摘要
目的采用HPLC和UV分光光度法测定阿奇霉素片溶出度,并从中选择适宜的方法。方法高效液相色谱法,采用ODS-50色谱柱,流动相为0.04M磷酸盐缓冲液(pH 11.0)-乙腈(400∶600),检测波长215 nm,柱温40℃,进样量20μL,溶出介质为磷酸盐缓冲液(pH6.5),转数为75 r/min,45 min取样。紫外分光光度法以磷酸盐缓冲液(pH6.0)为空白溶液及溶剂,检测波长为(482±2)nm,溶出介质为磷酸盐缓冲液(pH6.0),转速为100 r/min,45 min取样。结果溶出曲线方程式为:Y=0.001 3X+0.913,10 min时累计溶出量为92.3%(RSD=1.9%)。回归方程:A=0.007 5C-0.017 1,r=0.999 1。线性范围在17.9-89μg/mL,回收率为99.6%(RSD=0.4%)。结论两种方法都较为准确、简便,测得的溶出效果较好,溶出度结果也基本一致。
Objective To establish an effective and reliable determination for the dissolution of azithromycin tablet through comparing HPLC and UV methods.Methods ODS-50 column was used in HPLC method with phosphate buffer solution(pH 11.0)-acetonitrile(400∶600)as mobile phase.Column temperature was 40 ℃,detective wavelength was 215 nm,and injective volume was 20 μL.Phosphate buffer solution(pH 6.5)was used as dissolution medium.Rotatory speed was 75 rpm,and the sample was taken at 45 min.UV method:phosphate buffer solution(pH 6.0)was used as empty solvent,and detective wavelength was(482±2) nm.Dissolution medium for UV method was phosphate buffer solution(pH 6.0).Rotatory speed was 100 rpm,the sample was taken at 45 min.Results Linear range was 17.9~89.0 μg/mL,and recovery rate was 99.6%(RSD=0.4%).The average accumulated dissolution was 92.3%(RSD=1.9%)at 10 min.Conclusion The two methods are equivalent in determining dissolution of azithromycin tablet,and they are simple and accurate with good dissolution effect.
出处
《实用药物与临床》
CAS
2011年第4期303-305,共3页
Practical Pharmacy and Clinical Remedies