摘要 目的 探讨CTVision图像引导下的放射治疗系统的临床摆位误差.方法 应用西门子CTVision图像引导放射系统治疗33例患者,其中头颈部肿瘤20例,胸部肿瘤6例,腹盆部肿瘤7例.从首次放疗起每周1次于放疗前获取CT图像,将CT图像和计划图像及其靶中心匹配,获得靶中心x(左右)、y(头脚)、z(前后)方向的误差,分析误差及其分布规律.结果 头颈部肿瘤119次扫描中78次(65.5%)3个方向误差均<2mm,胸部肿瘤36次扫描中14次(38.9%)3个方向误差均<3mm,腹盆部肿瘤42次扫描中10次(23.8%)3个方向误差均<3 mm.头颈部、胸部、腹盆部肿瘤摆位误差均呈正态分布.结论 CTVision图像引导测量治疗前摆位误差并实时调整,大大提高患者的摆位精度。
出处 《肿瘤研究与临床》 CAS 2011年第10期709-711,共3页 Cancer Research and Clinic
  • 相关文献

参考文献8

  • 1Sweeney R,Bale R,Vogele M,et al.Repositioning accuracy:comparison of a noninvasive head holder with thermoplastic mask for fractionated radiotherapy and a case report.Int J Radiat Oncol Biolphys,1998,41:475-483.
  • 2Zhu Y,Stovall J,Butler L,et al.Comparison of two immobilization techniques using portal film and digitally reconstructed radiographs for pediatric patients with brain tumors.Int J Radiat Oncol Biol Phys,2000,48:1233-1240.
  • 3Gilbeau L,Octave PM,Lonco T,et al.Comparison of setup accuracy of three different thermoplastic masks for the treatment of brain and head and neck tumors.Radiother Oncol,2001,58:155-162.
  • 4Ohara K,Okumura T,Akisada M,et al.Irradiation synchronized with respiration gate.Int J Radiat Oncol Biol Phys,1989,17:853-857.
  • 5Shimizu SS,Shirato H,Kagei K,et al.Impact of respiratory movement in the computed tomographic images of small lung tumors in three-dimensional (3D) radiotherapy.Int J Radiat Oncol Biol Phys,2000,46:1127-1133.
  • 6Ekberg L,Holmberg O,Wittgren L,et al.What margins should be added to the clinical target volume in radiotherapy treatment planning for lung cancer.Radiother Oncol,1998,48:71-77.
  • 7Matthias G,Juergen M,Dirk V,et al.Magnitude and clinical relevance of translational and rotational patient setup errors:a cone-beam CT study.Int J Radiat Oncol Biol Phys,2006,5:934-942.
  • 8许峰,柏森,王瑾,张洪,钟仁明,蒋晓芹.用锥形束CT图像测量放疗摆位误差[J].中华放射肿瘤学杂志,2007,16(6):461-464. 被引量:60

二级参考文献15

  • 1Hurkmans CW, Remeijer P, Lebesque JV, et al. Set-up verification using portal imaging: review of current clinical practice. Radiother Oncol, 2001,58 : 105-120
  • 2van Herk M. Errors and margins in radiotherapy. Semin Radiat Oncol,2004,14:52-64
  • 3Kutcher GJ, Mageras GS, Liebel SA. Control, correction, and modeling of setup errors and organ motion. Semin Radiat Oncol,1995,5 : 134-145
  • 4van Herk M, Remeijer P, Lebesque JV. Inclusion of geometric uncertainties in treatment plan evaluation. Int J Radiat Oncol Biol Phys, 2002,52 : 1407-1422
  • 5Matthias G, Juergen M, Dirk V, et al. Magnitude and clinical relevance relevance of translational and rotational patient setup errors : a cone-beam CT study. Int J Radiat Oncol Biol Phys,2006,5:934-942
  • 6Catherine AM, Ann MH, Jonathan S, et al. X-ray volumetric imaging in image-guided radiotherapy: the new standard in on-treatment imaging. Int J Radiat Oncol Biol Phys, 2006,64:625-634.
  • 7Michalski J, Purdy JA. Image-guided radiation therapy committee.Int J Radiat Oncol Biol Phys,2001,51:60-65
  • 8Mackie TR, Kapatoes J, Ruchala K, et al. Image guidance for precise conformal radiotherapy. Int J Radiat Oncol Biol Phys, 2003,56 : 89-105
  • 9Sidhu K, Ford EC, Spirou S, et al. Optimization of conformal thoracic radiotherapy using cone-beam CT imaging for treatment verification. Int J Radiat Oncol Biol Phys, 2003,55:757-767
  • 10Brock KK, Dawson LA, Sharpe MB, et al. Feasibility of a novel deformable image registration technique to facilitate classification, targeting, and monitoring of tumor and normal tissue. Int J Radiat Oncol Biol Phys, 2006,64:1245-1254

共引文献59

同被引文献5

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部