期刊文献+

心电采集电路的设计与实现 被引量:10

Design and implementation of detection circuit for electrocardiogram
下载PDF
导出
摘要 介绍一种心电采集检测电路的设计,该设计以AD620和OP07为核心元件,针对心电信号的组成和干扰频率范围,进行了分析,对由电极采集到的心电信号,通过前置放大电路将微弱的心电心电信号放大,并通过低通滤波器、高通滤波器、及50Hz陷波电路滤除干扰最后通过后置放大电路进一步放大得到清晰的心电波形。系统具有高输入阻抗、高共模抑制比、低噪声、低漂移、和高信噪比、成本低等优点。 This paper introduces a specifice design of detection circuit in ECG acquisition system.This design employes AD620 and OP07 as its core.Through the major components of the ECG signal and the frequency range of interference were analysed,weak ECG signal collected by the electrodes was amplified by the preamplifier circuit,and then was filtered out the interference by using a low-pass filer,a high-pass filer,50Hz notch filer and back amplifier circuit,finally a right wave of ECG was gotten.The characteristics of the system show in the merits of high input impedance,high CMRR,low noise,less excursion and high SNR(signal to noise ratio),low cost and so on.
出处 《信息技术》 2011年第10期130-132,135,共4页 Information Technology
关键词 心电监测 放大电路 滤波器 AD620 detection circuit amplifier circuit filter AD620
  • 相关文献

参考文献7

二级参考文献12

  • 1姜苇,李刚,虞启琏,林凌,刘理,谢国明.新型高共模抑制比生理电信号前置放大器设计的研究[J].仪器仪表学报,2003,24(z1):184-186. 被引量:14
  • 2罗本成.高性能脑电波ERP信号放大器的研制[J].电子产品世界,2005,12(10A):90-93. 被引量:5
  • 3[2]Sheffield LT, Berson CA, Remschel DB, et al. Recommendations for standard of instrumentation and practice in the use of ambulstory electrocardiography [J]. Circulation, 1985, (3):626- 636.
  • 4冯应琨 主编.临床脑电图学[M].北京:人民卫生出版社,1979.14-22.
  • 5宁新宝.生物医学电子学[M].长沙:湖南科技出版社,1989.
  • 6余学飞.医学电子仪器原理与设计[M].第一版.广州:华南理工大学出版社,2001.
  • 7Bruce B.Winter John G.Webster.Driven-Right-Leg Circuit Design[C].IEEE Trans.1983
  • 8冈村迪夫[日].OP放大电路设计[M].王玲,徐雅珍,李武平,译.北京:科学出版社,2004.
  • 9佐藤一郎.集成运算放大器电路设计实用手册[M].韩健青,译.北京:学术书刊出版社,1989.
  • 10叶晓东,朱兆达.中值滤波的快速算法[J].信号处理,1997,13(3):227-230. 被引量:10