期刊文献+

基于动态阈值的肌电假手动作控制方法研究 被引量:4

Movement Pattern Control for Prosthetic Hand Based on a Method of Dynamic Threshold
下载PDF
导出
摘要 由于构成肌电信号采集电路的电子元器件性能不可能完全对称及干扰信号的存在,有时会导致两路肌电信号发生阈值不一致。这时仍采用固定阈值来对两路肌电信号控制的动作进行判别,会导致动作的误判率增加。为了提高对假手动作判别的正确率,本文提出了利用动态阈值对假手动作进行判别。实验结果表明,利用动态阈值对假手动作进行判别,能够提高对动作判别的正确率达约10%。 Due to the function asymmetry of electrical components used for EMG signal acquisition circuit and the existing of undesired signals,sometimes thresholds for judging EMG's occur may be incongruent.If a fixed threshold is used to recognize Prosthetic Hand's movement,the matching rate of prosthetic hand's movement pattern recognition will be increased.In order to improve prosthetic hand's movement pattern recognition,a method of dynamic threshold was used to recognize the prosthetic hand's movement.The exprimental results showed that the method of dynamic threshold can obviously improve Prosthetic Hand's movement pattern recognition by about 10%.
出处 《现代科学仪器》 2011年第3期43-45,共3页 Modern Scientific Instruments
关键词 假手 肌电信号 动作识别 Artifical hand Electromyography Movement recognition
  • 相关文献

参考文献5

二级参考文献15

  • 1姜明文,王人成,罗志增,金德闻,张济川.具有触滑觉功能的肌电假手[J].清华大学学报(自然科学版),2004,44(8):1051-1053. 被引量:22
  • 2SARIDIS G. EMG pattern analysis and classification for a prothetic arm [ J]. IEEE Trans. Biomed. Eng. , 1982,29 (6) :403 -412.
  • 3HERBERTS P, ALMSTROM C, CAINE K. Clinical application study of multifunctional prosthetics [ J ]. J. Bone Joint Surg. Br. ,1978,60(4) :552 -560.
  • 4KATSUTOSHI K, SHIMIZU S, OKIMURA K, et al. A discrimination system using neural network far EMG-controlled prostheses-integral type of EMG signal [ C ]//Processing. Proc. of the1993 IEEE/RSJ Int. Conference on Intelligent Robots and Systems' 93 [ Yokohama ] , 1983 : 1750 - 1755.
  • 5侯文生,许蓉,郑小林,马丽.握力大小与前臂肌肉表面肌电活动模式的相关性研究[J].航天医学与医学工程,2007,20(4):264-268. 被引量:25
  • 6查理.肌电假手的研究进展[J].国防科技,2007,28(9):6-13. 被引量:14
  • 7Vaiman M,Segal S,Eviatar E.Surface electromyographic studies of swallowing in normal children,age 4-12 years[J].International Journal of Pediatric Otorhinolaryngology,2004,68:65-73.
  • 8Sbriccoli P,Bazzucchi I,Rosponi A,et al.Amplitude and spectral characteristics of biceps Brachii sEMG depend upon speed of isometric force generation[J].Journal of Electromyography and Kinesiology,2003,13:139-147.
  • 9Choi K,Hirose H,Iijima T,et al.Prediction of four degrees of freedom arm movement using EMG signal[A].Proceedings of the 2005 IEEE.Engineering in Medicine and Biology 27th Annual Conference[C].Shanghai,2005.1018.
  • 10Hug F,Decherchi P,Marqueste T,et al.EMG versus oxygen uptake during cycling exercise in trained and untrained subjects[J].Journal of Electromyography and Kinesiology,2004,14:187-195.

共引文献6

同被引文献29

引证文献4

二级引证文献44

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...