体重指数适应性电流、电压结合前置门控技术对降低CCTA辐射剂量可行性的研究 被引量:2

Feasibility of Reducing Coronary CT Angiography Dosage:The Body Mass Index Dependent mA and kV Selection Combined
原文传递
导出
摘要 目的探讨体重指数(BMI)适应性电流、电压结合前置门控技术对降低冠状动脉CT血管成像(CCTA)辐射剂量及提高图像噪声均匀度的可行性。资料与方法将120例拟行CCTA检查的患者随机分成A、B两组。A组采用固定电流、电压的回顾性心电门控模式;B组采用BMI适应性电流、电压的前置心电门控模式,记录患者的身高、体重、BMI、电流、电压、噪声值、CT剂量指数(CTDI)、有效剂量(ED)等。图像质量评分采用4点评分法。采用SPSS13.0软件对数据进行统计学处理。结果对A组图像噪声与体重、身高、BMI行回顾性相关分析显示:噪声与BMI呈较显著正相关且相关性最大,相关系数为0.557。A组噪声均值小于B组,但B组噪声均匀度好于A组。A组ED为(21.27±1.53)mSv,B组ED为(6.04±1.45)mSv,两组间剂量差异均有统计学意义(P<0.05)。两组之间图像质量无差异。结论在心率<70次/min且心律稳定时,BMI适应性电流、电压结合前置门控技术可以在提高图像噪声均匀度的同时降低ED 71.6%。 Objective To evaluate the feasibility of reducing radiation dose of coronary CT angiography(CCTA) and improving the image noise uniformity through the body mass index(BMI) adapted mA and kV selection combined with prospective electrocardiographic(ECG)-gating technology.Materials and Methods 120 patients underwent CCTA were divided into control group(group A) and experimental group(group B).The retrospective ECG-gating protocol with fixed mA and kV was applied to the patients of group A.The prospective ECG-gating protocol with the BMI-adapted mA and kV was applied to group B.All of the patient's height,weight,BMI,electric current(mA),voltage(kV),noise value,CT dosage index(CTDI),effective dose(ED) were recorded.Four point scoring method was used for evaluating the image quality.Statistical processing was performed by the SPSS13.0 software.Results Correlation analysis was performed between image noise and the body weight,the height,the BMI respectively in group A.The result displayed that the image noise and BMI had significant correlation with the correlation coefficient of 0.557.The mean value of image noise in group A was smaller than that in group B,but Image noise uniformity in group B was better than that in group A.The ED of group A and group B was(21.27±1.53)mSv and(6.04±1.45)mSv.There was significant difference(P〈0.05) of radiation dosage between two groups,but there was no difference of image quality.Conclusion When the heart rate was lower than 70bpm and the heart rhythm was stable,the BMI-adapted mA and kV combined with prospective ECG-gating technique could reduce ED by 71.6% meanwhile the image noise uniformity be improved.
出处 《临床放射学杂志》 CSCD 北大核心 2011年第10期1532-1535,共4页 Journal of Clinical Radiology
关键词 体层摄影术 X线计算机 图像噪声 前置门控 冠状动脉 低剂量 Tomography X-ray computed Image noise Prospective ECG-gating Coronary artery Low dose
  • 相关文献

参考文献3

二级参考文献22

  • 1王怡宁,金征宇,孔令燕,宋兰,张竹花,张抒扬,林松柏,牟文斌,王沄,赵文敏,郭家武.64层螺旋CT冠状动脉成像初步研究[J].中华放射学杂志,2006,40(8):797-801. 被引量:99
  • 2董智,朱杰敏,刘喆,刘军波,李又洁,祁吉.心率过快、心律不齐及起搏器置入患者的电子束CT冠状动脉成像[J].中华放射学杂志,2006,40(3):273-276. 被引量:18
  • 3王继琛,邱建星,王鹤,高莉,孙晓伟,侯振亚,蒋学祥.心电图编辑软件在CT冠状动脉成像中的应用价值[J].中国医学影像技术,2006,22(10):1485-1488. 被引量:21
  • 4高建华,孙宪昶,李剑颖,王贵生,李娜,戴汝平.后置滤过器C_2对64层螺旋CT冠状动脉成像质量的影响[J].中国医学影像技术,2007,23(1):73-76. 被引量:31
  • 5Schoepf UJ, Becker CR. Ohnesorge BM, et al. CT of cornary artery, disease. Radiology.2004, 232 : 18-37.
  • 6Leschka S, Alkadhi H, Plass A, el al. Accuracy of MSCT coronary angiography with 64-slice technology: first experience. Eur Heart J,2005,26 : 1482-1487.
  • 7Herzog C, Arning-Erb M, Zangos S, et al. Multi-detector row CT coronary angiography: influence of reconstruction technique and heart rate on image quality. Radiology, 2006, 238:75-86.
  • 8Hoffmann MH, Shi H, Manzke R, et al. Noninvasive coronary angiography with 16-detector row CT: effect of heart rate. Radiology, 2005, 234:86-97.
  • 9Pannu HK, Flohr TG, Corl FM, et al. Current concepts in mutidetector row CT evaluation of the coromary arteries: principles, techniques, and anatomy. Radiographics ,2003,23 : 111-125.
  • 10Tack D, Maertelaer V, C, evenois PA. Dose reduction in muhidetector CT using attenuation-based online tube current modulation. AJR,2003 ,181:331-334.

共引文献170

同被引文献25

  • 1于康.体重指数和腰围用于肥胖的营养评定[J].中国医学科学院学报,2010,32(1):4-6. 被引量:25
  • 2Austen WG, Edwards JE, Frye RL, et al. A reporting system on patlents evaluated for coronary artery disease. Report of the Ad Hoc committee for grading of coronary artery disease, council on cardiovascular surgery, American heart association[J]. Circulation, 1975, 51($4): 5-40.
  • 3Brenner DJ, Hall EJ. Computed tomography: An increasing source of radiation exposure[J]. New England Journal of Medicine, 2007, 357(22): 2277-2284.
  • 4Menze] H, Schibilla H, Teunen D. European guidelines on quality criteria for computed tomography[M]. Luxembourg: European Commission, 2000, Publication No. EUR 16262.
  • 5EN. Lee EJ, Lee SK, Agid R, et al. Comparison of image quality and radiation dose between fixed tubecurrent and combined automatic tube current modulation angiography[J]. American Journal of Neuroradioloy, 2009, 30(9): in craniocervical CT 1754-1759.
  • 6Huda W, Scalzetti EM, Levin O. Technique factors and image quality as functions of patient weight at abdominal CT[J]. Radiology, 2000, 217(2): 430-435.
  • 7Aldrich JE, Chang SD, Bilawich AM, et al. Radiation dose in abdominal computed tomography: The role of patient size and the selection of tube current[J]. Canadian Association of Radiologists Journal, 2006, 57(3): 152-158.
  • 8Kalra MK, Maher:f:l, D' Souza RV, et al. Detection of urinary tract stones at low-radiation-dose CT with z-axis automatic tube current modulation: Phantom and clinical studies[J]. Radiology 2005, 235(2): 523 529.
  • 9W.P. Shuman,K.R. Branch,J.M. May,L.M. Mitsumori,D.W. Lockhart,T.J.Dubinsky,邱建星(译),唐光健(校).前瞻性和回顾性心电门控冠状动脉64层CT成像:影像质量和病人放射剂量的对照性研究[J].国际医学放射学杂志,2008,31(5):400-400. 被引量:129
  • 10李纬,朱绍成,史大鹏.体质量指数指导扫描条件降低体部CT剂量的可行性研究[J].中华放射医学与防护杂志,2011,31(2):241-242. 被引量:16

引证文献2

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部