期刊文献+

APDC-MIBK萃取石墨炉原子吸收光谱法测定痕量Se

DETERMINATION OF TRACE SELENIUM IN BIOLOGICAL AND ENVIRONMENTAL SAMPLES BY APDC-MINK EXTRACTION AND GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROMETRY
下载PDF
导出
摘要 对石墨炉原子吸收测定 Se 及其价态的条件进行了探讨.为防止 Se 的逸失对不同基体改进剂进行了试验.此外,对环境水样中 Se 及其价态,以及血清和发样中 Se 含量进行了分析.方法的特征含量为1.06×10^(-10)g,水样(n=10)及生物样品(n=4)的相对标准偏差(D_(RS))分别为1.9%及5%,回收率95%~114%. A convenient method for the determination of selenium and valency states of seledium in hair,serum and water is developed.Several matrix modifiers are investigated with respect to their stabilizing power for selen- ium in a graphite furnace.Possible interferences from cations and anions are tried and it is shown that their quantities below certain limit do not affect the analyses of selenium.Sensitivity is 1.06×10^(-10)g of selenium per 1% absorption and the relative standard deviation of 10 determinations for samples is less than 5%,whereas the recovery is 95%~114%.
出处 《北京师范大学学报(自然科学版)》 CAS CSCD 1989年第3期61-66,共6页 Journal of Beijing Normal University(Natural Science)
关键词 SE 测定 原子吸收光谱法 graphite furnace atomic absorption spectrometry, selenium, matrix modifier, hair, serum, water