期刊文献+

根管内血液对根尖定位仪Raypex5测量准确性的影响 被引量:10

Effect of blood in root canal on measurement accuracy of apex locator Raypex5
下载PDF
导出
摘要 目的通过体外实验评价根管内血液对根尖定位仪Raypex5测量牙齿工作长度准确性的影响。方法选择单根管离体牙30颗,用直接测量法取得根管实际长度。用生理盐水琼脂模型作为体外测试模型,用Raypex5分别在根管干燥、内有部分血液、充满血液3种情况下测量根管工作长度。结果直接测量法与电测法数据进行比较,测量值间差异无统计学意义(P>0.05)。在根管干燥、内有部分血液、充满血液3种情况下,Raypex5测量值间的差异无统计学意义(P>0.05)。结论 Raypex5测量根管工作长度的准确性不受根管内血液的影响,测量所得数值可作为根管工作长度参考值。 Objective To evaluate the influence of blood in root canal on measurement accuracy of apex locator Raypex5 in vitro.Methods Tirty extracted single root canal teeth were included in this study.The actual length of the root canals were obtained by direct manual measurement.The working length of the root canal was measured by Raypex5 in conditions of dry root canal,root canal filled partly with blood and with blood,respectively.Results There was no significant difference in root canal length between Raypex5 and direct manual measurement(P 0.05).Raypex5 measurement showed no significant difference among the three conditions(P0.05).Conclusion Blood in the root canal does not affect the measurement accuracy of Raypex5.
出处 《北京口腔医学》 CAS 2011年第3期147-148,共2页 Beijing Journal of Stomatology
关键词 根尖定位仪 血液 根管工作长度 Apex locator Blood Root canal working length
  • 相关文献

参考文献8

  • 1Moshonov J,Slutzky-Goldberg I.Apex locators:update and prospects for the future.Int J Comput Dent,2004,7(4):359-370.
  • 2Donnelly JC.A simplified model to demonstrate the operation of electronic root canal measuring device.J Endod,1993,19(11):579-580.
  • 3Herrera M,Abalos C,Planas AJ,et al.Influence of apical constriction diameter on Root ZX apex locator precision.J Endod,2007,33(8):995-998.
  • 4朱习华,董红,龙明生.ROOT ZX根管测量仪测量根管长度的临床分析[J].北京口腔医学,2006,14(1):63-63. 被引量:7
  • 5Czerw RJ,Fulkerson MS,Donnelly JC.An in vitro test of a simplified model to demonstrate the operation of electronic root canal measuring devices.J Endod,1994,20(12):605-606.
  • 6李政权,李俊,高原.根尖定位仪Raypex~ 5测量准确性实验研究[J].口腔医学研究,2008,24(6):695-696. 被引量:6
  • 7何俐,胡图强,赵红,林建琴,李鹏.4种液体对电子根尖定位仪测量准确性的影响[J].上海口腔医学,2008,17(3):304-307. 被引量:5
  • 8Ebrahim AK,Yoshioka T,Kobayashi C,et al.The effects of file size,sodium hypochlorite and blood on the accuracy of Root ZX apex locator in enlarged root canals:an in vitro study.Aust Dent J,2006,51(2):153-157.