CT-超声实时影像虚拟导航引导穿刺活检的实验研究 被引量:8

Biopsy guided by real-time virtual navigation system using computed tomography imaging data:a phantom study
原文传递
导出
摘要 目的 利用体模验证CT-超声实时影像虚拟导航(real-time virtual navigation system,RVS)引导活检的可行性及准确性.方法 体外模型内含30个红色球形胶块,胶块CT可见但超声不可见.两名不同经验操作者针对每个胶块以RVS引导与常规超声引导两种方法各活检一次,记录活检是否成功及检出的目标胶条长度.结果 两名操作者无论采用RVS引导还是常规超声引导,其活检成功率及检出的目标胶条长度差异均无统计学意义;RVS引导法及常规超声引导法活检成功率分别为96.7%(58/60)及43.3%(26/60),检出目标胶条长度的中位数分别为7.5(6~9)mm及0(0~5)mm.RVS引导法活检成功率及检出目标胶条的长度均高于常规超声引导法,差异有统计学意义(P<0.05).结论 RVS引导活检超声不可见目标是可行的,且具较高的准确性,有望为介入性超声提供一种新的引导方法. Objective To verify the feasibility and the accuracy for the biopsy of targets not visible on sonography but shown on CT in a phantom by using real-time virtual navigation system(RVS).Methods The tissue mimicking phantom,made of gel,contained 30 red spheres which were 10 mm in diameter,randomly placed.These spheres were invisible on sonography,but visible on CT.A junior operator and a senior operator performed the biopsy respectively,and each sphere was biopsied once guide by both RVS and conventional ultrasound for each operator.The red dye in the biopsied sample meant successful performance.The success rate of biopsy and the length of the red dyed sample were recorded.Results There was no statistically difference between the performance results of the two operators.The success rate and the median length of red dyed sample biopsy of RVS were respectively 96.7% (58/60) and 7.5 mm (rang,6-9 mm) which were statistically higher and longer than that of conventional ultrasound,43.3%(26/60) and 0 mm(rang,0 - 5 mm) ( P 〈0.05).Conclusions Biopsy guided by RVS for the target invisible to ultrasound is a feasible and accuracy method,which will be a new guided method for interventional ultrasound.
出处 《中华超声影像学杂志》 CSCD 北大核心 2011年第10期887-889,共3页 Chinese Journal of Ultrasonography
基金 广东省科技计划项目(20098060700026,20108031600041) 2007年公益性行业科研专项经费(2007-2-10)
关键词 超声检查 体模显象术 活组织检查 针吸 实时影像虚拟导航 Ultrasonography Phantoms, imaging Biopsy, needle Real-time virtual navigation system
  • 相关文献

参考文献6

  • 1Ewertsen C.Image fusion between ultrasonography and CT,MRI or PET/CT for image guidance and intervention-a theoretical and clinical study.Dan Med Bull,2010,57:B4172.
  • 2Molloy JA,Oldham SA.Benchmarking a novel ultrasound-CT fusion system for respiratory motion management in radiotherapy:assessment of spatio-temporal characteristics and comparison to 4DCT.Med Phys,2008,35:291-300.
  • 3Ewertsen C,Grossjohann HS,Nielsen KR,et al.Biopsy guided by real-time sonography fused with MRI:a phantom study.Am J Roentgenol,2008,190:1671-1674.
  • 4Kliewer MA,Sheafor DH,Paulson EK,et al.Percutaneous liver biopsy:a cost-benefit analysis comparing sonographic and CT guidance.Am J Roentgenol,1999,173:1199-1202.
  • 5Shin HJ,Amaral JG,Armstrong D,et al. Image-guided percutaneous biopsy of musculoskeletal lesions in children.Pediatr Radiol,2007,37:362-369.
  • 6Kang M,Kalra N,Gulati M,et al.Image guided percutaneous splenic interventions.Eur J Radiol,2007,64:140-146.

同被引文献66

  • 1李伟,朱学峰.医学图像融合技术及其应用[J].中国医学影像技术,2005,21(7):1126-1129. 被引量:24
  • 2刘广健,吕明德,谢晓燕,周旭辉,徐辉雄,匡铭,徐作峰,郑艳玲,梁瑾瑜,黄蓓.实时虚拟导航系统引导消融治疗肝癌[J].中华超声影像学杂志,2006,15(10):758-760. 被引量:15
  • 3徐静,杨向东,朱森强,鲁通,任贺,梁萍,陈恳.用于肝癌介入治疗的术中三维超声导航系统[J].中国生物医学工程学报,2007,26(5):719-723. 被引量:8
  • 4Zhu XL,Chen P,Guo H,et al. Contrast enhanced ultrasound for the diagnosis of hepatic adenoma. J Int Med Res. 2011 ,39:920-928.
  • 5Martie A, Sporea I, Popescu A, et al. Contrast enhanced ultrasound for the characterization of hepatocellular carcinoma. Med Ultrason,2011,13 : 108-113.
  • 6Sporea I, Badea R, Martie A, el al. Contrast enhanced ultrasound for the evaluation of focal liver lesions in daily practice. A multicentre study. Med Ultrason, 2012,14 : 95-100.
  • 7Seitz K, Bernatik T, Strobel D, et al. Contrast enhanced ultrasound (CEUS) for the characterization of focal liver lesions in clinical practice: CEUS vs MRI a prospective comparison in 269 patients. Ultraschall Med,2010,31:492-499.
  • 8Feier D, Socaciu M, Anton O, et al. The combined role of intravenous contrast enhanced ultrasound (CEUS) and computed tomography (CT) in liver abscess diagnosis. Chirurgia (Bucur), 2012,107:343-351.
  • 9Jung EM, Schreyer AG, Schacherer D, et al. New real-time image fusion technique for characterization of tumor vascularisation and tumor perfusion of liver tumors with contrast-enhanced ultrasound, spiral CT or MRI: First results. Clin Hemorheol Microcirc, 2009,43 : 57-69.
  • 10Ewertsen C, Henriksen BM, Torp Pedersen S, et al. Characterization by biopsy or CEUS of liver lesions guided by image fusion between ultrasonography and CT, PET/CT or MRI. Ultraschall Med, 2011,32 : 191-197.

引证文献8

二级引证文献26

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部