摘要 This study investigated the effect of velocity encoding on measurement of brain blood flow and blood volume of inflow and outflow using phase-contrast magnetic resonance angiography. A single two-dimensional phase-contrast magnetic resonance angiography slice was applied perpendicular to the internal carotid artery and the vertebral artery at C2 level. For each subject, the velocity encoding was set from 30 to 90 cm/s with an interval of 10 cm/s for a total of seven settings. Various velocity encodings greatly affected blood flow volume, maximal blood flow velocity and mean blood flow velocity in the internal carotid artery, but did not significantly affect vertebral arteries and jugular veins. When velocity encoding was 60-80 cm/s, the inflow blood volume was 655 _+ 118 mL/min, and the outflow volume was 506 _+ 186 mL/min. The ratio of outflow/inflow was steady at 0.78-0.83, and there was no aliasing in any of the images. These findings suggest that velocity encodings of 60 80 cm/s should be selected during measurement of cerebral blood flow volume using phase-contrast magnetic resonance angiography. This study investigated the effect of velocity encoding on measurement of brain blood flow and blood volume of inflow and outflow using phase-contrast magnetic resonance angiography. A single two-dimensional phase-contrast magnetic resonance angiography slice was applied perpendicular to the internal carotid artery and the vertebral artery at C2 level. For each subject, the velocity encoding was set from 30 to 90 cm/s with an interval of 10 cm/s for a total of seven settings. Various velocity encodings greatly affected blood flow volume, maximal blood flow velocity and mean blood flow velocity in the internal carotid artery, but did not significantly affect vertebral arteries and jugular veins. When velocity encoding was 60-80 cm/s, the inflow blood volume was 655 _+ 118 mL/min, and the outflow volume was 506 _+ 186 mL/min. The ratio of outflow/inflow was steady at 0.78-0.83, and there was no aliasing in any of the images. These findings suggest that velocity encodings of 60 80 cm/s should be selected during measurement of cerebral blood flow volume using phase-contrast magnetic resonance angiography.
出处 《Neural Regeneration Research》 SCIE CAS CSCD 2011年第23期1796-1800,共5页 中国神经再生研究(英文版)
基金 the Medical Program of the Scientific & Technical Foundation in Xiamen(MRI study of chronic cerebrovascular insufficiency) in 2008,No.3502Z20084028
关键词 cerebral blood flow velocity encoding PHASE-CONTRAST magnetic resonance imaging cerebral blood flow velocity encoding phase-contrast magnetic resonance imaging
  • 相关文献

参考文献30

  • 1Spilt A,Box FM,van der Geest RJ,et al.Reproducibility of total cerebral blood flow measurements using phase contrast magnetic resonance imaging.J Magn Reson Imaging.2002;16(1):1-5.
  • 2Buijs PC,Krabbe-Hartkamp MJ,Bakker CJ,et al.Effect of age on cerebral blood flow:measurement with ungated two-dimensional phase-contrast MR angiography in 250 adults.Radiology.1998; 209(3):667-674.
  • 3Taviani V,Patterson AJ,Graves MJ,et al.Accuracy and repeatability of fourier velocity encoded M-mode and two-dimensional cine phase contrast for pulse wave velocity measurement in the descending aorta.J Magn Reson Imaging.2010;31(5):1185-1194.
  • 4Amano Y,Takagi R,Suzuki Y,et al.Three-dimensional velocity mapping of thoracic aorta and supra-aortic arteries in Takayasu arteritis.J Magn Reson Imaging.2010;31(6):1481-1485.
  • 5Wentland AL,Wieben O,Korosec FR,et al.Accuracy and reproducibility of phase-contrast MR imaging measurements for CSF flow.AJNR Am J Neuroradiol.2010;31(7):1331-1336.
  • 6Stoquart-El Sankari S,Lehmann P,Gondry-Jouet C,et al.Phase-contrast MR imaging support for the diagnosis of aqueductal stenosis.AJNR Am J Neuroradiol.2009;30(1):209-214.
  • 7Chiang WW,Takoudis CG,Lee SH,et al.Relationship between ventricular morphology and aqueductal cerebrospinal fluid flow in healthy and communicating hydrocephalus.Invest Radiol.2009; 44(4):192-199.
  • 8Nedelmann M,Kaps M,Mueller-Forell W.Venous obstruction and jugular valve insufficiency in idiopathic intracranial hypertension.J Neurol.2009;256(6):964-969.
  • 9Hansen KL,Udesen J,Oddershede N,et al.In vivo comparison of three ultrasound vector velocity techniques to MR phase contrast angiography.Ultrasonics.2009;49(8):659-667.
  • 10Lotz J,Meier C,Leppert A,et al.Cardiovascular flow measurement with phase-contrast MR imaging:basic facts and implementation.Radiographics.2002;22(3):651-671.

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部