摘要
目的:探讨低仟伏值(kV)低辐射量MSCT扫描对泌尿系结石的检出率及对图像质量的影响。方法:将4种单一成分的泌尿系结石108粒放入猪里脊肉内制成结石模型,应用64层螺旋CT对模型进行扫描,扫描的管电压分别为120 kV1、00 kV和80 kV,并与不同的管电流进行组合,层厚为5 mm,层间距为5 mm,螺距为0.984。扫描后的原始数据(5 mm层厚)重建成1.25 mm层厚,使用GE ADW4.3工作站对数据进行处理。测量图像质量的噪声水平,记录X线辐射剂量指标CTDIvol。结果:与对照组120kV-250mAs相比,实验组80kV-150mAs及100kV-50mAs组的噪声SD值仅分别升高7.60%和8.62%,主观评分均为合格,结石检出率为100%,而辐射剂量降低率分别达到81.40%和87.60%。结论:合适的低kV-mAs组合对泌尿系结石的检出率可以达到100%,且明显降低辐射剂量。
Objective:To investigate the detectability of urinary calculi undergoing 64-slice CT scanning with reducing kV and to evaluate the image quality,so as to obtain the optimal scanning program.Methods:The calculi model was made by inserting 108 granules of urinary calculus containing four kinds of simple constituent into pork loin,and then scanned by 64-slice CT.The scanning parameters were as follows:three different dose-groups were chosen according to the fixed kV set 120kV,100kV,80kV;slice thickness and interval 5mm;the pitch is 0.984;the raw data were reconstructed into 1.25mm;all the reconstructed images were transferred to GE ADW4.3 workstation.To evaluate image quality according to different dose-groups and counted the number of detected calculus respectively.The standard deviation(SD) and signal to noise ratio(SNR) of the images were measured.CT dose index volume(CTDIvol) was obtained automatically.Results:120kV^250mAs was set to be the control group.Compared with the control group,80kV^150mAs,100kV^50mAs and 120kV^30mAs noise SD values were only increased by 7.60%,8.62% and 9.07%.The detection rate was 100%,while the radiation dose decreased significantly with the rate reaching 81.40%,87.60% and 87.98% respectively.Conclusion:It can reach 100% for the detection rate of urinary stones using the right combination of low kV-mAs scaning,and significantly reduce the radiation dose.
出处
《放射学实践》
北大核心
2011年第8期810-814,共5页
Radiologic Practice