虹膜定位技术在波前像差引导的LASIK手术过程的配合与要求 被引量:1

The investigation of the positioning of the iris and the cooperation of patients in wave front aberration guided LASIK
原文传递
导出
摘要 目的探讨虹膜定位联合波前像差引导个体化LASIK术前检查及手术过程中虹膜定位的正确方法。方法采用博士伦公司Zywave波前像差仪,采集在暗室瞳孔自然散大状态下的虹膜数据及波前像差数据,通过网络传输,在切削角膜瓣后未掀开角膜瓣前,启动虹膜定位功能,进行虹膜定位。结果采集虹膜图象时的要求:头位的额状面与波前像差仪的矢状面相垂直,虹膜定位后的正确图象显示为大圆圈与术眼角膜缘吻合,小圆圈与瞳孔吻合,小圆环内的6个反光白点清晰并分布均匀,瞳孔中心位的绿色“+”与坐标的黑色“+”中心位相吻,角膜瓣表面的水分要充足,虹膜纹理调至最清晰,捕捉时的瞳孔直径在3.5—5.0mm最佳。结论掌握好采集虹膜图象时的检查要求、作好虹膜定位的具体步骤,是保证虹膜定位联合波前像差引导的个体化LASIK效果的关键。 Objective To study right methods to localize iris before and during iris localization combined with wavefront aberration guided LASIK. Methods The data of iris and wavefront aberration were collected by the Zywave wavefront analyzer (USA Bausch & Lomb Company) under the dark room with naturally dilated pupil and were transmitted through the network. Iris localization started after cutting corneal flap and before openingit. Results Requirements of capturing iris images were stated as follow. Frontal plane of the head position and sagittal plane of the Zywave wavefront analyzer were perpendicular. The fight images after iris localization showed large circle coincided with the edge of the cornea, small circle coincided with the pupil, six reflective white spots in small circle were clear and evenly distributed. The green " + " in the center position of the pupil coincided with the center of black " + " in axis. The water on corneal flap surface should be sufficient, iris textures were the clearest, and pupil diameter from 3.5 to 5.0 mm during capturing images were optimal. Conclusions The keys to ensure the effect in iris localization combined with wave-front aberration guided LASIK were the inspection of requirements during capturing images, and the specific steps during iris localization were well made.
出处 《中华眼外伤职业眼病杂志》 2011年第10期781-783,共3页 Chinese Journal of Ocular Trauma and Occupational Eye Disease
关键词 虹膜定位 波前像差 Iris localization Wavefront aberration
  • 相关文献

参考文献10

  • 1Bahar I, Levinger S, Kremer I. Wavefront guided LASIK for myo- pia with the Technolas 217z; results at 3 years. J Refract Surg 2007,23:586 - 591.
  • 2Shaema M, Wachler BS, Chan CC. Higher order aberrations and relative risk of symptoms after LASIK. J Refract Surg 2007,23 : 252 - 256.
  • 3陈跃国.波阵面像差及其引导的准分子激光角膜屈光手术.见:王宁利,赵家良,谢立信,主编.眼科学[M].第1版.北京:中国协和医科大学出版社,2007:93-98.
  • 4Ciccio AE, Durrie DS, StahI JE. Ocular cyclotorsion during cus- tomized laser ablation. J Refract Surg, 2005, 21:772 -774.
  • 5周炜,何正瑞,郑洁,黄蕊,彭子春.近视眼LASIK术后屈光度客观评价的研究[J].国际眼科杂志,2007,7(1):133-135. 被引量:20
  • 6刘峰,黄永健,许曼,王旭辉,冯伟,廉井财.眼的高阶像差与视觉质量[J].国际眼科杂志,2007,7(4):1113-1115. 被引量:20
  • 7Chan CC, Boxer Wachter BS. Centration analysis of ablation over the coaxial corneal light reflex for hyperopic LASIK. J Refract Surg,2006,22 : 467 - 471.
  • 8Park SH, Kim M, Joo CK. Measurement of pupil eentroid shift and cyelotorsional displacement using iris registration. Ophthalmo- logica, 2009,223 : 166 - 171.
  • 9樊郑军,徐绍娟,贾泽昊.准分子激光近视矫正术后角膜前表面形态变化的研究[J].国际眼科杂志,2006,6(5):1092-1093. 被引量:15
  • 10贾丽,李金科,米生健,张愉,王晓鲁,魏亚荣.准分子激光术后近视回退和欠矫原因分析[J].国际眼科杂志,2005,5(6):1198-1200. 被引量:5

二级参考文献58

共引文献54

同被引文献6

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部