer,dynamic contour tonometry,and goldmann tonometry in keratokonic corneas[J].J Glaucoma,2010,19(8):540-545.
  • 9Asrani S,Chatterjee A,Wallace DK,et al.Evaluation of the ICare Rebound Tonometer as a Home Intraocular Pressure Monitoring Device[J].J Glaucoma,2011,20(2):74-79.
  • 10Nakamura M,Darhad U,Tatsumi Y,et al.Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers[J].Am J Ophthalmol,2006,142(2):332-334.

二级参考文献47

  • 1鲍捷,代艳,陈晓明.角膜厚度对两种眼压测量方法的影响[J].中国实用眼科杂志,2005,23(3):272-274. 被引量:15
  • 2刘杏,曾阳发,黄晶晶,郑小萍,凌运兰,蔡小于.相干光断层扫描仪检测正常人及青光眼患者中央角膜厚度[J].中华眼科杂志,2006,42(3):199-203. 被引量:37
  • 3陈卉.Bland-Altman分析在临床测量方法一致性评价中的应用[J].中国卫生统计,2007,24(3):308-309. 被引量:217
  • 4Avetisov SE,Petrov SIu,Bubnova IA,et al.Impact of the central thickness of the cornea on the results of tonometry (a review of literature)[J].Vestn Oftalmol,2008,124(5):3-7.
  • 5Harada H,Hirose N,Kubota T,et al.The influence of central corneal thickness and corneal curvature radius on the intraocular pressure as measured with different tonometers:noncontact and goldmann applanation tonometers[J].J Glaucoma,2008,17(8):619-625.
  • 6Kotecha A,White ET,Shewry JM,et al.The relative effects of corneal thickniss and age on Glodmann applanation tonometer and dynamic contour tonometry[J].Br J Ophthalmol,2005,89(12):1572-1575.
  • 7Kaufmann C,Bachmann LM,Thiel MA.Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis[J].Invest Ophthalmol Vis Sci,2003,44(9):3790-3794.
  • 8Jóhannesson G,Hallberg P,Eklund A,et al.Pascal,ICare and Goldmann applanation tonometry--a comparative study[J].Acta Ophthalmol,2008,86(6):614-21.
  • 9Pakrou N,Gray T,Mills R,et al.Clinical comparison of the Icare tonometer and Goldmann applanation tonometry[J].J Glaucoma,2008,17(1):43-47.
  • 10孟海林,冯梅艳,杜献芳,张东魁.LASIK术后眼压测量值的变化分析[J].眼科新进展,2007,27(8):616-618. 被引量:8

共引文献16

同被引文献46

  • 1鲍捷,陈晓明.角膜厚度对两种眼压测量方法的影响[J].华西医学,2005,20(1):4-5. 被引量:2
  • 2刘杏,曾阳发,黄晶晶,郑小萍,凌运兰,蔡小于.相干光断层扫描仪检测正常人及青光眼患者中央角膜厚度[J].中华眼科杂志,2006,42(3):199-203. 被引量:37
  • 3吴敏,叶剑,孙强,翟军印,曹书,喻箭.前节OCT与A超测量前房深度的比较[J].国际眼科杂志,2007,7(5):1341-1342. 被引量:8
  • 4Salim S. The role of anterior segment optical coherence tomography in glaucoma. J Ophthalmol 2012;2012:476801.
  • 5Izatt JA, Hee MR, Swanson EA, et al . Micrometer-scale resolution imaging of the anterior eye in vivo with optical coherence tomography. Arch Ophthalmol 1994 ; 112 ( 12 ) : 1584 - 1589.
  • 6Ehlers N, Hansen FK. Central corneal thickness in low- tension glaucoma. Acta Ophthalmol(Coperth) 1974 ;52 (5) :740-746.
  • 7Fern KD, Manny RE, Gwiazda J,et al. Intraoeular pressure and central corneal thickness in the COMET cohort. Optom Vis Sei 2012 ; 89 ( 8 ) : 1225-1234.
  • 8Fishman GR, Pons ME, Seedor JA,et al. Assessment of central corneal thickness using optical coherence tomography. J Cataract Refract Surg 2005 ;31 (4) :707-711.
  • 9Dada T, Sihota R, Gadia R, et al. Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy for assessment of the anterior segment. J Cataract Refract Surg 2007;33 (5) :837-840.
  • 10Garcia - Medina JJ, Garcia - Medina M, Garcia - Maturana C, et al . Comparative study of central corneal thickness using fourier- domain optical coherence tomography versus ultrasound pachymetry in primary open-angle glaucoma. Cornea 2013 ;32( 1 ) :9-13.

引证文献2

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部