摘要 目的通过比较不同扫描参数的图像质量,探讨全身背景抑制磁共振扩散成像(DWIBS)的最佳扫描技术。方法 15位健康志愿者、21例患者接受了DWIBS成像,通过ADC值、质量指数、SNR及AppCNR来评价不同扫描技术:体线圈和体部表面线圈扫描;SPAIR-EPI与STIR-EPI的比较。最后评价高分辨率MPR及MIP图像的可行性。结果体部表面线圈扫描的质量指数、SNR及AppCNR高于体线圈扫描(P<0.01),在b=800s/mm2时图像质量指数及AppCNR时最高(P=0.000);STIR-EPI质量指数较高(P<0.05);STIR-EPI的AppCNR优于SPAIR-EPI(P<0.05)。结论采用体部表面线圈,b=800s/mm2,层厚5mm,STIR脂肪抑制技术以及TR值在6000ms以上可以获得高质量的DWIBS图像。 Objective To study the optimal technique of MR diffusion weighted imaging with background body signal suppression(DWIBS) on whole-body diseases according to the comparison of image quality with different parameters.Methods Fifteen healthy volunteers and twenty-one patients underwent DWIBS on a 3.0T MRI,using different parameters including Body coil vs.phased array surface coil methods,SPAIR-EPI vs.STIR-EPI,ADC value,SNR and AppCNR were analyzed for different parameters.And the feasibility of high resolution MPR and MIP images were evaluated.Results The ADC value,SNR and AppCNR of phased array surface coils methods were better than Body coil methods(P0.05).The quality and AppCNR of STIR-EPI were better than SPAIR(P0.05).Conclusion High quality DWIBS can be achieved on 3.0T MRI with optimal parameters(phased array surface coil scans,b800,STIR-EPI and TR 6000).
出处 《中国医疗设备》 2011年第8期28-31,51,共5页 China Medical Devices
基金 无锡市科技局立项课题(CSE00812)
关键词 磁共振扫描仪 磁共振成像 背景抑制磁共振扩散成像 全身磁共振弥散加权成像 magnetic resonance imaging diffusion weighted imaging with background body signal suppression whole body
  • 相关文献

参考文献9

  • 1张秀梅,刘新纯.MRI影像质量评价指标及其优化[J].中国医疗设备,2008,23(1):46-49. 被引量:7
  • 2倪萍,陈自谦,钱根年,肖慧,许尚文,陈锦华.BOLD-fMRI研究过程若干质量控制因素分析[J].医疗设备信息,2007,22(12):6-9. 被引量:16
  • 3Petra Mürtz,Carsten Krautmacher,Frank Tr?ber,Jürgen Gieseke,Hans H. Schild,Winfried A. Willinek.Diffusion-weighted whole-body MR imaging with background body signal suppression: a feasibility study at 3.0 Tesla[J]. European Radiology . 2007 (12)
  • 4J. H. M. Chan,E. Y. K. Tsui,S. H. Luk,A. S. L. Fung,M. K. Yuen,M. L. Szeto,Y. K. Cheung,K. P. C. Wong.Diffusion-weighted MR imaging of the liver: distinguishing hepatic abscess from cystic or necrotic tumor[J]. Abdominal Imaging . 2001 (2)
  • 5T. Ichikawa,H. Haradome,J. Hachiya,T. Nitatori,T. Araki.Diffusion-weighted MR imaging with single-shot echo-planar imaging in the upper abdomen: preliminary clinical experience in 61 patients[J]. Abdominal Imaging . 1999 (5)
  • 6Muro I,Takahara T,Horie T, et al.Influence of respiratory.motion in body diffusion weighted imaging under free breathing (examination of a moving phantom). Nippon Hoshasen Gijutsu Gakkai Zasshi . 2005
  • 7Papanikolaou N,Moulopoulos LA,Gouliamos A,et al.Comparison of dual soin echo planar imaging(SE-EPI), turbo spin echo sequences for T2-weighted MR imaging of focal liver lesion. Magnetic Resonance Imaging . 2000
  • 8Takahara T,Imai Y,Yamashita T,et al.Diffusion weighted whole body imaging with background body signal suppression (DWIBS): technical improvement using free breathing, STIR and high resolution 3D display. Radiation Medicine . 2004