期刊文献+

Sysmex XE-5000自动血液分析仪的临床评价 被引量:4

Clinical Evaluation of Sysmex XE-5000 Hematology Analyzer
下载PDF
导出
摘要 目的 对Sysmex XE-5000全自动血细胞分析仪的主要性能进行评价.方法 对Sysmex XE-5000的精密度、线性范围、与镜检的相关性和交叉污染率进行测定,并与Sysmex XE-2100血细胞分析仪进行比较.同时,对微量血和体液模式进行了评估.结果 Sysmex XE-5000全自动血细胞分析仪精密度表现优异,重复性、稳定性好,线性范围宽,交叉污染率低.通过XE-5000分析,微量血与静脉血血细胞分析结果基本没有差异.体液模式脑脊液细胞计数、分类与人工镜检表现出良好的相关性.结论 Sysmex XE-5000全自动血细胞分析仪是一种较理想的血细胞分析仪. Objective To evaluate the major performance of Sysmex XE-5000 hematology analyzer. Methods Precision,carry-over rate,total reproducibility, leukocyte differential count reproducibility and linear range of Sysmex XE-5000 hematology analyzer were validated and we compared it with Sysmex XE-2100 hematology analyzer as well.Capillary blood and body fluid function were also evaluated. Results Compared with Sysmex XE-2100,the precision,total reproducibility and linear range of Sysmex XE-5000 hematology analyzer were better.Cross contamination rate of WBC,RBC,HGB,HCT and PLT examination were less than l%;There is no significant difference between capillary and venous blood count measurements using XE-5000.The capillary approach provides reliable blood counts.Cell count and differentiation of cerebrospinal fluid cells using XE-5000 showed good agreement between the automated and manual methods. Conclusion Sysmex XE-5000 hematology analyzer is ideal product for its exact and reliable performance.
出处 《中国血液流变学杂志》 CAS 2011年第3期493-497,502,共6页 Chinese Journal of Hemorheology
基金 天津市自然科学基金基础项目(08JCYBJC06500)
关键词 有核红细胞 血细胞分析仪 Nucleated red blood cell Blood cell analyzer
  • 相关文献

参考文献5

  • 1ICSH.Guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting and cell marker applicat- ions.International Council for Standardization in Haematology:prepared by the ICSH Expert Panel on Cytometry[J].Clin Lab Haematol,1994,16(2):157 - 174.
  • 2Gulati G,Behling E,Kocher W, et al.An evaluation ofthe performance of Sysmex XE-2100 in enumerating nucleated red cells in peripheral blood[J].Arch Pathol Lab Med,2007,131(7):1077 - 1083.
  • 3Field D,Taube E,Heumann S.Performance evaluation of the immature granulocyte parameter on the Sysmex XE-2100 automated hematology analyzer[J].Lab Hematol,2006,12(l): 11 - 14.
  • 4Novis DA,Walsh M,Wilkinson D,et al.Laboratory productivity and the rate of manual peripheral blood smear review:a College of American Pathologists Q-Probes study of 95,141 complete blood count determinations performed in 263 institutions[J]. Arch Pathol Lab Med,2006,130(5):596 - 601.
  • 5曹增,李勇,幕悦意,张云霞,夏永辉,田鹏,杨佳彧,李筱梅.Sysmex XE-5000血细胞分析仪白细胞分类性能评价[J].中国血液流变学杂志,2009(4):643-645. 被引量:10

二级参考文献6

  • 1陈梅,肖旺贤,段朝晖,付玉如,方伟祯.Sysmex XE-2100全自动血细胞分析仪对形态异常细胞提示功能评价[J].实用全科医学,2007,5(3):257-258. 被引量:12
  • 2National Committee for Clinical Laboratory Standards.Reference leukocyte differential count (proportional)and evaluation of instrument methods:approved standard[Z].NCCLS Document H20-A.NCCLS,Wayne(USA), 1992.
  • 3Field D,Taube E,Heumann S.Performance evaluation of the immature granulocyte parameter on the Sysmex XE-2100 automated hematology analyzer[J].Lab Hematol,2006,12(1): 11-14.
  • 4Novis DA, Walsh M, Wilkinson D, et al.Laboratory, productivity and the rate of manual peripheral blood smear review: a College of American Pathologists Q-Probes study of 95,141 complete blood count determinations performed in 263 institutions[J].Arch Pathol Lab Med,2006,130(5):596-601.
  • 5Barnes PW,McFadden SL,Machin SJ,et al.The international consensus group for hematology review:suggested criteria for action following automated CBC and WBC differential analysis[J].Lab Hematol,2005,11 (2):83-90.
  • 6江虹,曾婷婷,曾素根,郭曼英,朱新勤,左永太,李倚,贾永前.自动全血细胞分析和白细胞分类复检规则的制定及评价[J].中华检验医学杂志,2007,30(9):996-1000. 被引量:24