两种清洗方法对牙科手机使用寿命的影响分析 被引量:1

Evaluation of two cleaning methods for the service life of dental handpieces
下载PDF
导出
摘要 目的比较两种清洗方法对牙科手机使用寿命的影响。方法将20支牙科手机使用后随机分成甲乙两组各10支,甲组手机人工清洗后自然晾干、注油、灭菌;乙组手机人工清洗后置于全自动手机热力清洗消毒机内清洗、烘干、注油、灭菌。结果两组牙科手机使用清洗2000次后,检测甲组手机转速为(10.49±0.30)万转/m in、乙组手机为(27.51±2.90)万转/m in,差异有统计学意义(P<0.01);甲组手机扭矩为(0.41±0.08)kg、乙组手机为(1.53±0.13)kg,差异有统计学意义(P<0.01)。结论人工+机械清洗对牙科手机转速和扭矩的影响明显较单纯的人工清洗方法减低,采用人工+机械清洗的方法可在一定程度上降低手机损耗,延长其使用寿命。 Objective To compare two cleaning methods on the service life of dental handpieces.Methods 20 used handpieces were randomly divided into two groups,each 10 respectively.Group A handpieces dried naturally after taking artificially cleaning,oiled,sterilized;Group B were placed in automatic heat washing and disinfection machine after taking artificially cleaning,dried,oiled and sterilized.Results All dental handpieces were cleaned for 2000 times,the revolving speed of Group A and B handpieces was(10.49±0.30) million rev/min and(27.51±2.90) million rev/min,respectively,the difference was statistically significant(P0.01);the torque of Group A and B handpieces was(0.41±0.08) kg and(1.53±0.13) kg,respectively,the difference was also statistically significant(P0.01).Conclusion There're less impaction on the revolving speed and torque of dental handpieces using artificially plus mechanically cleaning method than simply manual cleaning.The former cleaning method could reduce the handpiece's wastage to some extent and extend its service life.
作者 王群 庄根鹰
出处 《中外医学研究》 2011年第19期15-16,共2页 CHINESE AND FOREIGN MEDICAL RESEARCH
关键词 牙科手机 清洗 全自动手机热力清洗消毒机 使用寿命 Dental handpiece Cleaning Automatic mobile heat cleaning and disinfecting machine Service life
  • 相关文献

参考文献4

二级参考文献18

  • 1李淑霞,邵泽注,周晓.临沂市口腔医疗机构消毒工作质量调查[J].中华医院感染学杂志,2005,15(3):302-304. 被引量:11
  • 2邓小虹,孙正,乔宏,邓宏燕,肖潇,苏静.牙科手机传播乙型肝炎病毒的可能性探讨[J].中华预防医学杂志,2005,39(3):199-202. 被引量:16
  • 3莫凡,杨雪英,王君,谭开威,唐松桥,李位民,甘少平.口腔科门诊医院感染管理现状[J].中华医院感染学杂志,2006,16(3):317-319. 被引量:21
  • 4王华生,孙雪莹,梁树森,徐然.影响医疗器械清洗质量的因素[J].中华医院感染学杂志,2007,17(5):553-554. 被引量:116
  • 5Andersen HK, Fiehn NE, Larsen T. Effect of steam sterilization inside the turbine chambers of dental turbines. Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 1999,87 (2) : 184-188.
  • 6AngeliUo IF, Bianco A, Nobile CG, et al. Evaluation of the efficacy of glutaraldehyde and peroxygen for disinfection of dental instruments. Lett Appl Microbiol, 1998,27 (5) :292-296.
  • 7Aasim SA, MeUor AC, Qualtrough AJ. The effect of pre-soaking and time in the ultrasonic cleaner on the cleanliness of sterilized endodontic files. Int Endod J,2006,39 (2) :143-149.
  • 8Linsuwanont P, Parashos P, Messer HH. Cleaning of rotary nickeltitanium endodontic instruments. Int Endod J,2004,37( 1 ) :19-28.
  • 9Coulter WA, Chew-Graham CA, Cheung SW, et al. Autoclave performance and operator knowledge of autoclave use in primary care : a survey of UK practices. J Hosp Infect,2001,48 (3) :180-185.
  • 10Bagg J, Smith AJ, Huh'ell D, et al. Pre-sterilisation cleaning of reusable instruments in general dental practice. Br Dent J, 2007,202 (9) :550-551.

共引文献20

同被引文献2

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部