呼吸过滤器对常规肺功能检测的影响观察

下载PDF
导出
摘要 目的探讨使用呼吸过滤器进行肺功能检查时对气道阻力及其它肺功能指标的影响。方法选择健康战士50名,每个受检者分别不使用和使用呼吸过滤器,进行常规肺功能检测,测定以下指标:FVC、SVC、FEV1、FEFMax、FEF25%、FEF50%、FEF25%~75%、FEF75%、MVV和气道阻力(Raw)。比较两种方法的差异。结果使用呼吸过滤器进行肺功能检测时气道阻力明显大于不使用呼吸过滤器,两者比较有显著差异性(P<0.05)。其他肺功能指标两种方法比较结果无显著性差异。结论使用呼吸过滤器增大肺功能测试时受检者的气道阻力,但对其他肺功能指标无明显影响。
出处 《临床肺科杂志》 2011年第7期1134-1135,共2页 Journal of Clinical Pulmonary Medicine
  • 相关文献

参考文献5

  • 1高怡,郑劲平.呼吸过滤器在肺功能实验室中的应用[J].国际呼吸杂志,2006,26(11):825-829. 被引量:9
  • 2Kendrick AH,Johns DP,Leeming JP.Infection control of lung function equipment:a practical approach[J].Respir Med,2003,97(11):1163-1179.
  • 3Miller MR,Crapo R,Hankinson J,et al.ATS/ ERS task force.General considerations for lung function testing[J].Eur Respir J,2005,26(1):153-161.
  • 4郑劲平.肺功能学[M].广东:广东科技出版社,2007:61.
  • 5Coates AL,Desmond KJ,Demizio D,et al.Sources of variationin FEV1[J].Am J Respir Crit Care Med,1994,149(2 pt 1):439-443.

二级参考文献18

  • 1Kamps AWA, Vermeer K, Roorda RJ, et al. Effect of bacterial filters on spirometry measurements. Arch Dis Child, 2001,85(4) :346-347.
  • 2Kendrick AH,Johns DP, Leeming JP. Infection control of lung function equipment: a practical approach. Respir Med, 2003,97(11) :1163-1179.
  • 3Demers RR, BS, RRT. Bacterial/ virual Filtration. Let the Breather Beware! Chest,2001,120(4) :1377-1389.
  • 4Townsend MC, Hankinson JL, Lindesmith LA, et al. Is My Lung Function Really That Good? Flow-Type Spirometer Problems That Elevate Test Results. Chest, 2004,125 (5) :1902-1909.
  • 5Dellamonica J, Boisseau N, Gonbaux B, et al. Comparison of manufactures specifications for 44types of heat and moisture exchanging filters. BJA, 2004,93(4) : 532-539.
  • 6Castel O, Planchon C, Denjean A, et al. Assessment of three filters for respiratory function tests. Rev Mal Respir, 1998, 15(6) :759-760.
  • 7Canakis AM, Ho B, Ho S, et al. Do in-line respiratory filters protect patients? Comparing bacterial removal efficiency of six filters. Pediatr Pulmonol,2002,34(5) :336-341.
  • 8Wilkes AR, Benough JE, Speight SE, et al. The bacterial and virus filtration performance of breathing system filters. Anaesthesia, 2000,55 (5):458-465.
  • 9Atkinson MC,Girgis Y, Broome IJ. Extent and practicalities of filter use in anaesthetic breathing circuits and attitudes towards their use : a postal survey of UK hospitals. Anaesthesia, 1999,54(1) :37-41.
  • 10British Standards Institution. Breathing system filters for anaesthetic and respiratory use. Salt test method to assess filtration performance (BS EN 13382-1,2001 ). London : British Standards Institution, 2001.

共引文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部