ICARE回弹式眼压计在婴幼儿青光眼的临床应用 被引量:5

Clinical application of ICARE rebound tonometer in glaucomatous infants and young children
原文传递
导出
摘要 目的探讨ICARE回弹式眼压计在婴幼儿青光眼患者的适用性及准确性。方法青光眼患儿11例(22只眼),其中婴幼儿型青光眼8例(16只眼),青光眼伴其它眼部发育异常3例(6只眼),年龄2.7~19.8个月,平均7.3个月。使用ICARE回弹式眼压计及Schiotz压陷式眼压计进行眼压测量。配对t检验比较两种眼压计之间的差异。结果所有患儿都可在非用药状态下使用ICARE眼压计测量眼压,Schiotz眼压计须在患儿用药后深睡眠状态下测量眼压,共获得34对可靠数值。ICARE眼压计和Schiotz眼压计的测量值分别为(13.95±10.69)mmHg和(14.60±10.78)mmHg(平均值±标准差),二者间的差异无统计学意义(P〉0.05)。结论ICARE回弹式眼压计是适合于婴幼儿青光眼患者眼压检查的眼压计,其测量结果与Schiom压陷式眼压计测量结果相似。 Objective To investigate the applicability and accuracy of ICARE rebound tonometer in glaucomatous infants and young children. Methods Eleven cases (22 eyes) of glaucomatous children were involved, including 8 cases (16 eyes) of infantile glaucoma and 3 cases (6 eyes) of glaucoma accompanying other maldevelopment, aged 2.7-19.8 months old. ICARE tonometer and Schiotz tonometer were used to measure intraocular pressure (IOP) of the patients. The measurements were compared used pared student-t test. Results Thirty-four pairs of reliable data were obtained. The values of ICARE and Scbiotz tonorneter were 13.95+ 10.69mmHg and 14.60+ 10.78mmHg (average+ SD), respectively. The difference had no statistic signification (P 〉0.05). Conclusions ICARE Rebound tonometer is suitable to glaucomous infants and young children to measure lOP. The measurements are comparable with those of Schiotz tonometer.
出处 《中国实用眼科杂志》 CSCD 北大核心 2011年第8期785-788,共4页 Chinese Journal of Practical Ophthalmology
关键词 眼压 回弹式眼压计 婴幼儿 Intraocular pressure Rebound tonometer Infants and young children
  • 相关文献

参考文献17

  • 1Ruokonen PC,Schwenteck T,Draeger J. Evaluation of the impedance tonometers TGDc-01 and ICARE according to the international ocular tonometer standards ISO 8612 [J]. Graefes Arch Ciin Exp Ophthalmol, 2007,245 ( 9 ) : 1259 - 1265.
  • 2van der Jagt LH,Jansonius NM. Three portable tonometers,the TGDc-01,the ICARE and the Tonopen XL,compared with each other and with Goldmann applanation tonometry [J]. Ophthalmic Physiol Opt, 2005,25 (5) : 429-435.
  • 3Poostchi A,Mitchell R,Nicholas S,et al. The ICARE rebound tonometer: comparisons with Goldmann tonometry,and influence of central corneal thickness [J]. Clin Experiment Ophthalmol, 2009,37 (7) : 687-691.
  • 4Rehnman JB,Martin L. Comparison of rebound and applanation tonometry in the management of patients treated for glaucoma or ocular hypertension [J]. Ophthalmic Physiol Opt,2008,28(4): 382-386.
  • 5Pakrou N,Gray T,Mills R,et al. Clinical comparison of the ICARE tonometer and Goldmann applanation tonometry [J]. J Glaucoma, 2008,17 ( 1 ) : 43-47.
  • 6Munkwitz S,Elkarmouty A,Hoffmann EM,et al. Comparison of the ICARE rebound tonometer and the Goldmann applanation tonometer over a wide IOP range [J]. Graefes Arch Clin Exp Ophthalmol, 2008,246( 6 ) : 875-879.
  • 7Brusini P,Salvetat ML,Zeppieri M,et al. Comparison of ICARE tonometer with Goldmann applanation tonometer in glaucoma patients [ J ]. J Glaucoma, 2006,15 ( 3 ) : 213-217.
  • 8Garcia-Resua C,Gonzalez-Meijome JM,Gilino J,et al. Accuracy of the new ICARE rebound tonometer vs. other portable tonometers in healthy eyes[J]. Optom Vis Sci,2006,83(2):102-107.
  • 9Fernandes P, Diaz-Rey JA, Queiros A,et al. Comparison of the I- CARE rebound tonometer with the Goldmann tonometer in a norreal population [J]. Ophthalmic Physiol Opt,2005,25 (5): 436-440.
  • 10Wingert TA,Bassi CJ,Mcalister W H,et al. Clinical evaluation of five portable tonometers [J].J Am Optom Assoc,1995,66(11): 670-674.

二级参考文献9

  • 1Schreiber W, Vorwerk CK, Langenbucher A, et al. A comparison of rebound tonometer(ICARE) with tonometer and GAT. Ophthalmologe 2007;104(4):299-304
  • 2Sahin A, Basmak H, Niyaz L, et al. Reproducibility and tolerability of the ICARE rebound tonometer in school children. J Glaucoma 2007;16(2):185-188
  • 3Toyohara K, Inoue K, et al. Comparison of ICARE rebound tonometer with GAT and non-contact tonometer. Atarashii (Journal of the Eye) 2007;24(3):355-359
  • 4López-Caballero C, Contreras I, Muos-Negrete FJ, et al. Rebound tonometry in a clinical setting.Comparison with applanation tonometry. Esp Oftalmol 2007;82(5):273-278
  • 5Queirós A, González-Méijome JM, Fernandes P, et al. A comparison of central and peripheral intraocular pressure using rebound tonometry. Ophthal Physiol Opt 2007;27(5):506-511
  • 6Davies LN, Bartlett H, Mallen EA, et al. Clinical evaluation of rebound tonometer. Acta Ophthalmol Scand 2006;84(2):206-209
  • 7傅伟才,张琳.开角型青光眼患者昼夜眼压监测及其意义[J].上海交通大学学报(医学版),2007,27(9):1163-1165. 被引量:3
  • 8王瑾,周和政,杨帆,曾波.PASCAL动态轮廓眼压计临床应用的初步报告[J].国际眼科杂志,2008,8(3):519-521. 被引量:8
  • 9周和政,沈政伟,金中秋,王柏川,鄢俊杰.辩证地认识眼压在青光眼诊断中的地位[J].美中国际眼科杂志,2002,2(2):20-23. 被引量:13

共引文献11

同被引文献46

  • 1林军,刘嘉萍,林瑶敏,吴京红,张秀兰,刘顺英,鲁沛才,卓业鸿.Goldmann眼压准确测理方法和技巧的探讨[J].现代医学仪器与应用,2005,17(12):62-65. 被引量:1
  • 2王涛,刘磊,李志辉,胡士敏,杨文利,朱晓青.应用超声生物显微镜研究睫状肌麻痹剂对眼前节结构的影响[J].中华眼科杂志,1998,34(2):137-140. 被引量:28
  • 3曾阳发,刘杏,黄晶晶,蔡小于,钟毅敏,毛真,王涛,肖辉.中央角膜厚度对Goldmann眼压计与非接触式眼压计测量值的影响[J].中国实用眼科杂志,2007,25(4):393-396. 被引量:8
  • 4彭大伟.眼压测量和眼压描记[M].临床青光眼.第二版.北京:人民卫生出版社,2000:40.62.
  • 5Jorge J, Fernandes P, Queiro's A, et al. Comparison of the IO- Pen and iCare rebound tonometers with the Goldmann tonome- ter in a normal population[J]. Ophthal Physiol Opt, 2010, 30 (1):108-112.
  • 6Abraham LM, Epasinghe NCR, Selva D, et al. Comparison of the ICares rebound tonometer with the Goldmann applanation tonometer by experienced and inexperienced tonometrists [J]. Eye, 2008,22 (4) : 503-506.
  • 7Moreno-Montanes J,Garcla N,FemOndez-Hortelano A,et al.Re- bound tonometer compared with goldmann tonometer in normal and pathologic eorneas [ J ].Cornea, 2007,26 : 427-430.
  • 8Tucker SM, Enzenauer RW, Levin AV, et al. Corneal diameter, axial length, and intraocular pressure in premature infants. Oph- thalmology, 1992, 99(8) : 1296-3000.
  • 9Anna L, Helena S, Enping C. Application of the ICare rebound tonometer in healthy infants. J Glaucoma, 2011, 20( 1 ) : 7-9.
  • 10Eisenberg DL, Sherman BG, McKeown CA, et al. Tonometry in adults and children. A manometric evaluation of pneumatonome- try, applanation, and TonoPen in vitro and in vivo. Ophthalmolo- gy, 1998, 105(7): 1173-1181.

引证文献5

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部