期刊文献+

CZT像素阵列核探测器的噪声分析 被引量:1

Noise Analysis of Array Pixellated CZT Detector
下载PDF
导出
摘要 碲锌镉(CdZnT e)晶体是一种新型核辐射探测材料,在室温探测环境下对X射线及低能量伽玛射线具有较高的探测效率及能量分辨率。运用低噪声快速前放模块,分析晶体内部性能以及前置放大器对探测系统噪声的影响,建立了2×2碲锌镉像素阵列探测系统。实验结果表明,探测系统输出信号噪声小、脉冲无堆积,读出电路电子学噪声得到明显抑制,能谱低能拖尾减少,电荷收集不完全噪声及漏电流噪声影响得到改善,同时对整个系统以及前置放大器进行了噪声分析并提出相关的改善方式,最后得到系统测试的结果。 As a novel nuclear radiation material,CdZnTe has many advantages of high efficiency detection and good energy resolution for detection of x-ray and low energy gamma-ray in room temperature.The inner crystal property and effects of preamplifier on the noise is analyzed using low noise preamplifier module,while a 2×2 array pixellated detection system is built.The results show a low noise of output signal and no pulse stack.The circuit electronic noise is obviously restrained.Trails of low power of energy spectrum are decreased.Noise analysis and the factors which produced the noise are given;improvement for the system is also discussed,together with the test result.
出处 《光电子技术》 CAS 北大核心 2011年第2期98-102,共5页 Optoelectronic Technology
基金 国家自然科学基金(NSAF基金)资助项目(10876044)
关键词 噪声 前置放大器 碲锌镉 noise preamplifier CdZnTe
  • 相关文献

参考文献1

二级参考文献6

共引文献9

同被引文献18

  • 1曾国强,葛良全,赖万昌.手持式辐射仪高压直流电源的设计[J].电测与仪表,2008,45(6):46-49. 被引量:10
  • 2肖无云,魏义祥,艾宪芸.多道脉冲幅度分析中的数字基线估计方法[J].核电子学与探测技术,2005,25(6):601-604. 被引量:16
  • 3徐慧超,张金洲,沈浩元.用于X荧光分析的半导体探测器[J].常熟理工学院学报,2006,20(2):91-93. 被引量:3
  • 4Cremat Inc. CR-150-R5 evaluation board application guide[OL], http://www.cremat.com/CR- 150-BOX-R5.pdf, 2014.
  • 5Spieler H. Introduction to radiation detectors and electronics[R]. USA: Lawrence Berkeley National Laboratory, 1998.
  • 6Noulis T, Fikos G, Siskos S, et aL Noise analysis of radiation detector charge sensitive amplifier architectures[C]. Topical Workshop on Electronics for Particle Physics, 2008:486-489.
  • 7Femandes L M P, Simoes P C P S, dos Santos J M F. Digital rise-time discrimination of large-area avalanche photodiode signals in X-ray detection[J]. IEEE Transactions on Nuclear Science, 2002, 49(4): 1699-1703 DOI: 10.1109/TNS.2002.801518.
  • 8Toke J, Quinlan M J, Gawlikowicz W. A simple method for rise-time discrimination of slow pulses from charge-sensitive preamplifiers[J]. Nuclear Instruments and Methods in Physics Research, 2008, A595: 460-463. DOI: 10.1016/j.nima.2008.07024.
  • 9Niemela A, Sipila H, Ivanov V I. Improving CdZnTe X-ray detector performance by cooling and rise time discrimination[J]. Nuclear Instruments and Methods in Physics Research, 1996, A377(2): 484-486. DOI: 10.1016/0168-9002(96)00028-9.
  • 10Jordanov V T, Pantazis J A, Huber A C. Compact circuit for pulse rise-time discrimination[J]. Nuclear Instruments and Methods in Physics Research, 1996, A380: 353-357. DOI: 10.1016/0168-583X(94)00570-2.