In-Vitro and In-Vivo Electrical Characteristics of a Penetrating Microelectrode Array for Optic Nerve Electrical Stimulation

In-Vitro and In-Vivo Electrical Characteristics of a Penetrating Microelectrode Array for Optic Nerve Electrical Stimulation
原文传递
导出
摘要 The Chinese C-Sight team aims to restore vision to blind patients by means of stimulating the optic nerve with a penetrating microelectrode array. A biocompatible, implantable microwire array was developed having four platinum-iridium shafts, each 100μm in diameter. This penetrating microwire array is described in this paper, including its fabrication techniques and its in-vitro electrical characteristics. Every set of four shafts was spaced 0.4mm from center to center, comprising two short shafts that were 0.3mm long and two that were 0.9mm long. This design was intended to stimulate ganglion cell axons at different depths within the optic nerve. In-vitro electrochemical impedance testing results showed that the impedance at 1kHz ranged from 8 to 10kΩ at room temperature. The voltage responses of the arrays to current pulse stimulation indicated a charge-injection capacity of 210μC/cm2. Finally, in-vivo acute animal experiments showed that the amplitude of the electrically evoked potentials (EEPs) measured in primary visual cortex could be as large as 100 μV upon direct stimulation of the optic nerve. The Chinese C-Sight team aims to restore vision to blind patients by means of stimulating the optic nerve with a penetrating microelectrode array. A biocompatible, implantable microwire array was developed having four platinum-iridium shafts, each 100μm in diameter. This penetrating microwire array is described in this paper, including its fabrication techniques and its in-vitro electrical characteristics. Every set of four shafts was spaced 0.4mm from center to center, comprising two short shafts that were 0.3mm long and two that were 0.9mm long. This design was intended to stimulate ganglion cell axons at different depths within the optic nerve. In-vitro electrochemical impedance testing results showed that the impedance at 1kHz ranged from 8 to 10kΩ at room temperature. The voltage responses of the arrays to current pulse stimulation indicated a charge-injection capacity of 210μC/cm2. Finally, in-vivo acute animal experiments showed that the amplitude of the electrically evoked potentials (EEPs) measured in primary visual cortex could be as large as 100 μV upon direct stimulation of the optic nerve.
出处 《Journal of Shanghai Jiaotong university(Science)》 EI 2011年第5期614-619,共6页 上海交通大学学报(英文版)
基金 the National Natural Science Foundation of China (Nos. 30700217 and 60971102) the National Basic Research Program (973) of China(Nos. 2011CB707502 and 2011CB707505)
关键词 optic nerve visual prosthesis platinum-iridium (Pt/Ir) microelectrode charge injection capacity impedance optic nerve visual prosthesis, platinum-iridium (Pt/Ir) microelectrode, charge injection capacity, impedance
  • 相关文献

参考文献19

  • 1DOBELLE W H, MLADEJOVSKY M G. Phosphenes pro- duced by electrical stimulation of human occipital cor- tex, and their application to the development of a prosthesis for the blind [J]. The Journal of Physiology, 1974, 243(2): 553-576.
  • 2TROYK P, BAK M, BERG J, et al. A model for intra- cortical visual prosthesis research [J]. Artificial Organs, 2003, 27(11): 1005-1015.
  • 3CHOW A Y, CHOW V Y. Subretinal electrical stimula- tion of the rabbit retina [J]. Neuroscience Letter, 1997, 225(1): 13-16.
  • 4ZRENNER E, MILICZEK K D, GABEL V P, et al. The development of subretinal microphotodiodes for replacement of degenerated photoreceptors [J]. Oph- thalmic Research, 1997, 29(5): 269-280.
  • 5HUMAYUN M S, WEILAND J D, FUJI1 G Y, et al. Visual perception in a blind subject with a chronic microelectronic retinal prosthesis [J]. Vision Research, 2003, 43(24): 2573-2581.
  • 6RIzzo J F, WYATT J, LOEWENSTEIN J, et al. Meth- ods and perceptual thresholds for short-term electrical stimulation of human retina with microelectrode ar- rays [J]. Investigative Ophthalmology and VisualScience, 2003, 44(12): 5355-5361.
  • 7VERAART C, RAFTOPOULOS C, MORTIMER J T, et al. Visual sensations produced by optic nerve stimulation using an implanted self-sizing spiral cuff electrode [J]. Brain Research, 1998, 813(1): 181-186.
  • 8BRELEN M E, DURET F, GERARD B, et al. Creating a meaningful visual perception in blind volunteers by optic nerve stimulation [J]. Journal of Neural Engi- neering, 2005, 2(1): 22-28.
  • 9SAKAGUCHI H, FUJIKADO T, FANG X Y, et al. Tran- sretinal electrical stimulation with a suprachoroidal multichannel electrode in rabbit eyes[J]. Japanese Journal of Ophthalmology, 2004, 48(3): 256-261.
  • 10FANG X Y, SAKAGUCHI H, FUJIKADO T, et al. Direct stimulation of optic nerve by electrodes implanted in optic disc of rabbit eyes [J]. Graefe's Archive for Clin- ical and Experimental Ophthalmology, 2005, 243(1): 49-56.

二级参考文献6

  • 1CHU Yin ping 1,CHEN Ru kun 2,FAN Yu bo 3 , PU Fang 31 Department of Thoracic & Cardiovascular Surgery Second Hospital Affiliated Shanxi Medical University, Taiyuan 030001,China 2 Second Affiliated Hospital of Zhejiang University Medical.Hydrodynamics of a Newly China-Made Jiuling Bileaflet Heart Valve Prostheses[J].Chinese Journal of Biomedical Engineering(English Edition),2000,9(4):168-179. 被引量:8
  • 2Reul H,van Son JAM,Steinseifer U,et al.In vitro comparison bileaflet aortic heart valve prostheses: St.Jude Medical , CarboMedics, modified Edwards-Duromedics, and Sorin-Bicarbon valves[].Journal of Thoracic and Cardiovascular Surgery.1993
  • 3Remadi JP,Baron O,Tribouilloy C,et al.Bivalvular mechanical mitral -aortic valve replacement in 254 patients: long-term results--a 22-year follow-up[].The Annals of Thoracic Surgery.2003
  • 4Jazayeri S,Gomez MC,Tatou E,et al.Clinical experience and Doppler echocardiographic assessment of the first one hundred ATS AP ( advanced performance ) prosthetic valve in the aortic position[].Journal of Heart Valve Disease.2003
  • 5Butterfield M,Fisher J,Davies GA,et al.Comparative study of the hydrodynamic function of the CarboMedics valve[].The Annals of Thoracic Surgery.1991
  • 6Roudaut R,Roques X,Lafitte S,et al.Surgery for prosthetic valve obstruction.A single center study of 136 patients[].European Journal of Cardio Thoracic Surgery.2003

共引文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部