期刊文献+

宽能谱高灵敏含硼塑料闪烁体探测器理论设计

Design of a high efficiency boron-loaded plastic scintillation detector for a wide spectrum of neutrons
下载PDF
导出
摘要 为提高含硼塑料闪烁体探测器对0.02–0.25 MeV能区中子的探测效率,提出了在含硼塑料闪烁体外包围聚乙烯慢化体的结构设计,并用MCNP程序模拟计算含硼塑料闪烁体探测器的探测效率,给出了探测效率与前端和侧面聚乙烯慢化体厚度的关系。结果表明,通过侧面聚乙烯慢化体的添加,可以大大提高含硼塑料闪烁体探测器对0.02–0.25 MeV能区中子的探测效率。当侧面聚乙烯厚度为80 mm时,对该能区中子的探测效率提高~35倍,而前端慢化体的添加会降低其探测效率。最后证明在含硼塑料闪烁体侧面添加聚乙烯慢化体的设计可以实现对宽能谱中子的高灵敏探测。 For improving detection efficiency of 0.02–0.25 MeV neutrons by boron-loaded plastic scintillation detector(BLPSD),surrounding the scintillator with polyethylene moderator is feasible way.In this paper,the detection efficiency of a BLPSD with front and side PE moderators of different thicknesses is simulated by MCNP code,and the thickness-dependant detection efficiencies are given.The results show the side PE moderator of 80 mm thick increases the BLPSD detection efficiency of 0.02–0.25 MeV neutrons by a factor of about,but the front PE moderator decreases the detection efficiency.Finally,the design structure of BLPSD surrounded by side PE moderator can realize high detection efficiency for a wide spectrum is proved.
出处 《核技术》 CAS CSCD 北大核心 2011年第8期599-603,共5页 Nuclear Techniques
关键词 含硼塑料闪烁体 蒙特卡罗 探测效率 慢化体 Boron-loaded plastic scintillator MCNP Detection efficiency Moderator
  • 相关文献

参考文献9

  • 1Waheed A, Aqil-ur-Rahman, Zakaullah M, et al. High efficiency neutron detector for low neutron flux measurement [J]. Journal of Fusion Energy, 2000, 19(1): 91-92.
  • 2胡青元,李波均,张德,郭洪生,王栋,杨高照,司粉妮,刘建.长中子计数管探测效率的模拟[J].原子能科学技术,2008,42(10):872-874. 被引量:3
  • 3李俊杰,郑春,邱东,代少丰.长计数器的特性研究及其在屏蔽性能测量中的应用[J].核动力工程,2007,28(z2):35-36. 被引量:3
  • 4Nina Hong, John Mullins, Keith Foreman, et al. Boron carbide-based solid state neutron detectors: The effects of bias and time constant on detection efficiency [J]. J Phys D: Appl Phys, 2010, 43(275101): 1-11.
  • 5BC-454 Boron-loaded premium plastic scintillator [J]. Bicron Company, 1999, 422(1-3): 89-94.
  • 6Briesmeister J F. MCNP-A General Monte Carlo N-particle Transport Code Version 4C [Z]. Los Alamos National Labortory, 2000.
  • 7Birks, Theory and practice of scintillation counting [M]. New York: Pergamon Press, 1964:194.
  • 8Stephane Normand, Brigitte Mouanda, Serge Haan, et al. Study of a new boron loaded plastic scintillator [J]. IEEE Transactions on Nuclear Science, 2002, 49(4): 1603-1608.
  • 9Douglas S. McGregor, Shultis J K. Reporting detection efficiency for semiconductor neutron detectors: A need for a standard [J]. Nucl Instru Methods Phys Res A, 2010, 632:167-174.

二级参考文献6

  • 1BEDDINGFIELD D H, JOHNSON N H, MENLOVE H O. ^3He neutron proportional counter performance in high gamma-ray dose environments[J]. Nucl Instrum Methods, 2000, A455: 670-682.
  • 2洪润生 蔡敦九 王大椿 等.长计数管效率的校准.原子能,1964,(10):926-932.
  • 3EAST L V, WALTON R B. Polyethylene moderated ^3He neutron detectors[J]. Nuel Instrum Methods, 1969, 72: 161-166.
  • 4HANSON A O, MCKIBBEN J L. A neutron detector having uniform sensitivity from 10 keV to 3 MeV[J]. Physical Review, 1947, 72: 673- 677.
  • 5SAMPSON T E, VINCENT D H. An absolute measurement of the efficiency of a BF3 proportional counter[J]. Nucl Instrum Methods, 1971, 95: 563-569.
  • 6FOWLER I L, TUNNICLIFFE P R. Boron trifluoride proportional counters [ J ]. Rev Sci Instrum, 1950, 21: 734-740.

共引文献4