动态调强放射治疗滑窗技术的剂量和机械稳定性评价 被引量:4

Dosimetric and Mechanical Accuracy Estimation of Sliding Window in Dynamic Intensity-modulated Radiotherapy
下载PDF
导出
摘要 目的利用二维电离室矩阵和动态治疗日志文件,评价动态多叶准直器(MLC)滑窗技术的剂量学稳定性和机械精度。方法基于滑窗技术设计5种动态射野A、B、C、D、E,重复按照5种射野对二维电离室矩阵MatriXX进行照射并分别记录其动态日志文件,对12次的测量数据进行统计分析。结果各相应照射野等中心点剂量重复性分别为±1.49%、±1.12%、±1.12%、±1.07%、±1.07%,最大相对标准偏差为±1.47,γ分析通过率(3%/3mm)均为100%;动态治疗日志文件显示所有叶片定位误差均<0.05cm,定位误差的标准差均值分别为0.0365、0.038、0.038、0.035、0.026cm。结论本实验中动态MLC剂量学稳定性和机械精度满足治疗要求,本方法简单方便,可作为常规MLC质量保证程序的补充。 Objective We estimated dosimetric and mechanical accuracy of sliding window in dynamic intensity-modulated radiotherapy with a 2D diode array and dynalog file. Methods Five kinds of MLC format files based on sliding window technique were designed. 2D array of MatriXX was radiated in the linac and dynalog files were generated twelve times. Measured data was analyzed. Results All the repeatability errors of isocenter point dose of every fields were ± 1.49%,± 1.12%, ± 1.12%, ± 1.07%,± 1.07%, respectively. The maximum of relative standard deviations were ±1.47. All of γ- analyze pass rates were 100%. Dynalog file indicated that all leaf deviations were less than 0.5mm. The averages of errors RMS were 0.0365, 0.038, 0.038, 0.035, 0.026cm, respectively. Conclusion The result of the present work reveals the ideal functioning of the linac for dynamic dose delivery. This method is simple and convenient, which can be introduced into routine QA procedure for MLC.
出处 《中国医疗设备》 2011年第10期21-24,共4页 China Medical Devices
基金 广东省科技计划项目(2010B031600146) 广州医学院博士留学回国人员基金项目(2010-1-11)
关键词 直线加速器 多叶准直器 滑窗技术 剂量仪 调强放疗 linear accelerator multi-leaf collimator sliding window technique dosimeter intensitymodulated radiotherapy therapty
  • 相关文献

参考文献14

  • 1Ezzell G A,Galvin J M, Low D,et al.Guidance document on delivery,treatment planning, and clinical implementation of IMRT:report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee[J].Med.Phys,2003, 30:2089-2115.
  • 2Palta J, Kim S, Li J,et al.Tolerance limits and action levels for planning and delivery of IMRT [J]. Med. Phys,2003,30:1395.
  • 3曾自力.医用加速器多叶准直器的质量保证和质量控制[J].中国辐射卫生,2008,17(4):449-450. 被引量:6
  • 4S.S.Sivakumar,K.Krishna Murthy,C.A.Davis, R. Ravichandran,et al.Clinical implementation of dynamic intensity-modulated radiotherapy:Dosimetric aspects and initial experience[J].Med. Phys,2008,33:64-71.
  • 5Anup K Bhardwaj,TS Kehwar,SK Chakarvarti,et al. Dosimetric and qualitative analysis of kinetic properties of millennium 80 multileaf collimator system for dynamic intensity modulated radiotherapy treatments[J].Journal of Cancer Research and Therapeutics.2007,(3):23-28.
  • 6Venencia CD,Besa P.Commissioning and quality assurance for intensity modulated radiotherapy with dynamic mnltileaf collimator:Experience of the Pontificia Universidad Cat o lica de Chile[J].ApplClin Med Phys,2004,(5):37-54.
  • 7Mohammad Mohammadi and Eva Bezak. Evaluation of" MLC leaf" positioning using a scanning liquid ionization chamber EPID[J].Phys.Med, Biol.2007,(52): 21-33.
  • 8M Okumura,Y Obata,K Shimomura,M Tamura andY Nishimura.The effect of gantry and collimator angles on leaf limited velocity and position in dynamic multileaf collimator intensity-modulated radiation therapy[J].Phys Med Biol, 2010,(55):3101 -3113.
  • 9Clarke MF,Budgell GJ.Use of amorphous silicon EPID for measuring MLC calibration at varying gantry angle[J].Phys Med Biol.2008,53:473-485.
  • 10Kumar MD,Thirumavalavan N,Kfishna DV,et al.QA of Intensity modulated beams using dynamic MLC log files[J]. Med Phys,2006,31:36-41.

二级参考文献20

  • 1马金利,蒋国梁,傅小龙,廖源,吴开良,周莉钧.窄带野胶片检查在多叶准直器质量保证中作用探讨[J].中华放射肿瘤学杂志,2004,13(2):123-127. 被引量:10
  • 2孙朝阳,肖锋,石梅,魏丽春,陆军.Varian多叶准直器(MLC)常规测量方法及质量保证[J].中国医学物理学杂志,2005,22(6):697-698. 被引量:10
  • 3李勤,韩军,邓光宇,魏黎,彭威,胡兵.电子束全身照射的剂量学特性[J].华中科技大学学报(医学版),2006,35(5):669-671. 被引量:2
  • 4林俊华,严惠玉.全身电子线照射辅助配套装置设计与制作[J].医疗设备信息,2007,22(2):15-15. 被引量:3
  • 5Klein EE, Tepper J, Sontag M, et al. Technology assessment of muhileaf collimation: A North American users survey. Int .J Radiat Oncol Bio1,1999,44(3) :705-710.
  • 6Pasquino M, Borca VC, Catuzzo P, et al. Transmission, penumbra and leaf positional accuracy in commissioning and quality assurance program of a multileaf collimator for step-and-shoot IMRT treatments. Tumori,2006,92(6) :511-516.
  • 7Sun Nuclear Corporation. MapCheck Model 1175 Manual. Melbourne (FL) : Sun Nuclear Corporation. 2004.
  • 8Yang Y, Xing L. Quantitative measurement of MLC leaf displacements using an electronic portal image device. Phys Med Biol,2004,49(8) : 1521-1533.
  • 9Jursinic P, Kainz K, Ahunbay E, et al. Use of a 2D array of diodes to test the accuracy of MLC leaf position and gap width. Med Phys, 2006,33(6) : 2147.
  • 10Yang Y, Xing L. Using the volumetric effect of a finite-sized detector for routine quality assurance of multileaf collimator leaf positioning. Med Phys,2003,30(3) :433-441.

共引文献33

同被引文献40

  • 1吴昊,蒋璠,韩树奎,包尚联.常见探测器的空间分辨率比较[J].中国医学物理学杂志,2011,28(2):2501-2504. 被引量:9
  • 2戴建荣,胡逸民,张红志,关莹,张可,王闯.针对患者调强放射治疗计划的剂量学验证[J].中华放射肿瘤学杂志,2004,13(3):229-233. 被引量:65
  • 3鞠忠建,王运来,马林,徐寿平,戴相昆,王连元.用二维电离室矩阵验证多叶准直器叶片到位精度[J].中华放射肿瘤学杂志,2006,15(4):335-338. 被引量:18
  • 4Woo MK,Nico A. Impact of multi-leaf collimator leaf positioning accuracy on intensity modulation radiation therapy quality assurance ion chamber measurements[J].Medical Physics,2005,(05):1440-1445.
  • 5Chui CS,spirou S,LoSaaso T. Testing of dynamic multi-leaf collimation[J].Medical Physics,1996,(05):635-641.
  • 6Simon TA,Kahler D,Simon WE,et a1. An MLC calibration method using a detector array[J].Medical Physics,2009,(10):4495-4503.
  • 7Varian Medica Systems. Varain user manual.Dynalng File Viewer reference guide[M].2002.
  • 8J W Lee,J H Part,J B Chung. Inverse verification of the dose distribution for intensity modulated radiation therapy patientspecific quality assurance using dynamic MLC log files[J].Journal of the Koream Physical Society,2009,(04):1649-1656.
  • 9Litzenberg DW,Moran JM,Fraass BA. Verification of dynamic and segmental IMRT delivery by dynamic log file analysis[J].J App Clin Med Phys,2002,(02):63-72.
  • 10Klein EE,Hanley J,Bayouth J,et a1. Task group 142 report:quality assurance of medical accelerators[J].Medical Physics,2009,(10):4197-4212.doi:10.1118/1.3190392.

引证文献4

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部