ATCM技术不同调制模式胸部CT扫描剂量的实验研究 被引量:3

Experimental study of radiation dose in chest CT scanning with different modulation patterns of ATCM technique
下载PDF
导出
摘要 目的通过仿真胸部体模CT扫描模拟实验,评价ATCM技术3种调制模式胸部扫描辐射剂量的特点。方法使用Philips公司Brilliance 16层螺旋CT,分别采用ATCM技术的角度调制、Z轴调制和联合调制模式,预置2组管电流(122 mA、296 mA),其他技术参数相同,扫描中国成人男性CDP-IC型仿真胸部体模。存储所有图像并记录CTDI和DLP值,以联合调制作为对照组进行配对资料t检验。结果角度调制和Z轴调制以其技术特点按照仿真胸部体模各层组织结构的变化调制管电流。预置296 mA组,角度调制、Z轴调制与联合调制比较,CTDI和DLP值比较差异有统计学意义(P<0.05)。结论 ATCM技术的3种模式均可不同程度地降低胸部CT扫描辐射剂量,使用联合调制模式降低辐射剂量更为显著。 Objective To evaluate the characteristic of radiation dose in chest CT scanning with different patterns of automatic tube current modulation(ATCM) technique.Methods The Chinese anthropomorphic chest phantom(CDP-IC)was scanned by Brilliance 16(Philips)with three patterns(angular modulation,z-axis modulation and combined modulation)of ATCM technique using two pre-set mA(122mA,296mA)and same others scan parameter.CT dose index(CTDI) and dose-length product(DLP)values were recorded.All data were analyzed by a paired t test.Results Angular modulation and z-axis modulation involves variation of the radiation dose among anatomic regions of chest anthropomorphic phantom by varying the tube current in accordance with its technique characteristic.In group of pre-set 296mA,compared with angular modulation and z-axis modulation,CTDI and DLP value had significant statistically differences(P0.05).Conclusion Three patterns of ATCM technique can decrease radiation dose in chest CT scanning,the radiation dose reduction of combined modulation technique is more significant than the others.
出处 《重庆医学》 CAS CSCD 北大核心 2011年第20期1979-1981,共3页 Chongqing medicine
基金 重庆市卫生局科研基金资助项目(08-2-29)
关键词 辐射剂量 自动管电流调制 仿真胸部体模 radiation dosage automatic tube current modulation anthropomorphic chest phantom
  • 相关文献

参考文献16

  • 1McCollough CH, Bruesewitz MR, Kofler B, et al. CT dose reduction and dose management tools:overview of available options[J].Radiographics, 2006,26 (2) : 503- 509.
  • 2王远萍,林大全,付涛.组织等效材料的组成结构对辐射等效性影响的研究[J].生物医学工程学杂志,1997,14(1):42-49. 被引量:2
  • 3曾勇明,吴富荣,张志伟,罗天友,欧阳羽,林大全,刘志宏.CR入射体表剂量与肺部病变检出相关性的实验研究[J].重庆医学,2009,38(2):130-133. 被引量:4
  • 4Kopka L, Funke M, Breiter N, et al. An anatomically adapted variation of the tube current in CT:studies on radiation dosage reduction and image quality [J ]. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr, 1995, 163 (2) : 383-389.
  • 5Gies M, Kalender WA, Wolf H, et al. Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies[J]. Med Phys, 1999, 26 (6) : 2235 -2239.
  • 6Westerman BR. Radiation dose from Toshiba CT scanners [J]. Pediatr Radiol,2002,32(2) :735-740.
  • 7Peng Y, Li J, Ma D, et al. Use of automatic tube current modulation with a standardized noise index in young children undergoing chest computed tomography scans with 64-slice multidetector computed tomography[J]. Acta Ra- diol,2009,50(10) : 1175-1179.
  • 8Lehmann K J, Wild J, Georgi M. Clinical use of softwarecontrolled X-ray tube modulation with"Smart-Scan" in spiral CT[J].Aktuelle Radiol, 1997,7(1) : 156-161.
  • 9Mulkens TH, Bellinck P, Baeyaert M, et al. Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations:clinical evaluation [J].adiology. 2005.237(1 ) , 213-21 g.
  • 10Rizzo S,Kalra M,Schmidt B,et al. Comparison of angular and combined automatic tube current modulation techniques with constant tube current CT of the abdomen and pelvis[J]. AJR Am J Roentgenol,2006,186(3) :673-680.

二级参考文献9

  • 1曾勇明,陈亮,陆云峰.CR图像噪声的原因及抑制方法[J].中国医学影像技术,2004,20(z2):148-149. 被引量:9
  • 2马骏,朱晓华,孙希文,彭刚.结节病肺部改变的CT征象分析[J].中华放射学杂志,2006,40(9):923-928. 被引量:31
  • 3Sinead EP, Patrick CB. Digital radiography: are the manu facturers'settings too high? Optimisation of the Kodak digital radiography system with aid of the computed radiography dose index[J]. Eur Radiol,2002,12(12): 2381.
  • 4Chotas HG,Floyd CE,Dobbins JT,et al. Digital chest radiography with photostimulable storage phosphors: signal-to-noise ratio as a function of kilovoltage with matched exposure risk[J]. Radiology, 1993,186(2) : 395.
  • 5Tingberg A,Sjostrom D. Search for optimal tube voltage for image plate radiography, in:Chakraborty DP, Krupinski EA, ed. Proceedings of SPIE: medical imaging 2003- image perception, observer performance, and technology assessment. Vol : 5034 [M]. Bellingham, wash : Internation al Society for Optial Engineering,2003:187.
  • 6王玉庆,中国科技与企业杂志,1995年,6卷,35页
  • 7林大全,Ploymers and biomaterials,1991年,391页
  • 8安智珠,高分子材料设计的数学模拟,1989年
  • 9布朗 P R,塑料试验方法手册(译),1988年

共引文献4

同被引文献19

引证文献3

二级引证文献31

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部