加速器机载锥形束CT输出量及稳定性分析 被引量:2

Radiation Output Evaluation of Kilovoltage Cone Beam CT Unit
下载PDF
导出
摘要 目的分析医用电子直线加速器机载kV级锥形束CT的输出剂量及其稳定性。方法用Unidos剂量仪和0.6cc电离室在空气中对Synergy机载的XVIkV级锥形束CT的空气比释动能、剂量输出稳定性、百分深度剂量进行测量。结果拍片模式下XVI空气比释动能随管电压呈二次方变化;空气比释动能随管电流和曝光时间线性变化;透视模式下,空气比释动能随图像桢数线性变化。输出剂量的重复性以及随机架角度的波动小于±1%。体模中百分深度剂量随管电压增加而变大。结论锥形束CT的剂量输出比较稳定,但随成像参数的设置变化范围大。临床应用时需要合理选择曝光条件,尽量减少患者接受的剂量。 Objective To evaluate the radiation output and stability of linac-integrated kV cone beam CT unit. Methods Air kermas in radiographic mode were measured with 0.6cc ion chamber and Unidos electrometer for Synergy-integrated XVI kV cone beam CT unit. Air kermas vs image frames were measured in fluoroscopic mode. Output stability and depth doses were measured. Results The air kerma increased quadratically with the increased tube voltage, while increasing linearly with the tube current, exposure time, and number of frames. The radiation output stability and its change with the gantry angle were within ± 1%. The percentage depth dose increased with higher tube voltage. Conclusion The radiation output of XVI is stable. The radiation outputs change considerably with the preset parameters. Parameters should be optimally chosen to reduce the patient dose.
出处 《中国医疗器械杂志》 CAS 2011年第5期383-385,共3页 Chinese Journal of Medical Instrumentation
关键词 锥形束CT 辐射剂量 空气比释动能 影像引导放射治疗 cone-beam CT, radiation dose, air kerma, image-guided radiotherapy
  • 相关文献

参考文献5

  • 1Saw CB, Yang Y, Li F, et al. Performance characteristic and quality assurance aspects of kilovoltage cone-beam CT on medical linear accelerator[J]. Med Dosim, 2007, 32(2): 80-85.
  • 2Oelfke U, Tucking T, Nill S, et al. Linac-integrated kv-cone beam CT: Technical features and first applications[J]. Med Dosim, 2006, 31(1):52-70.
  • 3Moores BM. Patient dose measurements in radiological practices[J]. Phys Med Biol, 2006, 51 : L31-L34.
  • 4ICRU. Patient dosimetry for x rays used in medical imaging[C]. ICRU Report 74, Bethesda, Maryland, ICRU,2005.
  • 5Islam MK, Purdie TG, Norrlinger BD, et al. Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy[J]. Med Phys, 2006, 33(6): 1573-1582.

同被引文献19

  • 1戴建荣,胡逸民.图像引导放疗的实现方式[J].中华放射肿瘤学杂志,2006,15(2):132-135. 被引量:184
  • 2潘才住,潘建基,陈传本,张秀春,李金莲.电子射野影像系统测量鼻咽癌调强放疗的摆位误差[J].福建医科大学学报,2007,41(1):49-52. 被引量:24
  • 3Zbijewski W,De Jean P,Prakash P,et al.A dedicated cone-beam CT system for musculoskeletal extremities imaging:Design,optimization,and initial performance characterization [ J ].Med Phys,2011,38(8):4700-13.
  • 4Hassan B,van der Steh P,Sanderink G.Accuracy of three-di-mensional measurements obtained from cone beam computed tomo-graphy surface-rendered images for cephalometric analysis:influ-ence of patient scanning position [ J ].Eur J Orthod,2009,31(2):129-34.
  • 5Hoffmann E,da Silva A M,Azevedo D,et al.Evaluation of im-age quality parameters of small and full field of view cone beam computed tomography dental imaging systems [ J ].Med Phys,2012,39(6):3627-8.
  • 6Ding G X,Duggan D M,Coffey C W.Characteristics of kilovolt-age x-ray beams used for cone-beam computed tomography in radi-ation therapy [ J ].Phys Med Biol,2007,52(6):1595-615.
  • 7Kan M W,Leung L H,Wong W,et al.Radiation dose from cone beam computed tomography for image-guided radiation therapy [J].Int J Radiat Oncol Biol Phys,2008,70(1):272-9.
  • 8Bissonnette J P,Baiter P A,Dong L,et al.Quality assurance for image-guided radiation therapy utilizing CT-based technologies:A report of the AAPM TG-179[J].Med Phys,2012,39(4):1946-63.
  • 9Mackie TR, Kaptoes J, Ruchala K, et al. Image guidance for precise conformal radiotherapy [ J ]. Int J Radiat Oncol Biol Phys,2003,56 ( 1 ) :89-105.
  • 10胡逸民,适形放射治疗[A]//胡逸民,杨定宇.肿瘤放射治疗技术[M].北京:北京医科大学中国协和医科大学联合出版社,1999:98.

引证文献2

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部