期刊文献+

高压注射器智能导流报警系统设计 被引量:3

Design on Intelligent Diversion & Alarm System of High Pressure Syringe
下载PDF
导出
摘要 为解决当前医用高压注射器注射药液时因注射速度快、压力高而易发生渗漏的问题,采用TI公司的微控制器MSP430F449及精量电子FS2050压力传感器,设计了一套高压注射器防渗漏智能导流报警系统,可实时监测高压注射过程。一旦监测到渗漏立即进行声光报警,同时迅速进行导流切换。本系统不仅克服了以往机械设计方案易松动的缺点,而且还增加了导流切换功能可有效减轻病人的痛苦。动物实验结果表明,智能导流报警系统稳定可靠、灵敏度高,可有效避免高压注射器的药液渗漏现象的发生。 The leakage of high-pressure syringe is usually happened during the course of clinical applications due to its rapid injection speed and high pressure.Based on Texas Instrument(TI)'s microcontroller MSP430F449 and Measurement Specialties Inc(MEAS)'s pressure sensor FS2050,an intelligent diversion alarm system is designed to prevent the leakage of high-pressure syringe designed.The system can offer real-time monitoring for the full injection process,and start sound light alarm,switch diversion immediately once the leakage is monitored.The intelligent system can not only overcome the shortcomings of easy to loose on mechanical designs,but also increase the switching function which can effectively reduce the pain of patients.Animal experimental results show that the intelligent diversion alarm system is reliable and sensitive and can effectively prevent the leakage of high-pressure syringes.
出处 《现代电子技术》 2011年第16期164-165,168,共3页 Modern Electronics Technique
基金 重庆市科技攻关计划项目(CSTC2009AC5021)
关键词 报警 压力 导流 传感器 微控制器 alarm pressure leakage sensor microcontroller
  • 相关文献

参考文献8

  • 1何晓琴,苟军.CT增强扫描造影剂外渗的预防和护理[J].西南军医,2009,11(5):975-976. 被引量:21
  • 2Texas Instruments. MSP430Fx44x mixed signal microcontroller[EB/OL]. [2007-10-14]. http://www. ti. com. cn.
  • 3Texas Instruments. MSP430x4xx family user's guide[EB/ OL]. [2009-04-21]. http,//www. ti. com. cn.
  • 4曹晓梅,何欣,陈贵海.传感器节点定位系统攻防机制研究[J].计算机科学,2008,35(7):36-41. 被引量:3
  • 5Texas Instruments. Datasheet of Nep1117[EB/OL]. [2004-07-21]. http://www. ti. com. cn.
  • 6舍茨.实用电子元器件与电路基础[M].夏建生,王仲奕,刘晓晖,等译.2版.北京:电子工业出版社,2009.
  • 7陈霖,陈云龙,黎海涛,等.自毁式导流阀:中国,ZL2000820099740.5[P/OL].[2009-06-03].http://www.sipo.gov.cn/sip02008/zljs/.
  • 8陈霖,陈云龙,黎海涛,等.高压注射器血管渗漏控制系统:中国,ZL2000820099741.X[P/OL].[2009-06-03].http://www.sipo.gov.cn/sip02008/zbs/.

二级参考文献30

  • 1王福豹,史龙,任丰原.无线传感器网络中的自身定位系统和算法[J].软件学报,2005,16(5):857-868. 被引量:672
  • 2Wellenhoff B H, Lichtenegger H, Collins J. Global positions system: theory and practice [M]. Fourth Edition, Springer-Verlag, 1997.
  • 3Priyantha N, Chakraborty A, Balakrishnan H. The CRICKET location-support system [C]//Proc. of the 6th Annual International Conference on Mobile Computing and Networking (MOBICOM' 00). Boston, MA, August 2000:32-43.
  • 4Nicelescu D, Nath B.. Ad hoc positioning (APS) using AOA. // Proc. of Joint Conference of the IEEE Computer and Communications Societies ( INFOCOM'03 ). San Francisco, CA, April 2003 : 1734-1743.
  • 5Bahl P, Padmanabhan V N. RADAR: An in-building RF-based user location and tracking system [C]//Proc. of Joint Conference of the IEEE Computer and Communications Societies (INFOCOM' 00). Tel Aviv, Israel, March 2000 : 775-784.
  • 6Bulusun N, Heidemann J, Estr I D. GPS-less low cost outdoor localization for very small devices [J]. IEEE Personal Communications,2000,7 (5) :28-34.
  • 7He T, Huang C, Bium B M, et al. Range - free localization scheroes in large scale sensor networks [C]//Proc. of the 9th Annual International Conference on Mobile Computing and Networking (MOBICOM'03). San Diego,CA, August 2003 : 81-95.
  • 8Niculescu D, Nath B. DV-based positioning in ad hoc networks [J]. Journal of Telecommunieation Systems, 200a, 22 (1/4): 267-280.
  • 9Nagpal R,Shrobe H,Bachrach J. Organizing a global coordinate system from local information on an ad hoe sensor network [C] //Proc, of the 2nd International Worksho Pon Information Processing in Sensor Networks ( IPSN ' 03 ). Palo Alto, California, April 2003: 151-152.
  • 10Sastry N,Shankar U,Wagner D. Secure verification of location claims [C]//Proc. of the 2003 ACM Worksho Pon Wireless Security (WISE'03). San diego,California,september 2003:1-10.

共引文献22

同被引文献26

  • 1李国英,张殿星,孙钢,张维青,李胜辉,李海军.静脉注射血管外渗报警装置的研制[J].医学影像学杂志,2004,14(9):785-786. 被引量:1
  • 2曹玉珍,刘洪涛,张力新,王玮.恒温恒湿药物稳定性试验箱的研制[J].医疗卫生装备,2004,25(12):9-10. 被引量:3
  • 3泰龙.MSP430单片机常用模块与综合系统实例精讲[M].北京:电子工业出版社,2007.
  • 4Texas Instruments.MSP430Fx44x family user's guide[EB/OL].[2009-04-21].http://www.ti.corn.cn.
  • 5Powell C C,Li J M,Rodino L,et al.A new device to limit extravasa-tion during contrast-enhanced CT[J].AJR Am J Roentgenol,2000,174(2):315-318.
  • 6Nelson R C,Anderson F A Jr,Birnbaum B A,et al.Contrast mediaextravasation during dynamic CT:detection with an extravasation detec-tion accessory[J].Radiology,1998,209(3):837-843.
  • 7赵小琼,邓卫警,沙立民,等.CT扫描增强剂外漏致肢体急性肿胀与缺血的应急治疗与护理6例[J].护理学杂志,2003,19(9):54.
  • 8ACIST Medical Systems,Inc.EDATM Extravasation DetectionTechnology[EB/OL].[2011-05-10].http://www.acist.com/radiology/eda.asp.
  • 9Medrad,Inc.How XDS Works:Reliable RF Wave Technology Makesthe Difference[EB/OL].[2011-05-10].http://www.medrad.com/en-us/resources/KnowledgeCenter/Pages/HowXDSWorksReliableRFWave TechnologyMakestheDifference.aspx.
  • 10Lowe D G.Distinctive image features from scale-invariant keypoints[J].International Journal of Computer Vision,2004,60(2):91-110.

引证文献3

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部