摘要
目的探讨雅培i2000全自动免疫发光仪用化学发光免疫分析法,与杭州中翰盛泰科技有限公司艾克美免疫荧光分析仪用免疫荧光干化学层析法检测总前列腺特异性抗原(TPSA)的可比对性。方法对免疫荧光干化学层析法定量检测TPSA进行了性能评价,内容包括批内重复性、线性、准确度。结果该法重复测定高、低值样本,CV<10%;线性回归方程Y=0.9748X+0.048;与化学发光免疫法的相关性分析,r=0.984。结论免疫荧光干化学层析法定量检测系统与雅培i2000全自动大型化学发光仪具有良好的重复性、线性和方法学的相关性,非常适合临床实验室门诊急诊及小批量样本时使用。
Objective To explore the comparability between ABBOTI'i2000 automatic luminometer and immune i-CHROMA Reader immunofluoreseence analyzer in the detection of total prostate specific antigen (TSPA). Methods Immunofluorescence dry chemical chromatography method was used for quantitative detection of TPSA in ovder to evaluate the performance,including intra-repeatability, linearity, accuracy. Results Determination of high- repetition, low-value samples, CV 〈 10% ; the linear regression equation Y = 0. 9748X + 0. 048 ; and Chemilumines- cence immunoassay analysis, r = 0. 984. Conclusion Immunofluorescence dry chemical chromatography quantitative detection systems and ABBTTi2000 automatic chemiluminescent instrument has good repeatability, linearity and methodology correlation and is very suitable for clinical laboratories and small-volmne samples of outpatient and emergency use.