摘要 以气相色谱法测定某空气样品中丙酮的浓度,并分析评定了测量过程中5个部分的不确定度。该样品空气中丙酮的浓度为0.74 mg/m3,扩展不确定度为U=0.04 mg/m3(k=2),结果可表示为:(0.74±0.04)mg/m3。此方法可用于测定空气中丙酮含量的不确定度评价,更有助于准确地表述测量结果。 The concentration of acetone in the air sample was measured by Gas Chromatography.The uncertainty of five parts of the measurement process was analyzed and evaluated.The concentration of acetone in the air sample was 0.74 mg/m3,extended uncertainty was 0.04 mg/m3(k=2),and the result of measuring acetone in the air sample was(0.74±0.04) mg/m3.The established method was applicable to the uncertainty analysis for the determination of acetone in air.It was favourable to record measurements more accurately.
出处 《广东化工》 CAS 2011年第10期157-158,共2页 Guangdong Chemical Industry
关键词 气相色谱法 丙酮 不确定度 gas chromatography acetone uncertainty
  • 相关文献

参考文献2

  • 1JJF:1059-1999,测量不确定度评定与表示指南[S].
  • 2国家环境保护总局,空气和废气监测分析方法编委会.《空气和废气监测分析方法(第四版)》[M].北京:中国环境科学出版社,2003,703:597-602.