^60Coγ射线空气比释动能量值国内比对 被引量:1

Comparison of Air Kerma in 60Co Gamma Rays
下载PDF
导出
摘要 中国计量科学研究院作为主导实验室,组织了全国15个实验室参加的^60Coγ射线空气比释动能量值比对。比对结果表明各参比实验室测量的比对点处空气比释动能率在0.6%以内吻合,核验比对给出的传递标准空气比释动能校准因子在0.8%以内吻合。 As the pilot laboratory, National Institute of Metrology organized the comparison for air kerma in 60Co gamma rays in China. There are 15 laboratories join the comparison. The results show that the comparison difference of air kerma rate in reference point is within 0. 6% , and the comparison difference of calibration factor for air kerma of transfer chambers is within 0. 8%.
出处 《计量学报》 CSCD 北大核心 2011年第4期377-380,共4页 Acta Metrologica Sinica
关键词 计量学 空气比释动能 ^60Coγ射线 比对 Metrology Air kerma 60 Co gamrrla ray Comparison
  • 相关文献

参考文献6

  • 1Allisy-Roberts P J, Kessler C, Burns D T , et al. Comparison of the standards for air kerma of the NIM and the BIPM for 60Co gamma rays [ J ]. Metrologia, 2009,46,06013.
  • 2JJG912-2010,治疗水平电离室剂量计检定规程[s].
  • 3JJF1117-2010,测量仪器比对规范[S].
  • 4JJG1059-1999.测量不确定度评定与表示[S].[S].,..
  • 5International Atomic Energy Agency. Absorbed Dose Determination in Photon and Electron Beams-An International Code of Practice [ R ]. IAEA Technical Report No. 277,1987.
  • 6AAPM TG--21. A protocol for the determination of absorbed dose from high-energy photon and electron beams[J]. Medical Physics, 1983,10(6) :741 - 771.

共引文献17

同被引文献8

  • 1IAEA. Absorbed Dose Determination in High Energy Electron and Photon Beams[R].IAEA Technical Report Series No277,1987.
  • 2IAEA. Absorbed Dose Determination in External Beam Radiotherapy[R].IAEA Technical Reports Series No398,2000.
  • 3Ratel G, Michotte C. BIPM comparison BIPM.RI(11)-K1 60Co of the activity measurements of the radionuclide 60Co[J].Metrologia,2003,40(1A):06007.
  • 4Allisy-roberts P J, Burns D T. Summary of the BIPM.RI(I)-K4 comparison for absorbed dose to water in 60Co gamma radiation[J].Metrologia,2005,42(1A):06002.
  • 5The BIPM key comparison database [DB].http://kcdb.bipm.org, 2015-11-01.
  • 6IAEA. IAEA/WHO network of secondary standard dosimetry laboratories comparison of therapy-level ionization chamber calibration coefficients[R].CMP-CPR/2014/01,2014.
  • 7王坤,李涛,杨小元.^(60)Co水吸收剂量的量值复现和国际比对验证[J].中国医学物理学杂志,2013,30(6):4476-4479. 被引量:12
  • 8李涛,王坤,杨小元,樊铁栓.基于石墨圆饼电离室的^(60)Co γ射线空气比释动能的绝对测量研究[J].原子能科学技术,2015,49(7):1153-1158. 被引量:1

引证文献1

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部